Determination of patient set-up error and optimal treatment margin for intensity modulated radiotherapy using image guidance system

被引:5
|
作者
Kanakavelu, Nithya [1 ,2 ]
Samuel, James Jebaseelan [2 ]
机构
[1] Medwin Canc Ctr, Dept Radiat Oncol, Hyderabad, Andhra Pradesh, India
[2] VIT Univ, Photon Nucl & Med Phys Div, Vellore 632014, Tamil Nadu, India
来源
JOURNAL OF BUON | 2016年 / 21卷 / 02期
关键词
image guided radiotherapy; intensity modulated radiotherapy; patient set-up errors; DOSIMETRIC CONSEQUENCES; QUALITY-ASSURANCE; PTV MARGIN; HEAD; IMRT;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The geometrical uncertainties in the patient positioning during intensity modulated radiotherapy (IMRT) are crucial as there is potential to underdose the tumor and overdose the nearby critical structures. Image guided techniques provide a solution to assess the patient set-up uncertainties and help determine the optimal planning target volume (PTV) margin to the clinical tumor volume (CTV). Methods: A retrospective study was conducted to evaluate patient set-up errors along the three translational directions at different treatment sites such as the brain, the head and neck (H&N) and the prostate. A total of 60 patients' set-up error data was analysed to evaluate the systematic and random errors and the optimal CTV-PTV margin. Results: For brain and H&N sites, more than 90, 80 and about 100% of the total image acquisitions were less than 3 mm in lateral, longitudinal and vertical directions respectively. For the prostate cases, the frequency of patient set-up error to be less than 3 mm were 79.7, 75.6 and 80% in lateral, longitudinal and vertical directions respectively. About 0.6% had more than 7 mm error in the lateral and longitudinal directions for the prostate site. CTV-PTV margin of 3.4, 3.4 and 1.9 mm for brain cases, 3.5, 3 and 1.8 mm for H&N cases and 5, 4.6 and 4.5 mm for the prostate cases in the lateral, longitudinal and vertical directions respectively were determined. Conclusion: Image guidance is an effective method to evaluate the accuracy of IMRT treatment delivery. The optimal CTV-PTV margin can be determined to ensure adequate dose to CTV, specific to the site.
引用
收藏
页码:505 / 511
页数:7
相关论文
共 50 条
  • [31] Set-up error and dosimetric analysis of HexaPOD evo RT 6D couch combined with cone beam CT image-guided intensity-modulated radiotherapy for primary malignant tumor of the cervical spine
    Jiang, Ping
    Zhang, Xile
    Wei, Shuhua
    Zhao, Tiandi
    Wang, Junjie
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (04): : 22 - 30
  • [32] Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study
    Saito, Masahide
    Ueda, Koji
    Suzuki, Hidekazu
    Komiyama, Takafumi
    Marino, Kan
    Aoki, Shinichi
    Sano, Naoki
    Onishi, Hiroshi
    JOURNAL OF RADIATION RESEARCH, 2022, 63 (03) : 435 - 442
  • [33] Evaluation of the validity of a convolution method for incorporating tumour movement and set-up variations into the radiotherapy treatment planning system
    McCarter, SD
    Beckham, WA
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (04): : 923 - 931
  • [34] Definition of the set-up error using 2D radiographic images and 3D computed tomography and dosimetric effect of set-up errors for head and neck/lung radiotherapy patients
    Ceylan, Cemile
    Arslan, Abdulkadir
    Kafkas, Baris
    Cekickesen, Yucel
    Molaoglu, Figen
    Tecik, Abidin
    Korkmaz, Nursad
    Ayata, Hande
    Kilic, Ayhan
    Ugur, Timur
    Guden, Metin
    Engin, Kayihan
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2015, 30 (03): : 118 - 132
  • [35] Spine radiosurgery for the local treatment of spine metastases: Intensity-modulated radiotherapy, image guidance, clinical aspects and future directions
    de Moraes, Fabio Ynoe
    Taunk, Neil Kanth
    Laufer, Ilya
    Pimenta Neves-Junior, Wellington Furtado
    Hanna, Samir Abdallah
    Carvalho, Heloisa de Andrade
    Yamada, Yoshiya
    CLINICS, 2016, 71 (02) : 101 - 109
  • [36] Set-up Error Correction Using Internal Markers on Patients with Pancreatic Adenocarcinoma Receiving Stereotactic Body Radiotherapy and its Association with the Patient's Body Mass Index
    Lin, C.
    Chen, S.
    Enke, C. A.
    Zhou, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S333 - S333
  • [37] A new device of real time dose monitoring system for preventing the set-up error with using flat panel detectors (FDP) in patient with breast cancer
    Kamikonya, N.
    Takada, Y.
    Fujiwara, M.
    Tsuboi, K.
    Nakao, N.
    Nakagawa, H.
    Tanooka, M.
    Yamamoto, E.
    Inoue, H.
    Yasumasa, K.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S628 - S629
  • [38] IMAGE-GUIDED STEREOTACTIC BODY RADIOTHERAPY FOR LUNG TUMORS USING BODYLOC WITH TOMOTHERAPY: CLINICAL IMPLEMENTATION AND SET-UP ACCURACY
    Zhou, Jining
    Uhl, Barry
    Dewitt, Kelly
    Young, Mark
    Taylor, Brian
    Fei, Ding-Yu
    Lo, Yeh-Chi
    MEDICAL DOSIMETRY, 2010, 35 (01) : 12 - 18
  • [39] Breast radiotherapy: comparation of set up error using All In One system and dedicated breast board
    Manestar, V.
    Karadza, V.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S994 - S994
  • [40] Modeling set-up error by daily MVCT for prostate adjuvant treatment delivered in 20 fractions: Implications for the assessment of the optimal correction strategies
    Broggi, Sara
    Cozzarini, Cesare
    Fiorino, Claudio
    Maggiulli, Eleonora
    Alongi, Filippo
    Cattaneo, Giovanni Mauro
    Di Muzio, Nadia
    Calandrino, Riccardo
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) : 246 - 252