Risk Factors for Aseptic Loosening After Total Knee Arthroplasty with a Rotating-Hinge Implant A Case-Control Study

被引:19
|
作者
Levent, Ali [1 ,2 ]
Suero, Eduardo M. [1 ,2 ,3 ]
Gehrke, Thorsten [1 ,2 ]
Citak, Mustafa [1 ,2 ]
机构
[1] ENDO Klin Hamburg, Hamburg, Germany
[2] ENDO Klin Hamburg, Dept Orthopaed Surg, Hamburg, Germany
[3] Ludwig Maximilians Univ Munchen, Univ Hosp, Dept Gen Trauma & Reconstruct Surg, Munich, Germany
来源
关键词
FAILURE;
D O I
10.2106/JBJS.20.00788
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Successful results have been reported in association with the use of a rotating-hinge prosthesis for primary total knee arthroplasty (TKA). The objective of the present study was to identify risk factors for aseptic loosening in patients who underwent primary TKA with rotating-hinge knee prostheses. Methods: The records of 1,235 patients who underwent primary TKA with a rotating-hinge prosthesis at our center were evaluated. A total of 125 patients who underwent revision were further evaluated according to the inclusion and exclusion criteria, and 33 patients who underwent revision because of aseptic loosening were then compared with a group of 30 patients who did not require revision surgery. All data, including radiographic measurements, were obtained from records prior to the primary TKA. Results: On the basis of our review of demographic, anthropometric, clinical, surgical, and radiographic findings, we found that higher body mass index (BMI) was associated with revision. The majority of patients with aseptic loosening had loosening of the femoral component. Furthermore, the inner femoral diameter at 20 cm proximal to the knee joint (on both anteroposterior and lateral images) was found to be predictive of revision among those with aseptic loosening. Receiver operating characteristic curve analysis showed that an inner diameter of the femur of >19 mm on anteroposterior images had a sensitivity of 91% and specificity of 87% for predicting the need for revision surgery in patients with aseptic loosening. Conclusions: This is one of few studies that has focused on determining risk factors for the failure of rotating-hinge prostheses following TKA surgery. Our findings indicate that a novel variable, the inner (diaphyseal) diameter of the femur at the point 20 cm proximal to the knee joint, is an extremely reliable predictor of revision surgery in patients with aseptic loosening.
引用
收藏
页码:517 / 523
页数:7
相关论文
共 50 条
  • [11] Letter to editor “Risk factors for aseptic loosening in complex revision total knee arthroplasty using rotating hinge implants”
    Orkhan Aliyev
    Cemil Burak Demirkıran
    Fatih Yıldız
    İbrahim Tuncay
    International Orthopaedics, 2021, 45 : 801 - 802
  • [12] Letter to editor "Risk factors for aseptic loosening in complex revision total knee arthroplasty using rotating hinge implants"
    Aliyev, Orkhan
    Demirkiran, Cemil Burak
    Yildiz, Fatih
    Tuncay, Ibrahim
    INTERNATIONAL ORTHOPAEDICS, 2021, 45 (03) : 801 - 802
  • [13] Aseptic Tibial Implant Loosening After Total Knee Arthroplasty: Preventable?
    Menken, Luke G.
    Fleuriscar, Jean
    Weiner, Travis
    Berliner, Zachary P.
    Rodriguez, Jose A.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2021, 29 (08) : 326 - 330
  • [14] Aseptic loosening after total hip arthroplasty and the risk of cardiovascular disease: A nested case-control study
    Rysinska, Agata
    Skoldenberg, Olof
    Garland, Anne
    Rolfson, Ola
    Aspberg, Sara
    Eisler, Thomas
    Garellick, Goran
    Stark, Andreas
    Hailer, Nils
    Gordon, Max
    PLOS ONE, 2018, 13 (11):
  • [15] Risk Factors for aseptic loosening after primary total knee arthroplasty with cemented knee implants
    Holy, Chantal E.
    Daccach, Juan
    Leopold, John
    Ray, Bidusee
    Coplan, Paul
    Hoeffel, Daniel
    Wright, John
    PHARMACOEPIDEMIOLOGY AND DRUG SAFETY, 2020, 29 : 350 - 350
  • [16] Endo-Modell Rotating-hinge Total Knee for Revision Total Knee Arthroplasty
    Bistolfi, Alessandro
    Rosso, Federica
    Crova, Maurizio
    Massazza, Giuseppe
    ORTHOPEDICS, 2013, 36 (10) : E1299 - E1306
  • [17] Outcomes of Rotating-hinge Total Knee Arthroplasty Following Complex Primary and Revision Total Knee Arthroplasty
    Pfeufer, Daniel
    Gililland, Jeremy
    Anderson, Mike B.
    Monteiro, Priscila
    Peters, Christopher L.
    Stagg, Marissa
    Pelt, Christopher E.
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2021, 38
  • [18] Indications for primary rotating-hinge total knee arthroplasty. Is there consensus?
    Dauwe, Jan
    Vandenneucker, Hilde
    ACTA ORTHOPAEDICA BELGICA, 2018, 84 (03): : 245 - 250
  • [19] Mid-term results of complex primary total knee arthroplasty using a rotating-hinge implant
    Spranz, David M.
    Koch, Kevin-Arno
    Reiner, Tobias
    Hetto, Pit
    Gotterbarm, Tobias
    Merle, Christian
    KNEE, 2022, 34 : 34 - 41
  • [20] Rotating-Hinge Revision Total Knee Arthroplasty for Treatment of Severe Arthrofibrosis
    Bingham, Joshua S.
    Bukowski, Brandon R.
    Wyles, Cody C.
    Pareek, Ayoosh
    Berry, Daniel J.
    Abdel, Matthew P.