A 3D printed phantom to assess MRI geometric distortion

被引:2
|
作者
Ramachandran, Prabhakar [1 ,2 ]
Noble, Christopher [1 ]
Langton, Christian [3 ,4 ]
Perrett, Ben [1 ]
Cox, Joshua [4 ]
Chapman, Mike [4 ]
McGill, George [5 ]
机构
[1] Princess Alexandra Hosp, Dept Radiat Oncol, Woolloongabba, Qld 4102, Australia
[2] Queensland Univ Technol, Sci & Engn Fac, Brisbane, Qld 4000, Australia
[3] Queensland Univ Technol, Fac Hlth, Clin Sci, Kelvin Grove, Qld 4059, Australia
[4] 3D One Australia Pty Ltd, Salisbury, Qld 41076, Australia
[5] Princess Alexandra Hosp, Biomed Tech Serv, Woolloongabba, Qld 4102, Australia
来源
关键词
magnetic resonance imaging; distortion; 3D-printing;
D O I
10.1088/2057-1976/abeb7e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Geometric distortions in magnetic resonance can introduce significant uncertainties into applications such as radiotherapy treatment planning and need to be assessed as part of a comprehensive quality assurance program. We report the design, fabrication, and imaging of a custom 3D printed unibody MR distortion phantom along with quantitative image analysis. Methods: The internal cavity of the phantom is an orthogonal three-dimensional planar lattice, composed of 3 mm diameter rods spaced equidistantly at a 20 mm centre-centre offset repeating along the X, Y, and Z axes. The phantom featured an overall length of 308.5 mm, a width of 246 mm, and a height of 264 mm with lines on the external surface for phantom positioning matched to external lasers. The MR phantom was 3D printed in Nylon-12 using an advancement on traditional selective laser sintering (SLS) (HP Jet Fusion 3D-4200 machine). The phantom was scanned on a Toshiba Aquilion CT scanner to check the integrity of the 3D print and correct for any resultant issues. The phantom was then filled with NiSO4 solution and scanned on a 3T PET-MR Siemens scanner for selected T1 and T2 sequences, from which distortion vectors were generated and analysed using in-house software written in Python. Results: All deviations of the node positions from the print design were less than 1 mm, with an average displacement of 0.228 mm. The majority of the deviations were smaller than the 0.692 mm pixel size for this dataset. Conclusion: A customised 3D printed MRI-phantom was successfully printed and tested for assessing geometric distortion on MRI scanners. 3D printed phantoms can be considered for clinics wishing to assess geometric distortions under specific conditions, but require resources for design, fabrication, commissioning, and verification.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A modular phantom and software to characterize 3D geometric distortion in MRI
    Slagowski, Jordan M.
    Ding, Yao
    Aima, Manik
    Wen, Zhifei
    Fuller, Clifton D.
    Chung, Caroline
    Debnam, J. Matthew
    Hwang, Ken-Pin
    Kadbi, Mo
    Szklaruk, Janio
    Wang, Jihong
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (19):
  • [2] Geometric Distortion Correction of MRI-Linac Phantom with 3d Printing Phantom
    Liu, X.
    Jia, C.
    Li, H.
    Shi, L.
    Lu, W.
    Gong, G.
    Li, Z.
    Yin, Y.
    Qiu, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E141 - E141
  • [3] Compare the 3D Geometric Distortion of MRI, MR-sim and MR-Linac Systems by 3D Printed Phantom
    Liu, X.
    Li, Z.
    Jia, C.
    Yin, Y.
    Qiu, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E259 - E259
  • [4] Geometric distortion in clinical MRI systems - Part I: evaluation using a 3D phantom
    Wang, D
    Strugnell, W
    Cowin, G
    Doddrell, DM
    Slaughter, R
    MAGNETIC RESONANCE IMAGING, 2004, 22 (09) : 1211 - 1221
  • [5] Geometric distortion in clinical MRI systems - Part II: correction using a 3D phantom
    Wang, D
    Strugnell, W
    Cowin, G
    Doddrell, DM
    Slaughter, R
    MAGNETIC RESONANCE IMAGING, 2004, 22 (09) : 1223 - 1232
  • [6] A Comparison of the Distortion in the Same Field MRI and MR-Linac System With a 3D Printed Phantom
    Liu, Xuechun
    Li, Zhenjiang
    Rong, Yi
    Cao, Minsong
    Li, Hongyu
    Jia, Chuntao
    Shi, Liting
    Lu, Weizhao
    Gong, Guanzhong
    Yin, Yong
    Qiu, Jianfeng
    FRONTIERS IN ONCOLOGY, 2021, 11
  • [7] DESIGN OF A NEW 3D GEOMETRIC DISTORTION PHANTOM FOR MR IMAGING
    KESSLER, ML
    CARSON, PL
    CHENEVERT, TL
    TENHAKEN, RK
    RADIOLOGY, 1992, 185 : 223 - 224
  • [8] Assessment of geometric distortion in six clinical scanners using a 3D-printed grid phantom
    Jafar M.
    Jafar Y.M.
    Dean C.
    Miquel M.E.
    Miquel, Marc E. (marc.miquel@bartshealth.nhs.uk), 2017, MDPI (03)
  • [9] Estimation of system-related geometric distortion in 7T MRI using a 3D anthropomorphic head phantom
    Peerlings, J.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S455 - S455
  • [10] Characterization of 3D geometric distortion of MRI scanners commissioned for RT planning
    Torfeh, T.
    Perkins, G.
    Hammoud, R.
    McGarry, M.
    Celik, A.
    Hwang, K.
    Stancanello, J.
    Al-Hammadi, N.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S842 - S842