Weighted schemes applied to 3D-OSEM reconstruction in PET

被引:0
|
作者
Michel, C [1 ]
Liu, X [1 ]
Sanabria, S [1 ]
Lonneux, M [1 ]
Sibomana, M [1 ]
Bol, A [1 ]
Comtat, C [1 ]
Kinahan, PE [1 ]
Townsend, DW [1 ]
Defrise, M [1 ]
机构
[1] Catholic Univ Louvain, PET Lab, B-3000 Louvain, Belgium
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In order to better preserve the Poisson characteristics of PET data when using 3D OSEM iterative reconstruction, we have compared images reconstructed by various data weighting schemes: unweighted, attenuation weighted, attenuation and normalization weighted OSEM 3D, FORE + unweighted and attenuation weighted OSEM 2D using 3DRP and FORE + FBP Images as reference. Due to better noise modeling, 3D weighted OSEM schemes are expected to produce better images (i.e, noise-bias trade-off) especially at low statistics, in short Frames of dynamic studies and when bed overlap is suboptimal, as in 3D wholebody studies. These reconstruction methods were implemented on the ECAT Exact HR and quantification was validated an clinical data. Their influence on physiological parameter measurements was examined on a dynamic brain receptor study with C-11-Flumazenil and a wholebody study with (18)FDG. The 3D weighted OSEM schemes provide unbiased results in volume of interests. The resulting noise reduction may lead to more accurate registration, acquisition of shorter frames and generation of parametric images.
引用
收藏
页码:1152 / 1157
页数:2
相关论文
共 50 条
  • [1] Comparison of 3-D reconstruction with 3D-OSEM and with FORE plus OSEM for PET
    Liu, X
    Comtat, C
    Michel, C
    Kinahan, P
    Defrise, M
    Townsend, D
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (08) : 804 - 814
  • [2] Spatial resolution of the HRRT PET scanner using 3D-OSEM PSF reconstruction
    Olesen, Oline Vinter
    Sibomana, Merence
    Keller, Sune Hogild
    Andersen, Flemming
    Jensen, Jorgen
    Holm, Soren
    Svarer, Claus
    Hojgaard, Liselotte
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3789 - 3790
  • [3] 3D-OSEM versus FORE? plus ?OSEM: Optimal Reconstruction Algorithm for FDG PET with a Short Acquisition Time
    Tsuda, Keisuke
    Suzuki, Takayuki
    Toya, Kazuhito
    Sato, Eisuke
    Fujii, Hirofumi
    [J]. WORLD JOURNAL OF NUCLEAR MEDICINE, 2023, 22 (03) : 234 - 243
  • [4] Full 3D-OSEM reconstruction with compressed response of the system
    Herraiz, JL
    España, S
    Vaquero, JJ
    Desco, M
    Udias, JM
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2952 - 2956
  • [5] 3D-OSEM reconstruction from truncated data in pinhole SPECT
    Zeniya, Tsutomu
    Watabe, Hiroshi
    Sohlberg, Antti
    Inomata, Toru
    Kudo, Hiroyuki
    Iida, Hidehiro
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4205 - 4207
  • [6] 3D-OSEM vs FORE plus OSEM: The optimal reconstruction algorithm in FDG PET tests with short acquisition time
    Tsuda, Keisuke
    Fukushi, Masahiro
    Hirayama, Akira
    Suzuki, Takayuki
    Toya, Kazuhito
    Koyama, Kazuya
    Iwabuchi, Yuto
    Nemoto, Koichi
    Moriyama, Noriyuki
    Fujii, Hirofumi
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [7] 3D-OSEM iterative image reconstruction for high-resolution PET using precalculated system matrix
    Ortuno, Juan E.
    Guerra-Gutiérrez, Pedro
    Rubio, Jose L.
    Kontaxakis, George
    Santos, Andres
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02): : 440 - 444
  • [8] Multi-grid 3D-OSEM reconstruction technique for high resolution rotating-head PET scanners
    Ortuno, Juan E.
    Rubio, Jose L.
    Guerra, Pedro
    Kontaxakis, George
    Santos, Andres
    [J]. 2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2215 - 2218
  • [9] Accelerated 3D-OSEM image reconstruction using a Beowulf PC cluster for pinhole SPECT
    Zeniya, Tsutomu
    Watabe, Hiroshi
    Sohlberg, Antti
    Iida, Hidehiro
    [J]. ANNALS OF NUCLEAR MEDICINE, 2007, 21 (09) : 537 - 543
  • [10] Accelerated 3D-OSEM image reconstruction using a Beowulf PC cluster for pinhole SPECT
    Tsutomu Zeniya
    Hiroshi Watabe
    Antti Sohlberg
    Hidehiro Iida
    [J]. Annals of Nuclear Medicine, 2007, 21 : 537 - 543