Monte Carlo study of dosimetric impact of gadolinium contrast medium in transverse field MR-Linac system

被引:0
|
作者
Ohno, Takeshi [1 ]
Kubota, Takahiro [2 ]
Yano, Masayuki [3 ]
Fujiwara, Yasuhiro [1 ]
Araki, Fujio [1 ]
机构
[1] Kumamoto Univ, Fac Life Sci, Dept Hlth Sci, 4-24-1 Kuhonji, Kumamoto, Japan
[2] Kokura Mem Hosp, 3-2-1 Asano, Fukuoka, Japan
[3] Saga Heavy Ion Med Accelerator Tosu, Koga Machi, Tosu, Saga, Japan
关键词
Gadolinium contrast medium; MR-Linac; Magnetic field; Monte Carlo technique; Dose distribution; MV RADIOTHERAPY ACCELERATOR; MEGAVOLTAGE PHOTON BEAMS; HIGH SIGNAL INTENSITY; DENTATE NUCLEUS; MAGNETIC-FIELD; DOSE INCREASE; SCANNER; NANOPARTICLES; TRANSPORT; AGENT;
D O I
10.1016/j.ejmp.2021.05.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study is to evaluate the dosimetric impact of gadolinium contrast medium (Gadovist) in a transverse MR-Linac system using Monte Carlo methods. The dose distributions were calculated using two heterogeneous multi-layer phantoms consisting of Gadovist, water, bone, and lung. The photon beam was irradiated with a filed size of 5 x 5 cm2, and a transverse magnetic field of 0-3.0 T was applied perpendicular to the incident photon beam. Next, dose distributions for brain, head and neck (H&N), and lung cancer patients were calculated using a patient voxel-based phantom with and without replacing the patient's GTV with Gadovist. The dose at the water-Gadovist interface increased by 8% without a magnetic field. A similar dose increment was observed at 0.35 T. In contrast, the dose increment at the water-Gadovist interface was small at 1.5 T and a dose decrement of 5% was observed at 3.0 T. The dose variation at the lung-Gadovist interface was larger than that at the water-Gadovist interface. The mass collision stopping power ratio for Gadovist was 7% lower than that for water, whereas, the electron fluence spectra at the water-Gadovist interface increased by 17.5%. In a patient study, Gadovist increased the Dmean for brain, H&N, and lung cancer patients by 0.65-8.9%. The dose variation due to Gadovist grew large in the low-dose region in H&N and lung cancer. The GTV dose variation due to Gadovist in all treatment site was below 2% at 0-3 T if the Gadovist concentration was lower than 0.2 mmol/ml-1.
引用
收藏
页码:19 / 30
页数:12
相关论文
共 50 条
  • [21] Towards magnetic-field-modulated radiotherapy (MagMRT) with an MR-LINAC-a Monte Carlo study
    Chu, Vivien W. S.
    Kan, Monica W. K.
    Wong, Kenneth C. W.
    Lee, Louis K. Y.
    Chan, Anthony T. C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (20):
  • [22] The impact of a 1.5 T MR-Linac fringe field on neighbouring linear accelerators.
    Perik, T.
    Kaas, J.
    Wittkamper, F.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S230 - S231
  • [23] The dosimetric impact of geometric image distortions in slice-based 4D-MRI on the MR-linac
    Fast, M.
    Keesman, R.
    Van den Wollenberg, W.
    Juan-Cruz, C.
    Van de Lindt, T.
    Van der Heide, U.
    Sonke, J.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S537 - S538
  • [24] Dosimetric Evaluation for MR-Linac Online Adaptive Planning Approaches: A Case Study of Recurrent Adrenocortical Carcinoma
    Niedzielski, J.
    Koong, A.
    Sobremonte, A.
    Castillo, P.
    Lee, B.
    Yang, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [25] Online 3D prostate cine-MRI on the MR-linac: dosimetric impact of intrafraction motion
    Kontaxis, C.
    Keizer, D. De Muinck
    Willigenburg, T.
    Den Hartogh, M.
    Kerkmeijer, L.
    van Zyp, J. Van der Voort
    Raaymakers, B.
    Lagendijk, J.
    De Boer, H.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S65 - S65
  • [26] Evaluation of Gastrointestinal Air Cavity Variations and Dosimetric Impact for Pancreas SBRT on a 0.35T MR-Linac
    Gao, Y.
    Skinner, L.
    Hristov, D.
    Chuang, C. F. Y.
    Liu, L.
    Pham, D.
    Ngo, N.
    Nguyen, N.
    Vitzthum, L.
    Pollom, E.
    MEDICAL PHYSICS, 2024, 51 (10) : 7784 - 7784
  • [27] Experimental verification the electron return effect around spherical air cavities for the MR-Linac using Monte Carlo calculation
    Shortall, J.
    Osorio, E.
    Aitkenhead, A.
    Berresford, J.
    Agnew, J.
    Budgell, G.
    Chuter, R.
    McWilliam, A.
    Kirkby, K.
    Mackay, R.
    Herk, M.
    MEDICAL PHYSICS, 2020, 47 (06) : 2506 - 2515
  • [28] Development and validation of a 1.5 T MR-Linac full accelerator head and cryostat model for Monte Carlo dose simulations
    Friedel, M.
    Nachbar, M.
    Moennich, D.
    Dohm, O.
    Thorwarth, D.
    MEDICAL PHYSICS, 2019, 46 (11) : 5304 - 5313
  • [29] Dosimetric Impact of Bulk Assignment Versus Deformed CT-Based Electron Density in Abdominal MR-Linac Planning
    Romaguera, T.
    Godley, A.
    Alvarez, D.
    Mishra, V.
    Gutierrez, A.
    Luciani, G.
    MEDICAL PHYSICS, 2019, 46 (06) : E318 - E318
  • [30] Clinical acceptance and dosimetric impact of automatically delineated elective target and organs at risk for head and neck MR-Linac patients
    Koteva, Vesela
    Eiben, Bjorn
    Dunlop, Alex
    Gupta, Amit
    Gangil, Tarun
    Wong, Kee Howe
    Breedveld, Sebastiaan
    Nill, Simeon
    Harrington, Kevin
    Oelfke, Uwe
    FRONTIERS IN ONCOLOGY, 2024, 14