Evaluation of the i-STAT Alinity v in a veterinary clinical setting

被引:3
|
作者
Burke, Jasper E. [1 ]
Nguyen, Thuy Hien T. [2 ]
Davis, Tanya [2 ]
Koenig, Amie [1 ]
Lane, Selena L. [1 ]
Good, Jennifer [1 ]
Brainard, Benjamin M. [1 ]
机构
[1] Univ Georgia, Coll Vet Med, Dept Small Anim Med & Surg, Athens, GA USA
[2] Zoetis Inc, Parsippany, NJ USA
关键词
blood gas; cats; dogs; electrolytes; emergency; point-of-care; BLOOD-GASES; ANALYZER; PH; ELECTROLYTES; GUIDELINES; ACCURACY;
D O I
10.1177/10406387211019710
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Many point-of-care (POC) analyzers are available for the measurement of electrolytes and acid-base status in animals. We assessed the precision of the i-STAT Alinity v, a recently introduced POC analyzer, and compared it to 2 commonly used and previously validated POC analyzers (i-STAT 1, Stat Profile pHOx Ultra). Precision was evaluated by performing multiple analyses of whole blood samples from healthy dogs, cats, and horses on multiple i-STAT Alinity v analyzers. For comparison between analyzers, whole blood samples from dogs and cats presented to the emergency room were run concurrently on all 3 POC instruments. Reported values were compared by species (dogs and cats only) using Pearson correlation, and all values from all species were analyzed together for the Bland-Altman analysis. Results suggested that the i-STAT Alinity v precision was very good, with median coefficients of variability <2.5% for all measured parameters (except the anion gap), with variable ranges of coefficients of variation. In addition, good-to-excellent correlation was observed between the i-STAT Alinity v and i-STAT 1, and between the i-STAT Alinity v and Stat Profile pHOx Ultra for all parameters in both cats and dogs, respectively. In this cohort, the i-STAT Alinity v had clinically acceptable bias compared to the currently marketed analyzers and can be used for monitoring measured analytes in cats and dogs, although serial measurements in a single animal should be performed on the same analyzer whenever possible.
引用
收藏
页码:703 / 710
页数:8
相关论文
共 50 条
  • [21] Analysis of Blood Biochemistry of Free Ranging and Human-Managed Southern White Rhinoceros (Ceratotherium simum simum) Using the i-STAT Alinity v®
    Trivedi, Shweta
    Burnham, Christina M.
    Capobianco, Christian M.
    Boshoff, Christiaan
    Zheng, Yaxin
    Pettiglio, Jordan Wood
    Ange-van Heugten, Kimberly
    Bissell, Heidi D.
    Minter, Larry J.
    VETERINARY MEDICINE INTERNATIONAL, 2021, 2021
  • [22] EVALUATION OF BLOOD-GAS AND PH METHODS ON THE I-STAT(TM) PORTABLE CLINICAL ANALYZER
    BOTTJEN, PL
    HAVEN, MC
    CLINICAL CHEMISTRY, 1995, 41 (06) : S196 - S196
  • [23] i-STAT teams up with Abbott
    不详
    BIOSENSORS & BIOELECTRONICS, 1998, 13 (11): : I - I
  • [24] Infringement suit brought by Nova against i-STAT dismissed - i-STAT plans rule 11 action
    不详
    BIOTECHNOLOGY LAW REPORT, 1998, 17 (02): : 298 - 298
  • [25] I-STAT: The Instant Portable Analyzer
    Wynn, Alice
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2012, 138 : A058 - A058
  • [26] Evaluation of the Next Generation i-STAT® Point of Care Prothrombin Time
    Ball, S.
    Hahn, S.
    Kornfeld, S. F.
    CLINICAL CHEMISTRY, 2024, 70
  • [27] ANALYTICAL EVALUATION OF I-STAT PORTABLE CLINICAL ANALYZER AND USE BY NONLABORATORY HEALTH-CARE PROFESSIONALS
    JACOBS, E
    VADASDI, E
    SARKOZI, L
    COLMAN, N
    CLINICAL CHEMISTRY, 1993, 39 (06) : 1069 - 1074
  • [28] Comparison of Hematocrit and Biochemical Analytes among Two Point-of-Care Analyzers (EPOC and i-STAT Alinity v) and a Veterinary Diagnostic Laboratory in the African Savanna Elephant (Loxodonta africana) and the Southern White Rhinoceros (Ceratotherium simum simum)
    Stratton, Hayley S.
    Ange-van Heugten, Kimberly D.
    Minter, Larry J.
    JOURNAL OF ZOOLOGICAL AND BOTANICAL GARDENS, 2022, 3 (04): : 653 - 664
  • [29] Accuracy of the Glucose Test in the i-STAT® G Cartridge Using the i-STAT 1 System for Venous, Arterial and Capillary Whole Blood Specimens in Clinical Settings
    Kornfeld, S. F.
    Aguirre, M. Gonzalez
    Zhang, H.
    Hahn, S.
    CLINICAL CHEMISTRY, 2024, 70
  • [30] Analytical evaluation of i-STAT portable clinical analyzer for blood gas and electrolyte analyses in pediatric samples.
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL CHEMISTRY, 1996, 42 (06) : 198 - 198