Cross-calibration of a GE iDXA and Prodigy for Total and Regional Body Bone Parameters: The Importance of Using Cross-calibration Equations for Longitudinal Monitoring After a System Upgrade

被引:9
|
作者
Rhodes, Laura A. [1 ]
Cooper, William [1 ]
Oldroyd, Brian [1 ]
Hind, Karen [2 ]
机构
[1] Univ Leeds, Div Med Phys, Leeds LS2 9NL, W Yorkshire, England
[2] Leeds Metropolitan Univ, Carnegie Res Inst, Leeds LS1 3HE, W Yorkshire, England
关键词
Cross-calibration; iDXA; prodigy; regional total body; X-RAY ABSORPTIOMETRY; MINERAL DENSITY; PRECISION;
D O I
10.1016/j.jocd.2013.09.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We aimed to determine if cross-calibration equations could be applied to convert GE Lunar Prodigy total and regional bone measurements to the GE iDXA model to support longitudinal monitoring of subjects. The cross-calibration group comprised 63 adults (age 45.1 [12.8] yr; body mass index: 25.6 [3.7] kg/m(2)) and the validation group comprised 25 adults (age 40.5 [11.5] yr; body mass index: 25.7 [3.5] kg/m(2)). The parameters reported were total and regional bone mineral density (BMD), bone mineral content, and bone area. There were significant differences between densitometers for all anatomical regions and reported bone parameters (p < 0.0001); iDXA reported lower BMD than the Prodigy apart from the ribs. Linear regression indicated good agreement for all measurements. Bland-Altman analyses indicated significant bias for all measurements and that cross-calibration equations were required. The derived cross-calibration equations were effective in reducing differences between predicted and measured results for each parameter and at each region apart from leg BMD, where the difference remained significant (0.013 g/cm(2); p < 0.05). Our results indicate that cross-calibration is important to maintain comparability of total body-derived regional bone measurements between the Lunar Prodigy and iDXA.
引用
收藏
页码:496 / 504
页数:9
相关论文
共 42 条
  • [21] Development and cross-calibration of hand bone densitometry using two fan beam DXA machines
    Deodhar, A
    Volk, H
    Duyek, N
    Sultany, G
    OSTEOPOROSIS INTERNATIONAL, 2000, 11 : S47 - S47
  • [22] Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers - bone mineral density and body composition studies
    Oldroyd, B
    Smith, AH
    Truscott, JG
    EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2003, 57 (08) : 977 - 987
  • [23] Cross-calibration of GE/Lunar pencil and fan-beam dual energy densitometers—bone mineral density and body composition studies
    B Oldroyd
    A H Smith
    J G Truscott
    European Journal of Clinical Nutrition, 2003, 57 : 977 - 987
  • [24] Cross-Calibration of the Total Ozone Unit (TOU) With the Ozone Monitoring Instrument (OMI) and SBUV/2 for Environmental Applications
    Wang, Weihe
    Flynn, Lawrence
    Zhang, Xingying
    Wang, Yongmei
    Wang, Yingjian
    Jiang, Fang
    Zhang, Yan
    Huang, Fuxiang
    Li, Xiaojing
    Liu, Ruixia
    Zheng, Zhaojun
    Yu, Wei
    Liu, Guoyang
    IEEE TRANSACTIONS ON GEOSCIENCE AND REMOTE SENSING, 2012, 50 (12): : 4943 - 4955
  • [25] Cross-calibration of stroke disability measures: Bayesian analysis of longitudinal ordinal categorical data using negative dependence
    Parmigiani, G
    Ashih, HW
    Samsa, GR
    Duncan, PW
    Lai, SM
    Matchar, DB
    JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 2003, 98 (462) : 273 - 281
  • [26] Cross-calibration of three types of DXA hologic devices using the ESP phantoms in the longitudinal multicentre Strontium Ranelate trial
    Hans, DB
    Zawadynski, S
    Perron, C
    Pittet, N
    Slosman, DO
    BONE, 2003, 32 (05) : S163 - S164
  • [27] Validation of a calibration method using the cross-calibration factor and system planar sensitivity in quantitative single-photon emission computed tomography imaging
    Matsutomo N.
    Matsumoto S.
    Yamamoto T.
    Sato E.
    Radiological Physics and Technology, 2017, 10 (4) : 439 - 445
  • [28] In Situ Cross-Calibration of In-Fiber Bragg Grating and Electrical Resistance Strain Gauges for Structural Monitoring Using an Extensometer
    Grattan, Simon K. T.
    Taylor, Su E.
    Sun, Tong
    Basheer, P. A. Muhammed
    Grattan, Kenneth T. V.
    IEEE SENSORS JOURNAL, 2009, 9 (11) : 1355 - 1360
  • [29] A cross-calibration study using a novel dual X-ray absorptiometry system for bone mineral density measurements with the European Spine Phantom
    Jiang, Yaojun
    Wu, Yan
    Zhang, Yonggao
    Yang, Xiaopeng
    Gao, Jianbo
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (04) : 2119 - 2127
  • [30] Multicenter study of quantitative PET system harmonization using NIST-traceable 68Ge/68Ga cross-calibration kit
    Miwa, Kenta
    Wagatsuma, Kei
    Iimori, Takashi
    Sawada, Koichi
    Kamiya, Takashi
    Sakurai, Minoru
    Miyaji, Noriaki
    Murata, Taisuke
    Sato, Eisuke
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 52 : 98 - 103