High-intensity focused ultrasound (HIFU) ablation by frequency chirp excitation: reduction of the grating lobe in axial focus shifting

被引:9
|
作者
Wang, Mingjun [1 ]
Zhou, Yufeng [1 ]
机构
[1] Nanyang Technol Univ, Sch Mech & Aerosp Engn, 50 Nanyang Ave, Singapore 639798, Singapore
关键词
high-intensity focused ultrasound; phased array transducer; grating lobe; frequency chirp excitation; axial focus shifting; temperature elevation; thermal dose; PHASED-ARRAY TRANSDUCER; MEDICAL ULTRASOUND; TISSUE ABLATION; FIELD; SURGERY; SIGNALS; PATTERN; DESIGN;
D O I
10.1088/1361-6463/aacaed
中图分类号
O59 [应用物理学];
学科分类号
摘要
High-intensity focused ultrasound (HIFU) ablation has gained popularity in tumor treatment. However, the pre-focal heating due to the presence of the grating lobe may generate unintended heating in clinical applications and limit the steering abilities of the phased array design. In order to solve this problem, frequency chirp excitation was proposed and evaluated both numerically and experimentally. First, propagation of an HIFU burst through multi-layer media was simulated in a nonlinear acoustic model using the angular spectrum approach. Influences of axial focus shifting on the acoustic intensity at the both grating and main lobes at different excitation frequencies were investigated. It was found that different excitation frequencies result in variations in the position and magnitude of the grating lobe in the pre-focal region, but have little influence on the main lobe. Thus, the frequency chirp excitation can spread the acoustic energy in the pre-focal region, and subsequently reduce the temperature elevation. The performances of downward and upward frequency chirps in different sweeping times were compared to those of conventional excitation at the constant frequency. Then, HIFU plane ablation was evaluated. Even without the cooling time between treatment spots when using the frequency chirps, the pre-focal heating was still less than that at the constant frequency. Finally, the thermocouple measurement in the gel phantom showed an similar to 40% reduction in pre-focal temperature elevations produced by a single-element concave HIFU transducer using the frequency chirps. Altogether, the frequency chirp excitation could effectively reduce the pre-focal grating lobe at a large axial focus shifting distance, which may enhance both the efficacy and safety of clinical HIFU ablation.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] High-intensity focused ultrasound (HIFU) ablation by the frequency chirps: Enhanced thermal field and cavitation at the focus
    Wang, Mingjun
    Lei, Yisheng
    Zhou, Yufeng
    ULTRASONICS, 2019, 91 : 134 - 149
  • [2] High-Intensity Focused Ultrasound Ablation by the Dual-Frequency Excitation
    Law, Siu Kwun Berlinda
    Zhou, Yufeng
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2019, 66 (01) : 18 - 25
  • [3] High-intensity focused ultrasound (HIFU) ablation in patients with hepatocellular carcinoma
    Jung, S. E.
    Cho, S. H.
    Han, J. Y.
    Hahn, S. T.
    Piao, X.
    JOURNAL OF HEPATOLOGY, 2008, 48 : S147 - S147
  • [4] A Nonlinear Method for High-Intensity Focused Ultrasound (HIFU) Aberration Reduction
    White, P. J.
    von Pattenberg, P.
    Clement, G. T.
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 2059 - 2061
  • [5] Numerical study of high-intensity focused ultrasound (HIFU) in fat reduction
    Mortazavi, Sare
    Mokhtari-Dizaji, Manijhe
    SKIN RESEARCH AND TECHNOLOGY, 2023, 29 (01)
  • [6] Noninvasive cardiac arrhythmia therapy using High-Intensity Focused Ultrasound (HIFU) ablation
    Wu, Qi
    Zhou, Qi
    Zhu, Que
    Rong, Shunkang
    Wang, Qi
    Guo, Rui
    Deng, Changming
    Liu, Dichuan
    Yang, Gang
    Jiang, Yonghong
    Wang, Zhigang
    Lei, Han
    He, Tong-Chuan
    Wang, Zhibiao
    Huang, Jing
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (17) : B224 - B224
  • [7] Influence of High-Intensity Focused Ultrasound (HIFU) Ablation on Arteries: Ex Vivo Studies
    Zhou, Yufeng
    Lim, Wei Chun Daniel
    MICROMACHINES, 2021, 12 (05)
  • [8] Noninvasive cardiac arrhythmia therapy using High-Intensity Focused Ultrasound (HIFU) ablation
    Wu, Qi
    Zhou, Qi
    Zhu, Que
    Rong, Shunkang
    Wang, Qi
    Guo, Rui
    Deng, Changming
    Liu, Dichuan
    Yang, Gang
    Jiang, Yonghong
    Wang, ZhiGang
    Lei, Han
    He, Tong-Chuan
    Wang, ZhiBiao
    Huang, Jing
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 166 (02) : E28 - E30
  • [9] In vivo photoacoustics and high frequency ultrasound imaging of mechanical high intensity focused ultrasound (HIFU) ablation
    Daoudi, Khalid
    Hoogenboom, Martijn
    den Brok, Martijn
    Eikelenboom, Dylan
    Adema, Gosse J.
    Futterer, Jurgen J.
    de Korte, Chris L.
    BIOMEDICAL OPTICS EXPRESS, 2017, 8 (04): : 2235 - 2244
  • [10] High-intensity focused ultrasound (HIFU) therapy for pancreatic cancer
    Asai, Yasutsugu
    Sofuni, Atsushi
    Itoi, Takao
    Tsuchiya, Takayoshi
    Tsuji, Shujiro
    Kamada, Kentaro
    Tanaka, Reina
    Honjyou, Mitsuyoshi
    Mukai, Shuntaro
    Yamamoto, Kenjiro
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2016, 31 : 261 - 261