Implementation of in-vivo diode dosimetry for intensity modulated radiotherapy as routine patients' quality assurance

被引:0
|
作者
AL-Shareef, Jamal M. [1 ,2 ,3 ]
Attalla, Ehab M. [1 ]
El-Gebaly, Reem H. [2 ]
Deiab, Nashaat A. [1 ]
Abdelmajeed, M. [1 ]
Fathy, Mohamed M. [2 ]
机构
[1] Cairo Univ, Natl Canc Inst, Dept Radiotherapy & Nucl Med, Giza, Egypt
[2] Cairo Univ, Fac Sci, Dept Biophys, Giza, Egypt
[3] Sanaa Univ, Dept Phys, Sanaa, Yemen
关键词
In-vivo dosimetry; In-vitro dosimetry; Diode; IMRT; QA; EXIT DOSE MEASUREMENTS; MOSFET DETECTORS; ENTRANCE; ERRORS;
D O I
10.1016/j.radphyschem.2021.109564
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The implementation of an in-vivo diode for dose verification of the intensity-modulated radiotherapy (IMRT) during the actual treatment being rare. This work aims to implement in-vivo and in-vitro diode dosimetry for the IMRT technique to verify the accuracy of the dose delivered to the patients undergoing radiation treatment. In total, 279 fields have been investigated for dose verification from 33 cases, including head and neck, prostate, bladder, and breast cancer patients. Additionally, these fields were compared with in-vitro measurements on a perspex phantom. Beforehand, the measurements were tested for an Alderson Rando phantom. In-vivo measurements revealed that 87.1% of the measured doses were within +/- 5% of the calculated doses. Of the other 12.9%, about 11.1% were more than +/- 5% and less +/- 10% and only 1.8% were outside the +/- 10% of the calculated doses. The corresponding fields for in-vitro measurements show that 96.1% of the measured doses within +/- 5% and only 3.9% were outside the +/- 5% of the calculated doses. The results in this study indicated that in-vivo diode dosimetry for IMRT is reliable as a quality assurance (QA) tool in the radiotherapy departments. Moreover, measurements with diodes are a real-time readout during actual treatment compared to the verification tools that have been used for routine IMRT QA. Additionally, the measurements were carried out comfortably and well tolerated by the patients.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Implementation of in-vivo diode dosimetry for intensity modulated radiotherapy as routine patients' quality assurance
    AL-Shareef, Jamal M.
    Attalla, Ehab M.
    El-Gebaly, Reem H.
    Deiab, Nashaat A.
    Abdelmajeed, M.
    Fathy, Mohamed M.
    [J]. Radiation Physics and Chemistry, 2021, 187
  • [2] In vivo diode dosimetry for routine quality assurance in IMRT
    Higgins, PD
    Alaei, P
    Gerbi, BJ
    Dusenbery, KE
    [J]. MEDICAL PHYSICS, 2003, 30 (12) : 3118 - 3123
  • [3] In vivo diode dosimetry for routine quality assurance in IMRT
    Higgins, P
    Alaei, P
    Gerbi, B
    Dusenbery, K
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1500 - 1501
  • [4] In vivo diode dosimetry in standard and intensity modulated radiotherapy
    Marinello, G
    Mège, JP
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S15 - S15
  • [5] The use of diode in in-vivo dosimetry quality assurance in IMRT
    Kim, H
    Wang, Z
    Lalonde, R
    Sontag, M
    Chen, H
    Li, F
    Smith, R
    Huq, M
    Heron, D
    Yue, N
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1988 - 1989
  • [6] A practical approach to diode based in vivo dosimetry for intensity modulated radiotherapy
    Kadesjo, Nils
    Nyholm, Tufye
    Olofsson, Jorgen
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 98 (03) : 378 - 381
  • [7] Diode dosimetry for the routine quality assurance in IMRT
    Xu, Z.
    Yang, G.
    Li, W.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2437 - 2437
  • [8] A monitor unit verification calculation in intensity modulated radiotherapy as a dosimetry quality assurance
    Kung, JH
    Chen, GTY
    Kuchnir, FK
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2226 - 2230
  • [9] Implementation of in vivo dosimetry using diodes as part of quality assurance in radiotherapy
    Villegas, L. I.
    Espinoza, W. P.
    Bayas, M. V.
    [J]. IX LATIN AMERICAN SYMPOSIUM ON NUCLEAR PHYSICS AND APPLICATIONS, 2012, 1423
  • [10] A deformable head and neck phantom with in-vivo dosimetry for adaptive radiotherapy quality assurance
    Graves, Yan Jiang
    Smith, Arthur-Allen
    Mcilvena, David
    Manilay, Zherrina
    Lai, Yuet Kong
    Rice, Roger
    Mell, Loren
    Jia, Xun
    Jiang, Steve B.
    Cervino, Laura
    [J]. MEDICAL PHYSICS, 2015, 42 (04) : 1490 - 1497