Evaluation of the i-STAT Alinity v in a veterinary clinical setting

被引:3
|
作者
Burke, Jasper E. [1 ]
Nguyen, Thuy Hien T. [2 ]
Davis, Tanya [2 ]
Koenig, Amie [1 ]
Lane, Selena L. [1 ]
Good, Jennifer [1 ]
Brainard, Benjamin M. [1 ]
机构
[1] Univ Georgia, Coll Vet Med, Dept Small Anim Med & Surg, Athens, GA USA
[2] Zoetis Inc, Parsippany, NJ USA
关键词
blood gas; cats; dogs; electrolytes; emergency; point-of-care; BLOOD-GASES; ANALYZER; PH; ELECTROLYTES; GUIDELINES; ACCURACY;
D O I
10.1177/10406387211019710
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Many point-of-care (POC) analyzers are available for the measurement of electrolytes and acid-base status in animals. We assessed the precision of the i-STAT Alinity v, a recently introduced POC analyzer, and compared it to 2 commonly used and previously validated POC analyzers (i-STAT 1, Stat Profile pHOx Ultra). Precision was evaluated by performing multiple analyses of whole blood samples from healthy dogs, cats, and horses on multiple i-STAT Alinity v analyzers. For comparison between analyzers, whole blood samples from dogs and cats presented to the emergency room were run concurrently on all 3 POC instruments. Reported values were compared by species (dogs and cats only) using Pearson correlation, and all values from all species were analyzed together for the Bland-Altman analysis. Results suggested that the i-STAT Alinity v precision was very good, with median coefficients of variability <2.5% for all measured parameters (except the anion gap), with variable ranges of coefficients of variation. In addition, good-to-excellent correlation was observed between the i-STAT Alinity v and i-STAT 1, and between the i-STAT Alinity v and Stat Profile pHOx Ultra for all parameters in both cats and dogs, respectively. In this cohort, the i-STAT Alinity v had clinically acceptable bias compared to the currently marketed analyzers and can be used for monitoring measured analytes in cats and dogs, although serial measurements in a single animal should be performed on the same analyzer whenever possible.
引用
收藏
页码:703 / 710
页数:8
相关论文
共 50 条
  • [41] Precision Evaluation of the pH, PO2 and PCO2 tests in the i-STAT® G3+ Cartridge using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Thibodeau, L.
    Aguirre, M. Gonzalez
    Mohammed, A.
    Kovacevic, S.
    Gupta, M.
    Zhang, H.
    CLINICAL CHEMISTRY, 2024, 70
  • [42] Analytical performance of the i-STAT cardiac troponin I assay.
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    CLINICAL CHEMISTRY, 2004, 50 (06) : A9 - A9
  • [43] Evaluation of a portable clinical analyzer in a veterinary hospital setting
    Grosenbaugh, DA
    Gadawski, JE
    Muir, WW
    JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1998, 213 (05) : 691 - 694
  • [44] EVALUATION AND IMPLEMENTATION OF THE I-STAT SYSTEM IN THE EMERGENCY DEPARTMENT AND NEWBORN INTENSIVE-CARE UNIT
    GOOGINS, MK
    KRAFT, C
    CHANEY, K
    KREYER, J
    NARLOCK, C
    APPLE, F
    CLINICAL CHEMISTRY, 1995, 41 (06) : S187 - S187
  • [45] Evaluation of the blood-saving Kaolin i-STAT ACT technique in paediatric cardiac anaesthesia
    Ulmer, F.
    Bettex, D.
    Baulig, W.
    Buerki, C.
    Weiss, M.
    SWISS MEDICAL WEEKLY, 2008, 138 (41-42) : 11S - 11S
  • [46] Evaluation of portable point of care instrument:: the i-Stat®.: Report of 7,000 analyses
    Feuillu, A
    Morel, I
    Mollard, JF
    ANNALES DE BIOLOGIE CLINIQUE, 2002, 60 (02) : 153 - 164
  • [47] Evaluation of the Abbott i-STAT®1 as a supporting instrument for the Abbott Architect ci8200
    Kerekes, R.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A225 - A225
  • [48] Evaluation of the i-STAT hand-held chemical analyser during treadmill and endurance exercise
    Silverman, S. C.
    Birks, E. K.
    EQUINE VETERINARY JOURNAL, 2002, 34 : 551 - 554
  • [49] Abbott公司收购i-STAT公司附视频
    胡子
    中国医疗器械杂志, 2004, (02) : 127 - 127
  • [50] EVALUATION OF THE I-STAT PORTABLE CLINICAL ANALYZER FOR MEASUREMENT OF IONIZED CALCIUM AND SELECTED BLOOD CHEMISTRY VALUES IN ASIAN ELEPHANTS (ELEPHAS MAXIMUS)
    Tarbert, Danielle K.
    Behling-Kelly, Erica
    Priest, Heather
    Childs-Sanford, Sara
    JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2017, 48 (02) : 319 - 327