Validation of A&D UA-767 device for the self-measurement of blood pressure

被引:82
|
作者
Rogoza, AN [1 ]
Pavlova, TS [1 ]
Sergeeva, MV [1 ]
机构
[1] Dept New Diagnost Methods, Moscow 121552, Russia
关键词
A&D UA-767; self blood pressure measurement; validation; AAMI and BHS protocol;
D O I
10.1097/00126097-200008000-00006
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The validation of self-measurement devices has been recommended. Automatic monitor A & D UA-767 (A&D Company, Ltd, Tokyo, Japan) is well known and widely used, but not tested according to the Association for Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) recommendations. Objective To perform a clinical validation for use by adults of the A&D UA-767 device according to the criteria of the AAMI and a modified BHS protocol. Methods The test concerned 101 subjects (57 men and 44 women) aged 15-85 years with arm circumferences of 22-39 cm, a systolic blood pressure (SBP) range of 89-206 mmHg, and a diastolic blood pressure (DBP) range of 53 - 122 mmHg. For each subject, three readings of the UA-767 were compared with simultaneous auscultatory measurements by two trained independent observers who used a mercury manometer and dual stethoscope. The results were graded according to the BHS (1990 and 1993) and AAMI recommendations. Results Observers showed close agreement, with mean differences of 1.1. +/- 2.4 mmHg for SEP and -0.7 +/- 2.0 mmHg for DBP. The proportion of Values agreeing to within 5, 10 and 15 mmHg were 93, 100, and 100% for SEP and 97, 100, and 100% for DBP for the two observers. The sphygmomanometer measurements were 132 +/- 24/79 +/- 14 mmHg (mean +/- SD). The average difference between the mercury sphygmomanometer and A&D UA-767 readings for SEP and DBP were, respectively, - 0.4 +/- 5.4 and -0.4 +/- 4.8 mmHg. The proportion of values agreeing to within 5,10, and 15 mmHg were 82, 94, and 98% for SEP and 80, 95, and 98% for DBP for the observers and device (A/A grade for BHS). Conclusions For an adult population, the A&D UA-767 device for the self-measurement of blood pressure satisfied the AAMI criteria, achieved a BHS grade of A/A and can therefore be recommended for monitoring blood pressure in home and clinical conditions by patients with mild-to-moderate arterial hypertension. (C) 2000 Lippincott Williams & Wilkins.
引用
收藏
页码:227 / 231
页数:5
相关论文
共 50 条
  • [21] Device for the self-measurement of blood pressure that can monitor blood pressure during sleep
    Chonan, K
    Kikuya, M
    Araki, T
    Fujiwara, T
    Suzuki, M
    Michimata, M
    Hashimoto, J
    Ohkubo, T
    Hozawa, A
    Yamamoto, N
    Miyawaki, Y
    Matsubara, M
    Imai, Y
    BLOOD PRESSURE MONITORING, 2001, 6 (04) : 203 - 205
  • [22] VALIDATION OF A SELF BLOOD PRESSURE MEASUREMENT DEVICE
    FOUQUERAY, B
    JULIEN, J
    PAGNY, JY
    JEUNEMAITRE, X
    SASSANO, P
    BATTAGLIA, C
    PLOUIN, PF
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1988, 81 : 231 - 234
  • [23] Challenges in Blood Pressure Self-Measurement
    Wagner, Stefan
    Toftegaard, Thomas Skjodeberg
    Bertelsen, Olav W.
    INTERNATIONAL JOURNAL OF TELEMEDICINE AND APPLICATIONS, 2012, 2012
  • [24] SELF-MEASUREMENT OF BLOOD-PRESSURE
    WEBER, F
    ANLAUF, M
    MUNCHENER MEDIZINISCHE WOCHENSCHRIFT, 1983, 125 (45): : 1017 - 1021
  • [25] SELF-MEASUREMENT OF BLOOD-PRESSURE
    不详
    BULLETIN OF THE WORLD HEALTH ORGANIZATION, 1988, 66 (02) : 155 - 159
  • [26] SELF-MEASUREMENT OF BLOOD-PRESSURE
    KOENIGSBERGER, CW
    LANCET, 1988, 2 (8613): : 739 - 739
  • [27] SELF-MEASUREMENT OF BLOOD-PRESSURE
    不详
    LANCET, 1988, 2 (8607): : 377 - 378
  • [28] Validation of the A&D UM-101 professional device for office blood pressure measurement according to the international protocol
    Stergiou, G.
    Giovas, P.
    Gkinos, C.
    Tzamouranis, D.
    JOURNAL OF HYPERTENSION, 2007, 25 : S164 - S164
  • [29] TECHNICAL VALIDATION OF OMRON 705IT AND A&D UA-767PLUSBT USING OSCILLOMETRIC WAVEFORM SIMULATORS
    Hoffmann-Petersen, N.
    Newell, S. Rose
    Brantlov, S.
    Pedersen, E. Bjerregaard
    JOURNAL OF HYPERTENSION, 2015, 33 : E285 - E285
  • [30] Validation of the Stabil-O-Graph® blood pressure self-measurement device according to bhs and esh criteria
    Westhoff, T.
    Schmidt, S.
    Zidek, W.
    van der Giet, M.
    JOURNAL OF HYPERTENSION, 2007, 25 : S161 - S161