Effects of Motion Artifact on the Blood Oxygen Saturation Estimate in Pulse Oximetry

被引:0
|
作者
Clarke, Geoffrey W. J. [1 ]
Chan, Adrian D. C. [1 ]
Adler, Andy [1 ]
机构
[1] Carleton Univ, Dept Syst & Comp Engn, Ottawa, ON K1S 5B6, Canada
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oxygen saturation estimates from pulse oximeters (SpO(2)) have been shown to be unreliable in the presence of motion artifact. This may cause errors in the clinical environment if the device falsely detects normal or desaturated conditions. This paper seeks to investigate the failure modes of the standard SpO(2) calculation algorithm in the presence of motion artifact. A Texas Instruments AFE4400 evaluation module was used to collect data. The board is designed for pulse oximetry applications and allows access to the raw photoplethysmograph signals. Measurements were taken from a single subject with a finger probe. Signals were collected both while moving the instrumented hand and while moving the sensor without moving the hand. These were compared to a control signal where the subject remained motionless. Oxygen saturation was constant as verified by a Clevemed Bioradio SpO(2) probe on the subject's other hand, which remained motionless for all of the measurements. The results showed a significant decrease of measured SpO(2) during motion of the hand but not during motion of the sensor. This was likely due to the probe detecting the movement of venous blood, or failure to correctly detect peaks in the PPG signals. The variability of the measured SpO(2) increased during motion of the hand and motion of the sensor, likely due to variation of the optical path length through the tissue. This work will help future development of algorithms to improve the performance of pulse oximetry in ambulatory conditions.
引用
收藏
页码:275 / 278
页数:4
相关论文
共 50 条
  • [1] Quantifying Blood-Oxygen Saturation Measurement Error in Motion Contaminated Pulse Oximetry Signals
    Clarke, G. W. J.
    Chan, A. D. C.
    Adler, A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 986 - 989
  • [2] Blood haemoglobin and pulse oximetry oxygen saturation measurement error
    Crooks, C.
    West, J.
    Morling, J.
    Simmonds, M.
    Juurlink, I.
    Briggs, S.
    Cruickshank, S.
    Hammond-Pears, S.
    Shaw, D.
    Card, T.
    Fogarty, A.
    EUROPEAN RESPIRATORY JOURNAL, 2022, 60
  • [3] Motion Artifact Mitigation for Wearable Pulse Oximetry
    Williamson, James R.
    Patel, Tejash
    Singh, Ninoshka
    Siegel, Andrew
    Telfer, Brian
    Trebicka, Ray
    Welsh, Brendon
    Hoyt, Reed
    2018 IEEE 15TH INTERNATIONAL CONFERENCE ON BIOMEDICAL AND HEALTH INFORMATICS (BHI) AND THE WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS (BSN), 2018, : 74 - 77
  • [4] PULSE OXIMETRY AND MOTION ARTIFACT - A STUDY OF ACCURACY AND RELIABILITY
    BARKER, SJ
    SHAH, NK
    HYATT, J
    ANESTHESIOLOGY, 1995, 83 (3A) : A450 - A450
  • [5] AN UNUSUAL CAUSE OF MOTION ARTIFACT DURING PULSE OXIMETRY
    CONE, AM
    SANSOME, AJ
    ANAESTHESIA, 1992, 47 (10) : 917 - 917
  • [6] Design of filter to reject motion artifact of pulse oximetry
    Lee, J
    Jung, W
    Kang, I
    Kim, Y
    Lee, G
    COMPUTER STANDARDS & INTERFACES, 2004, 26 (03) : 241 - 249
  • [7] Pulse Oximetry Overestimates Oxygen Saturation in COPD
    Amalakanti, Sridhar
    Pentakota, Mohan Rao
    RESPIRATORY CARE, 2016, 61 (04) : 423 - 427
  • [8] Oxygen Saturation in Pulse Oximetry in Hemoglobin Anomalies
    Zur, B.
    Bagci, S.
    Ludwig, M.
    Stoffel-Wagner, B.
    KLINISCHE PADIATRIE, 2012, 224 (04): : 259 - 265
  • [9] Effects of henna dye on oxygen saturation reading using pulse oximetry
    Samman, YS
    Rahimi, JS
    Wali, SO
    Krayem, AB
    Abdelaziz, MM
    SAUDI MEDICAL JOURNAL, 2006, 27 (02) : 268 - 269
  • [10] Investigating Motion Artifact in Reflective Pulse Oximetry During Random Motion
    Nawn, Corinne
    Warren, Kristen
    2014 40TH ANNUAL NORTHEAST BIOENGINEERING CONFERENCE (NEBEC), 2014,