Subcutaneous implantable cardioverter defibrillators in patients with left ventricular assist devices: case report and comprehensive review

被引:10
|
作者
Lopez-Gil, Maria [1 ,2 ]
Fontenla, Adolfo [1 ,2 ]
Delgado, Juan F. [2 ,3 ,4 ,5 ]
Rodriguez-Munoz, Daniel [1 ]
机构
[1] 12 Octubre Univ Hosp, Cardiol Dept, Cardiac Electrophysiol & Arrhythmia Unit, Avda Cordoba S-N, Madrid 28041, Spain
[2] 12 Octubre Univ Hosp, Avda Cordoba S-N, Madrid 28041, Spain
[3] 12 Octubre Univ Hosp, Cardiol Dept, Heart Failure & Heart Transplantat Program, Avda Cordoba S-N, Madrid 28041, Spain
[4] Univ Complutense, Fac Med, Avda Seneca 2, Madrid 28040, Spain
[5] CIBER CV, C Melchor Fernandez Almagro 3, Madrid 28029, Spain
关键词
Left ventricular assist devices; Subcutaneous implantable cardioverter defibrillator; Electromagnetic interferences; Undersensing; Case report; ELECTRONIC DEVICES;
D O I
10.1093/ehjcr/ytz057
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Left ventricular assist devices (LVAD) are increasingly used in patients with advanced heart failure, many of whom have been or will be implanted with an implantable cardioverter defibrillator (ICD). Interaction between both devices is a matter of concern. Subcutaneous ICD (S-ICD) obtains its signals through subcutaneous vectors, which poses special challenges with regards to adequate performance following LVAD implantation. Case summary We describe the case of a 24-year-old man implanted with an S-ICD because of idiopathic dilated cardiomyopathy, severe biventricular dysfunction, and self-limiting sustained ventricular tachycardias. After the implantation of a HeartMate 3TM (Left Ventricular Assist System, Abbott) several months later, the S-ICD became useless because of inappropriate sensing due to electromagnetic interference and attenuation of QRS voltage. Discussion We reviewed the reported cases in PubMed about the concomitant use of S-ICD and LVAD. Seven case reports about the performance of S-ICD in patients with an LVAD were identified, with discordant results. From these articles, we analyse the potential causes for these differing results. Pump location and operating rates in LVAD, as well as changes in the subcutaneous-electrocardiogram detected by the S-ICD after LVAD implantation are related to sensing disturbances when used in the same patient.
引用
下载
收藏
页数:6
相关论文
共 50 条
  • [31] Subcutaneous implantable cardioverter-defibrillator troubleshooting in patients with a left ventricular assist device: A case series and systematic review
    Black-Maier, Eric
    Lewis, Robert K.
    Barnett, Adam S.
    Pokorney, Sean D.
    Sun, Albert Y.
    Koontz, Jason I.
    Daubert, James P.
    Piccini, Jonathan P.
    HEART RHYTHM, 2020, 17 (09) : 1536 - 1544
  • [32] Implantable Cardioverter Defibrillators in Patients with Continuous Flow Left Ventricular Assist Devices: Utilization Patterns, Shock Therapies and Complications
    Alvarez, Paulino
    Sperry, Brett
    Perez, Antonio
    Starling, Randall C.
    JOURNAL OF CARDIAC FAILURE, 2018, 24 (08) : S115 - S116
  • [33] The role of implantable cardioverter-defibrillators in patients with continuous flow left ventricular assist devices - A meta-analysis
    Agrawal, Sahil
    Garg, Lohit
    Nanda, Sudip
    Sharma, Abhishek
    Bhatia, Nirmanmoh
    Manda, Yugandhar
    Singh, Amitoj
    Fegley, Mark
    Shirani, Jamshid
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 222 : 379 - 384
  • [34] Prognostic impact of implantable cardioverter defibrillators and associated adverse events in patients with continuous flow left ventricular assist devices
    Pausch, Jonas
    Mersmann, Julian
    Bhadra, Oliver D.
    Barten, Markus J.
    Toennis, Tobias
    Yildirim, Yalin
    Pecha, Simon
    Reichenspurner, Hermann
    Bernhardt, Alexander M.
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
  • [35] Implantable Cardioverter Defibrillators in Patients With Continuous Flow Left Ventricular Assist Devices: Utilization Patterns, Related Procedures, and Complications
    Alvarez, Paulino A.
    Sperry, Brett W.
    Perez, Antonio L.
    Yaranov, Dmitry M.
    Randhawa, Varinder
    Luthman, Jacob
    Cantillon, Daniel J.
    Starling, Randall C.
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2019, 8 (14):
  • [36] Implantable Cardioverter-Defibrillator Events in Patients with Left Ventricular Assist Devices
    Ambardekar, Amrut V.
    Sauer, William H.
    Brieke, Andreas
    Lowery, Christopher M.
    Allen, Larry A.
    JOURNAL OF CARDIAC FAILURE, 2009, 15 (06) : S47 - S48
  • [37] Should subcutaneous implantable cardioverter-defibrillators be implanted in patients who are candidates for continuous flow left ventricular assist device?
    Raffa, Giuseppe Maria
    Morsolini, Marco
    Gentile, Giovanni
    Coppola, Giuseppe
    Sciacca, Sergio
    EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2017, 43 : E30 - E32
  • [38] Implantable Cardioverter Defibrillators and Survival in Continuous-Flow Left Ventricular Assist Device Patients
    Kutyifa, Valentina
    Fernandez, Genaro
    Sherazi, Saadia
    Aktas, Mehmet
    Huang, David
    McNitt, Scott
    Papernov, Anna
    Wang, Meng
    Massey, H. Todd
    Chen, Leway
    Alexis, Jeffrey D.
    ASAIO JOURNAL, 2019, 65 (01) : 49 - 53
  • [39] Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators and left heart assist devices
    Bandorski, Dirk
    Hoeltgen, Reinhard
    Stunder, Dominik
    Keuchel, Martin
    ANNALS OF GASTROENTEROLOGY, 2014, 27 (01): : 3 - 8
  • [40] Electrophysiological interactions between implantable cardioverter defibrillators and left ventricular assist device
    Maffe, Stefano
    Perucca, Antonello
    Signorotti, Fabiana
    Pistono, Massimo
    Paffoni, Paola
    Parravicini, Umberto
    Dellavesa, Pierfranco
    Cucchi, Lorenzo
    Paino, Anna Maria
    Zenone, Franco
    Pardo, Nicolo Franchetti
    Bielli, Massimo
    Zanetta, Marco
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2010, 145 (03) : 620 - 623