Comparison of three different chemiluminescence assays and a rapid liquid chromatography tandem mass spectrometry method for measuring serum aldosterone

被引:16
|
作者
Yin, Yicong [1 ,2 ]
Ma, Chaochao [1 ,2 ]
Yu, Songlin [1 ,2 ]
Liu, Wenjing [3 ]
Wang, Danchen [1 ,2 ]
You, Tingting [1 ,2 ]
Cheng, Qian [1 ,2 ]
Qiu, Ling [1 ,2 ]
机构
[1] Peking Union Med Coll, Peking Union Med Coll Hosp, Dept Clin Lab, Beijing 100730, Peoples R China
[2] Chinese Acad Med Sci, Beijing 100730, Peoples R China
[3] Waters Corp Shanghai Sci & Technol Co Ltd, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
aldosterone; chemiluminescence immunoassay; liquid chromatography tandem mass spectrometry; inconsistency; primary aldosteronism; recalibration; PRIMARY HYPERALDOSTERONISM; IMMUNOASSAY; PLASMA; PREVALENCE; OUTCOMES;
D O I
10.1515/cclm-2019-0706
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: This study aimed to quantify and compare serum aldosterone (sALD) levels through three different chemiluminescence immunoassays (CLIAs) and liquid chromatography tandem mass spectrometry (LC-MS/MS) analysis. Methods: Serum samples from 221 patients with suspected primary aldosteronism (PA) were retrospectively included in this study conducted at the Peking Union Medical College Hospital from June to August in 2017. sALD levels were determined using the LC-MS/MS method and three different CLIA systems, viz., DiaSorin (R) XL, iSYS and Auto Lumo A2000. Pooled fresh serum samples were used for recalibration. Passing-Bablok regression analysis, correlation matrix, and Bland-Altman plots were used to evaluate the concurrence among ALD levels determined using the three CLIAs. Results: Within-laboratory precision of the four assays ranged from 2.1% to 9.4%, except the coefficient variation (CV) of one of the CLIAs, which exceeded 20.0% for samples with low sALD levels. sALD levels determined using LC-MS/MS were significantly lower than those determined using the other three CLIAs (p <0.0001). Spearman's correlation coefficient of the four assays ranged from 0.745 to 0.950 (p < 0.0001). The Bland-Altman plot showed that the average bias WO for the three CLIAs and LC-MS/MS ranged from -69.3 to -49.2. After recalibration, this correlation did not improve among the assays. However, the bias and bias percentage at the medical decision level improved between LC-MS/MS and DiaSorin (R) XL/iSYS. Conclusions: Significant inconsistencies between the results of CLIAs and LC-MS/MS indicate that different sALD measures cannot be used interchangeably.
引用
收藏
页码:95 / 102
页数:8
相关论文
共 50 条
  • [1] Establishment of a rapid and simple liquid chromatography tandem mass spectrometry method for measuring aldosterone in urine
    Yin, Yicong
    Yu, Songlin
    Qiu, Ling
    Wang, Xiaoxia
    Wang, Danchen
    Ma, Chaochao
    Zhang, Kai
    You, Tingting
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2019, 1113 : 84 - 90
  • [2] Establishment of a rapid and simple liquid chromatography tandem mass spectrometry method for measuring aldosterone in urine
    Yin, Y.
    Yu, S.
    Qiu, L.
    CLINICA CHIMICA ACTA, 2019, 493 : S48 - S48
  • [3] Calibrators for measuring aldosterone by liquid chromatography-tandem mass spectrometry
    Taylor, Paul J.
    Gordon, Richard D.
    Stowasser, Michael
    CLINICA CHIMICA ACTA, 2012, 413 (1-2) : 346 - 347
  • [4] Determination of aldosterone in serum by liquid chromatography-tandem mass spectrometry
    Turpeinen, Ursula
    Hamalainen, Esa
    Stenman, Ulf-Hakan
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 862 (1-2): : 113 - 118
  • [5] Comparison of plasma aldosterone measured by chemiluminescence immunoassay and liquid chromatography-tandem mass spectrometry in screening test for primary aldosteronism
    Chen, Wenzhan
    Lai, Fenghua
    Huang, Xiaoyu
    Yu, Shuang
    Chen, Nan
    Xu, Changliu
    Wang, Chenxue
    Liang, Shuhui
    Li, Yanbing
    Xiao, Haipeng
    Cao, Xiaopei
    PRACTICAL LABORATORY MEDICINE, 2024, 39
  • [6] Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid-liquid extraction method for the ABSCIEX API-5000 mass spectrometry system
    Van der Gugten, Jessica Grace
    Dubland, Joshua
    Liu, Hua-Fen
    Wang, Alex
    Joseph, Christine
    Holmes, Daniel T.
    JOURNAL OF CLINICAL PATHOLOGY, 2012, 65 (05) : 457 - 462
  • [7] Rapid liquid chromatography-electrospray tandem mass spectrometry method for serum free and total carnitine
    Ho, CS
    Cheng, BSS
    Lam, CWK
    CLINICAL CHEMISTRY, 2003, 49 (07) : 1189 - 1191
  • [8] Rapid and sensitive liquid chromatography-tandem mass spectrometry method for the determination of leuprolide in human serum
    Zhan, Yan
    Chen, Xiaoyan
    Zhao, Xiaohua
    Zhong, Dafang
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (27): : 3194 - 3200
  • [9] A Rapid Method to Determine Tetrabromobisphenol A in Rat Serum and Urine by Liquid Chromatography-Tandem Mass Spectrometry
    Kang, Mi Jeong
    Kim, Ju Hyun
    Lee, Sang Kyu
    Kang, Wonku
    Kim, Hyung Sik
    Lyoo, Won Seok
    Jeong, Tae Cheon
    ARCHIVES OF PHARMACAL RESEARCH, 2010, 33 (11) : 1797 - 1803
  • [10] A rapid method to determine tetrabromobisphenol a in rat serum and urine by liquid chromatography-tandem mass spectrometry
    Mi Jeong Kang
    Ju Hyun Kim
    Sang Kyu Lee
    Wonku Kang
    Hyung Sik Kim
    Won Seok Lyoo
    Tae Cheon Jeong
    Archives of Pharmacal Research, 2010, 33 : 1797 - 1803