Improving the dimensional accuracy of 3D x-ray microscopy data

被引:10
|
作者
Villarraga-Gomez, Herminso [1 ]
Kotwal, Naomi [2 ]
Parwani, Rachna [2 ]
Weiss, Daniel [3 ]
Krenkel, Martin [3 ]
Kimmig, Wolfgang [3 ]
Vom Hagen, Christoph Graf [2 ]
机构
[1] Carl Zeiss Ind Metrol LLC, Wixom, MI 48393 USA
[2] Carl Zeiss Xray Microscopy Inc, Dublin, CA USA
[3] Carl Zeiss Ind Messtech GmbH, Oberkochen, Germany
关键词
x-ray microscopy; 3D x-ray microscopes; computed tomography; dimensional metrology; dimensional accuracy; geometric magnification; optical magnification; COMPUTED-TOMOGRAPHY; CT; RESOLUTION;
D O I
10.1088/1361-6501/ac5a30
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
X-ray microscopy instruments have the unique ability to achieve nondestructive imaging with higher spatial resolutions than traditional x-ray computed tomography (CT) systems. This unique ability is of interest to industrial quality control entities, as they deal with small features in precision manufactured parts (with tolerances in the order of +/- 25 mu m or better). Since many of today's technology and manufacturing companies demand increasingly higher levels of precision, accuracy, and reliability for dimensional measurements on feature sizes that are much smaller than 5 mm, it would be ideal to further expand the imaging capabilities of x-ray microscopy to the field of precision metrology. To address such demand, this paper describes the development of a measurement workflow, through a package consisting of hardware and software, to improve the accuracy of dimensional data obtained with 3D x-ray microscopes (XRMs)-also known as sub-micrometer CT systems. The measurement workflow, called Metrology Extension (MTX), was designed to adjust and configure the XRM instrument work-zone to perform dimensional measurement tasks. The main adjustments of an XRM instrument through the MTX workflow, which must be implemented before scanning parts of interest for dimensional evaluation, include applying a distortion map correction on the image projections produced by the x-ray detector and a voxel scale correction. The main purpose of this article is to present, evaluate, and analyze the experimental results of various measurement tests to verify the metrological performance of several XRM systems operating with the MTX workflow. The main results show that these systems can produce repeatable and reproducible measurements, with repeatability standard deviations of the order of 0.1 mu m and reproducibility standard deviations of less than 0.5 mu m. In addition, coordinate-based 3D XRM measurements produce dimensional accuracies comparable to those offered by high-precision tactile coordinate measurement machines (with deviations within the range of +/- 0.95 mu m). Therefore, once the MTX workflow is executed, XRM instruments can be used to measure small volumes, in the order of (5 mm)(3) or less, with improved dimensional accuracy.
引用
收藏
页数:18
相关论文
共 50 条
  • [21] 3D Analysis of Porosity in a Ceramic Coating Using X-ray Microscopy
    Klement, Uta
    Ekberg, Johanna
    Kelly, Stephen T.
    [J]. JOURNAL OF THERMAL SPRAY TECHNOLOGY, 2017, 26 (03) : 456 - 463
  • [22] 3D elemental sensitive imaging using transmission X-ray microscopy
    Liu, Yijin
    Meirer, Florian
    Wang, Junyue
    Requena, Guillermo
    Williams, Phillip
    Nelson, Johanna
    Mehta, Apurva
    Andrews, Joy C.
    Pianetta, Piero
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 404 (05) : 1297 - 1301
  • [23] 3D Analysis of Porosity in a Ceramic Coating Using X-ray Microscopy
    Uta Klement
    Johanna Ekberg
    Stephen T. Kelly
    [J]. Journal of Thermal Spray Technology, 2017, 26 : 456 - 463
  • [24] Multiscale 3D characterization with dark-field x-ray microscopy
    Simons, Hugh
    Jakobsen, Anders Clemen
    Ahl, Sonja Rosenlund
    Detlefs, Carsten
    Poulsen, Henning Friis
    [J]. MRS BULLETIN, 2016, 41 (06) : 454 - 459
  • [25] 3D elemental sensitive imaging using transmission X-ray microscopy
    Yijin Liu
    Florian Meirer
    Junyue Wang
    Guillermo Requena
    Phillip Williams
    Johanna Nelson
    Apurva Mehta
    Joy C. Andrews
    Piero Pianetta
    [J]. Analytical and Bioanalytical Chemistry, 2012, 404 : 1297 - 1301
  • [26] X-ray Micro-Computed Tomography for Nondestructive Three-Dimensional (3D) X-ray Histology
    Katsamenis, Orestis L.
    Olding, Michael
    Warner, Jane A.
    Chatelett, David S.
    Jones, Mark G.
    Sgalla, Giacomo
    Smit, Bennie
    Larkin, Oliver J.
    Haig, Ian
    Richeldi, Luca
    Sinclair, Ian
    Lackie, Peter M.
    Schneider, Philipp
    [J]. AMERICAN JOURNAL OF PATHOLOGY, 2019, 189 (08): : 1608 - 1620
  • [27] What's inside my USB drive? - X-Ray Microscopy and X-Ray nano CT for 3d packaging
    Oppermann, Martin
    Roth, Holger
    Neubrand, Tobias
    Zerna, Thomas
    [J]. PROCEEDINGS OF THE 2013 IEEE 15TH ELECTRONICS PACKAGING TECHNOLOGY CONFERENCE (EPTC 2013), 2013, : 88 - 92
  • [28] High-resolution 3D X-ray diffraction microscopy: 3D mapping of deformed metal microstructures
    Kutsal, Mustafacan
    Poulsen, Henning Friis
    Winther, Grethe
    Sorensen, Henning Osholm
    Detlefs, Carsten
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2022, 55 (05) : 1125 - 1138
  • [29] 3D X-RAY IMAGE COMPOSITION
    Armeanu, Constantin Catalin
    [J]. UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2016, 78 (01): : 309 - 320
  • [30] Accuracy evaluation of direct navigation with an isocentric 3D rotational X-ray system
    van de Kraats, EB
    van Walsum, T
    Kendrick, L
    Noordhoek, NJ
    Niessen, WJ
    [J]. MEDICAL IMAGE ANALYSIS, 2006, 10 (02) : 113 - 124