Clinical Evaluation of MR-Gated Respiratory Motion Correction in Simultaneous PET/MRI

被引:1
|
作者
Li, Hui [1 ]
Meng, Xiangxi [1 ]
Guan, Xiangping [1 ]
Zhou, Nina [1 ]
Liu, Hui [2 ]
Zhang, Yan [1 ]
Yu, Boqi [1 ]
Zhu, Hua [1 ]
Li, Nan [1 ]
Yang, Zhi [1 ]
机构
[1] Peking Univ Canc Hosp & Inst, Key Lab Carcinogenesis & Translat Res, Minist Educ Beijing, Dept Nucl Med, 52 Fucheng Rd, Beijing 100142, Peoples R China
[2] United Imaging Healthcare, Shanghai, Peoples R China
关键词
PET; MRI; motion correction; lesion detection; lesion quantification; CARDIAC MOTION; PET;
D O I
10.1097/RLU.0000000000003510
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The recently available gated T1-weighted imaging with the Dixon technique enables the synchronized gating signal for both MR acquisition and PET reconstruction. Herein, we evaluated the clinical value of this MR-gated PET reconstruction in the thoracic-abdominal PET/MRI compared with non-MR-gated method. Methods Twenty patients (28 hypermetabolic target lesions) underwent PET/MRI. Four types of PET images were reconstructed: non-MR-gating + gated attenuation correction (AC) (group A), MR-gating + gated AC (group B), non-MR-gating + breath-hold (BH) AC (group C), and MR-gating + BH AC (group D). A 4-point objective scale (from well match to obvious mismatch was scored from 3 to 0) was proposed to evaluate the mismatch. The detection rate and quantitative metrics were also evaluated. Results In the patient-based analysis, for groups A through D, the detection rates were 90%, 100%, 85%, and 90% as well as 95%, 100%, 85%, and 85%, assessed by readers 1 and 2, respectively, and significant difference of mismatch score was observed with the highest proportion of 3 points in group B (85%, 90%, 35%, and 40%, and 80%, 90%, 35%, and 20%, assessed by readers 1 and 2, respectively). The lesion-based analysis demonstrated significant differences in quantitative metrics for groups A through D (all P's < 0.05), with the highest quantitative metrics in group B (SUVmax: 7.49 +/- 3.37, 8.45 +/- 3.82, 6.90 +/- 3.24, and 7.69 +/- 3.50; SUVmean: 3.90 +/- 1.60, 4.34 +/- 1.84, 3.67 +/- 1.61, and 4.03 +/- 1.81; SUVpeak: 5.60 +/- 2.50, 6.10 +/- 2.80, 5.22 +/- 2.40, and 5.65 +/- 2.68; signal-to-noise ratio: 136.06 +/- 90.58, 136.24 +/- 81.63, 99.52 +/- 53.16, and 107.57 +/- 69.05). Conclusions The MR-gated reconstruction using gated AC reduced the mismatch between MR and PET images and improved the thoracic-abdominal PET image quality in simultaneous PET/MRI systems.
引用
下载
收藏
页码:297 / 302
页数:6
相关论文
共 50 条
  • [31] Assessment of motion correction strategies in simultaneous PET-MR
    Chun, Se Young
    Guerin, Bastien
    Cho, Sanghee
    Reese, Timothy
    Zhu, Xuping
    Ouyang, Jinsong
    Catana, Ciprian
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [32] Fast tagged MR acquisition for PET motion correction in simultaneous cardiac PET-MR
    Huang, Chuan
    Petibon, Yoann
    Ouyang, Jinsong
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [33] Comparison of MR and PET derived respiratory motion estimates for dual respiratory and cardiac motion correction in cardiac PET
    Feng, Tao
    Ahlman, Mark
    Kolbitsch, Christoph
    Guo, Liheng
    Tsui, Benjamin
    Bluemke, David
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [34] MR-GATED INTRACRANIAL CSF DYNAMICS - EVALUATION OF CSF PULSATILE FLOW
    NJEMANZE, PC
    BECK, OJ
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1989, 10 (01) : 77 - 80
  • [35] Development of a motion correction system for respiratory-gated PET study
    Woo, SK
    Choi, JY
    Song, TY
    Choi, Y
    Lee, KH
    Kim, BT
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2366 - 2369
  • [36] Correction of Respiratory Motion in Dual Gated Cardiac Imaging in PET/CT
    Lamare, F.
    Teras, M.
    Kokki, T.
    Fayad, H.
    Rimoldi, O.
    Camici, P. G.
    Knuuti, J.
    Visvikis, D.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4530 - +
  • [37] Respiratory compensated PET with DOTATOC PET/MRI using motion registered and gated reconstruction
    Tao, Dora
    Pampaloni, Miguel Hernandez
    Ravanfar, Vahid
    Nakakura, Eric
    Bergsland, Emily
    Hope, Thomas
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [38] Respiratory motion correction for enhanced quantification of hepatic lesions in simultaneous PET and DCE-MR imaging
    Ippoliti, Matteo
    Lukas, Mathias
    Brenner, Winfried
    Schatka, Imke
    Furth, Christian
    Schaeffter, Tobias
    Makowski, Marcus R.
    Kolbitsch, Christoph
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (09):
  • [39] Clinical evaluation of a local respiratory motion correction (LRMC) for PET/CT imaging
    Lamare, Frederic
    Rimoldi, Ornella
    Bettinardi, Valentino
    Camici, Paolo
    Visvikis, Dimitris
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [40] Simultaneous PET–MR acquisition and MR-derived motion fields for correction of non-rigid motion in PET
    Charalampos Tsoumpas
    Jane E. Mackewn
    Philip Halsted
    Andrew P. King
    Christian Buerger
    John J. Totman
    Tobias Schaeffter
    Paul K. Marsden
    Annals of Nuclear Medicine, 2010, 24 : 745 - 750