The influence of patient centering on CT dose and image noise

被引:160
|
作者
Toth, Thomas
Ge, Zhanyu
Daly, Michael P.
机构
[1] GE Co, CT Syst Engn W 1140, GE Healthcare, Milwaukee, WI 53201 USA
[2] Univ Illinois, Dept Elect & Comp Engn, Electromagnet Lab, Urbana, IL 61801 USA
关键词
D O I
10.1118/1.2748113
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Although x-ray intensity shaping filters (bowtie filters) have been used since the introduction of some of the earliest CT scanner models, the clinical implications on dose and noise are not well understood. To achieve the intended dose and noise advantage requires the patient to be centered in the scan field of view. In this study we explore the implications of patient centering in clinical practice. We scanned various size and shape phantoms on a GE LightSpeed VCT scanner using each available source filter with the phantom centers positioned at 0, 3, and 6 cm below the center of rotation (isocenter). Surface doses were measured along with image noise over a large image region. Regression models of surface dose and noise were generated as a function of phantom size and centering error. Methods were also developed to determine the amount of miscentering using a scout scan projection radiograph (SPR). These models were then used to retrospectively evaluate 273 adult body patients for clinical implications. When miscentered by 3 and 6 cm, the surface dose on a 32 cm CTDI phantom increased by 18% and 41% while image noise also increased by 6% and 22%. The retrospective analysis of adult body scout SPR scans shows that 46% of patients were miscentered in elevation by 20-60 mm with a mean position 23 mm below the center of rotation (isocenter). The analysis indicated a surface dose penalty of up to 140% with a mean dose penalty of 33% assuming that tube current is increased to compensate for the increased noise due to miscentering. Clinical image quality and dose efficiency can be improved on scanners with bowtie filters if care is exercised when positioning patients. Automatically providing patient specific centering and scan parameter selection information can help the technologist improve workflow, achieve more consistent image quality and reduce patient dose. (C) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:3093 / 3101
页数:9
相关论文
共 50 条
  • [21] Optimizing CT Dose and Image Quality for Different Patient Sizes
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2010, 37 (06) : 3333 - +
  • [22] Optimizing CT Dose and Image Quality for Different Patient Sizes
    Cody, D.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [23] Robustness and Accuracy of An Automated Solution for Measuring Image Noise, Spatial Resolution, Contrast, and Dose in Clinical CT Images: Efforts Towards Patient-Specific Quantifications of CT Image Quality and Radiation Dose
    Ding, A.
    Smith, T.
    Abadi, E.
    Ria, F.
    Wilson, J.
    Samei, E.
    MEDICAL PHYSICS, 2017, 44 (06) : 3023 - 3024
  • [24] Effect of technologist and patient attributes on centering for body CT examinations: Influence of cultural and ethnic factors
    Aly, Antar
    Ebrahimian, Shadi
    Kharita, Mohammed H.
    Heidous, Mahmoud
    Ashruf, Mohammad Zaya
    Kumar, Davendra
    Kalra, Mannudeep K.
    Al Naemi, Huda Mohd
    PLOS ONE, 2022, 17 (08):
  • [25] Automatic patient centering for MDCT: Effect on radiation dose
    Li, Jianhai
    Udayasankar, Unni K.
    Toth, Thomas L.
    Seamans, John
    Small, William C.
    Kalra, Mannudeep K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (02) : 547 - 552
  • [26] A method to extract image noise level from patient images in CT
    Malkus, Annelise
    Szczykutowicz, Timothy P.
    MEDICAL PHYSICS, 2017, 44 (06) : 2173 - 2184
  • [27] Patient-specific predictors of image noise in coronary CT angiography
    Schuhbaeck, Annika
    Schaefer, Marcella
    Marwan, Mohamed
    Gauss, Soeren
    Muschiol, Gerd
    Lell, Michael
    Pflederer, Tobias
    Ropers, Dieter
    Rixe, Johannes
    Hamm, Christian
    Daniel, Werner G.
    Achenbach, Stephan
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2013, 7 (01) : 39 - 45
  • [28] Thorax CT Dose Reduction Based on Patient Features: Effect of Patient Characteristics on Image Quality and Effective Dose
    Koc, Gizem Gul
    Koc, Zafer
    Kaniyev, Tahir
    Kokangul, Ali
    HEALTH PHYSICS, 2019, 116 (05): : 736 - 745
  • [29] Deep Convolutional approach for Low-Dose CT Image Noise Reduction
    Badretale, Seyyedomid
    Shaker, Fariba
    Babyn, Paul
    Alirezaie, Javad
    2017 24TH NATIONAL AND 2ND INTERNATIONAL IRANIAN CONFERENCE ON BIOMEDICAL ENGINEERING (ICBME), 2017, : 142 - 146
  • [30] Fully Convolutional Architecture for Low-Dose CT Image Noise Reduction
    Badretale, S.
    Shaker, F.
    Babyn, P.
    Alirezaie, J.
    2017 INTERNATIONAL CONFERENCE ON ARTIFICIAL INTELLIGENCE APPLICATIONS AND TECHNOLOGIES (AIAAT 2017), 2017, 261