Performance evaluation of the Inveon PET Scanner using GATE based on the NEMA NU-4 Standards

被引:0
|
作者
Lu, Lijun [1 ]
Karakatsanis, Nicolas
Ma, Jianhua [1 ]
Bian, Zhaoying [1 ]
Han, Yanjiang
Tang, Jing
Rahmim, Arman
Chen, Wufan [1 ]
机构
[1] Southern Med Univ, Sch Biomed Engn, Guangzhou, Guangdong, Peoples R China
基金
中国国家自然科学基金; 美国国家科学基金会;
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The purpose of this study is to validate the application of GATE (Geant4 Application for Tomographic Emission) Monte Carlo simulation toolkit to model the performance characteristics of Siemens Inveon small animal PET system. GATE has been utilized to accurately create a complete PET imaging system model, including system geometry, physics process, signal processing and time-dependent effects. The GATE output data were histogrammed into 3D sinograms and subsequently processed by the Software of Tomography Reconstruction (STIR) to reconstruct 3D PET images. The simulation results were validated against experimental data in accordance with the NEMA NU-4 2008 protocol for standardized evaluation of spatial resolution, sensitivity, scatter fraction (SF) and noise equivalent counting rate (NECR) of a preclinical PET system. An agreement of less than 3% was obtained between the spatial resolution of the simulated and experimental results. The simulated peak NECR of mouse-size phantom agreed with the experimental result, while for the rat-size phantom simulated value was higher than experimental result. The simulated and experimental SFs of mouse- and rat- size phantom both reach an agreement of less than 2%. Therefore, it has been shown the feasibility of our GATE model to accurately simulate, within certain limits, all major performance characteristics of Inveon PET system. In future, the validated model can be utilized to validate, assess and further develop reconstruction algorithms and PET acquisition protocols.
引用
收藏
页数:4
相关论文
共 50 条
  • [41] Performance evaluation of a novel high performance pinhole array detector module using NEMA NU-4 image quality phantom for four head SPECT Imaging
    Rahman, Tasneem
    Tahtali, Murat
    Pickering, Mark R.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [42] Performance measurements for the PET/CT Discovery-600 using NEMA NU 2-2007 standards
    De Ponti, E.
    Morzenti, S.
    Guerra, L.
    Pasquali, C.
    Arosio, M.
    Bettinardi, V.
    Crespi, A.
    Gilardi, M. C.
    Messa, C.
    MEDICAL PHYSICS, 2011, 38 (02) : 968 - 974
  • [43] NEMA NU 4-Optimized Reconstructions for Therapy Assessment in Cancer Research with the Inveon Small Animal PET/CT System
    Lasnon, Charline
    Dugue, Audrey Emmanuelle
    Briand, Melanie
    Blanc-Fournier, Cecile
    Dutoit, Soizic
    Louis, Marie-helene
    Aide, Nicolas
    MOLECULAR IMAGING AND BIOLOGY, 2015, 17 (03) : 403 - 412
  • [44] Performance evaluation of a whole-body PET-CT scanner according the NEMA NU 2-2001 standard
    Tosi, G.
    Leonardi, L.
    Chiti, A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S261 - S261
  • [45] NEMA NU 2-2001 performance testing of a Philips Gemini GXL PET/CT scanner
    Chithradevi Sathiakumar
    Seu Som
    Stefan Eberl
    Peter Lin
    Australasian Physical & Engineering Sciences in Medicine, 2010, 33 : 199 - 209
  • [46] Performance evaluation of two quantitative gamma cameras using NEMA NU-01 2018 standards
    Prasad, Rameshwar
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [47] NEMA NU 4-Optimized Reconstructions for Therapy Assessment in Cancer Research with the Inveon Small Animal PET/CT System
    Charline Lasnon
    Audrey Emmanuelle Dugue
    Mélanie Briand
    Cécile Blanc-Fournier
    Soizic Dutoit
    Marie-hélène Louis
    Nicolas Aide
    Molecular Imaging and Biology, 2015, 17 : 403 - 412
  • [48] Optimisation of I-124 acquisition parameters with the NEMA NU 4-2008 performances assessment for the Inveon PET system
    Anizan, N.
    Carlier, T.
    Hindorf, C.
    Bardies, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S334 - S334
  • [49] Comparative performance measurement of 124I and 18F PET on Siemens Inveon PET using NEMA NU4-2008: Toward optimal parameters in 124I PET
    Yu, A. Ram
    Kim, Jin Su
    Kim, Kyeong Min
    Lee, Young Sub
    Kim, Jong Guk
    Park, JiAe
    Woo, Sang-Keun
    Kim, Hee-Joung
    Cheon, Gi Jeong
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [50] NEMA NU 2-2001 performance testing of a Philips Gemini GXL PET/CT scanner
    Sathiakumar, Chithradevi
    Som, Seu
    Eberl, Stefan
    Lin, Peter
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (02) : 199 - 209