Multicenter study of quantitative PET system harmonization using NIST-traceable 68Ge/68Ga cross-calibration kit

被引:9
|
作者
Miwa, Kenta [1 ]
Wagatsuma, Kei [2 ]
Iimori, Takashi [3 ]
Sawada, Koichi [3 ]
Kamiya, Takashi [4 ]
Sakurai, Minoru [5 ]
Miyaji, Noriaki [6 ]
Murata, Taisuke [3 ]
Sato, Eisuke [7 ]
机构
[1] Int Univ Hlth & Welf, Sch Hlth Sci, Dept Radiol Sci, Otawara, Tochigi, Japan
[2] Tokyo Metropolitan Inst Gerontol, Res Team Neuroimaging, Tokyo, Japan
[3] Chiba Univ Hosp, Dept Radiol, Chiba, Japan
[4] Osaka Univ Hosp, Dept Med Technol, Osaka, Japan
[5] Nippon Med Sch, Clin Imaging Ctr Healthcare, Tokyo, Japan
[6] Japanese Fdn Canc Res, Canc Inst Hosp, Dept Nucl Med, Tokyo, Japan
[7] Kyorin Univ, Fac Hlth Sci, Dept Med Radiol Technol, Tokyo, Japan
关键词
NIST; Ge-68/Ga-68; SUV; PET/CT; MAP-EM; Harmonization; Standardization; POSITRON-EMISSION-TOMOGRAPHY; DESIGNATED CANCER CENTERS; CLINICAL-TRIALS; F-18-FDG PET/CT; FDG-PET/CT; IMAGING NETWORK; VERSION; 2.0; QUALIFICATION; PHANTOM; RECONSTRUCTION;
D O I
10.1016/j.ejmp.2018.07.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The present study aimed to define the errors in SUV and demonstrate the feasibility of SUV harmonization among contemporary PET/CT scanners using a novel National Institute of Standards and Technology (NIST)-traceable Ge-68/Ga-68 source as the reference standard. Methods: We used Ge-68/Ga-68 dose calibrator and PET sources made with same batch of Ge-68/Ga-68 embedded in epoxy that is traceable to the NIST standard. Bias in the amount of radioactivity and the radioactive concentrations measured by the dose calibrators and PET/CT scanners, respectively, was determined at five Japanese sites. We adjusted optimal dial setting of the dose calibrators and PET reconstruction parameters to close the actual amount of radioactivity and the radioactive concentration, respectively, of the NIST-traceable Ge-68/Ga-68 sources to harmonize SUV. Errors in SUV before and after harmonization were then calculated at each site. Results: The average bias in the amount of radioactivity and the radioactive concentrations measured by dose calibrator and PET scanner was - 4.94% and - 12.22%, respectively, before, and - 0.14% and - 4.81%, respectively, after harmonization. Corresponding averaged errors in SUV measured under clinical conditions were underestimated by 7.66%, but improved by - 4.70% under optimal conditions. Conclusion: Our proposed method using an NIST-traceable 68Ge/68Ga source identified bias in values obtained using dose calibrators and PET scanners, and reduced SUV variability to within 5% across different models of PET scanners at five sites. Our protocol using a standard source has considerable potential for harmonizing the SUV when contemporary PET scanners are involved in multicenter studies.
引用
收藏
页码:98 / 103
页数:6
相关论文
共 21 条
  • [1] Validation study of new cross calibration method using NIST traceable 68Ge/68Ga source kit
    Wagatsuma, Kei
    Sakata, Muneyuki
    Oda, Keiichi
    Sasaki, Masayuki
    Ishii, Kenji
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [2] Optimal quality control of dose calibrator using a traceable 68Ge/68Ga calibration source in the operating environment
    Miyaji, Noriaki
    Miwa, Kenta
    Wagatsuma, Kei
    Murata, Taisuke
    Umeda, Takuro
    Takiguchi, Tomohiro
    Koizumi, Mitsuru
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [3] A practical method of measuring 68Ge breakthrough from a 68Ge/68Ga generator using a PET scanner
    Stavrinou, M.
    Sherwani, A. G.
    Gridset, L. E.
    Siksjo, M. K.
    Stokke, C.
    Mikalsen, L. T. G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S663 - S664
  • [4] Optimal quality control of dose calibrator using a traceable 68Ge/68Ga calibration source in the operating environment
    Miyaji, Noriaki
    Miwa, Kenta
    Wagatsuma, Kei
    Murata, Taisuke
    Umeda, Takuro
    Takiguchi, Tomohiro
    Koizumi, Mitsuru
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [5] Validation of a Novel Calibration Method Using a Traceable 68Ge/68Ga Point-Like Source On Eleven Types of PET Scanners
    Koyama, S.
    Hasegawa, T.
    Miyatake, H.
    Inoue, Y.
    Kikuchi, K.
    Wagatsuma, K.
    Miyaji, N.
    Watabe, H.
    Shidahara, M.
    Watanuki, S.
    Tsuda, K.
    Muramatsu, Y.
    Yanagisawa, K.
    Wada, Y.
    Oda, K.
    Sato, Y.
    Yamada, T.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [6] Cross-calibration of gamma counters and PET scanners using Ge-68/Ga-68
    Lodge, Martin
    Dahlbom, Magnus
    Wahl, Richard
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [7] Traceable point-like 68Ge/68Ga source with a spherically symmetric positron absorber for PET scanners
    Tomoyuki Hasegawa
    Mio Okamoto
    Takahiro Yamada
    Hidetake Ishizu
    Takahiro Mikamoto
    Yasushi Sato
    Hiroki Miyatake
    Kei Kikuchi
    Yusuke Inoue
    Radiological Physics and Technology, 2020, 13 : 170 - 176
  • [8] Traceable point-like 68Ge/68Ga source with a spherically symmetric positron absorber for PET scanners
    Hasegawa, Tomoyuki
    Okamoto, Mio
    Yamada, Takahiro
    Ishizu, Hidetake
    Mikamoto, Takahiro
    Sato, Yasushi
    Miyatake, Hiroki
    Kikuchi, Kei
    Inoue, Yusuke
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2020, 13 (02) : 170 - 176
  • [9] Clinical experience with indigenous kit-based preparation of 68Ga-DOTATOC using commercial 68Ge/68Ga generator
    Pandey, Usha
    Korde, Aruna
    Mukherjee, Archana
    Shinto, Ajit
    Kamaleswaran, K. K.
    Jose, Raghi P.
    Gamre, Naresh
    Dash, Ashutosh
    APPLIED RADIATION AND ISOTOPES, 2018, 136 : 59 - 64
  • [10] Novel 68Ge/68Ga generator system using polysaccharide polymers containing N-methylglucamine groups as adsorbents for 68Ge
    Fuchigami, Takeshi
    Ono, Hokuto
    Oyadomari, Kohta
    Iwatake, Mayumi
    Harigami, Syunpei
    Haratake, Mamoru
    Nakayama, Morio
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55