Preparation and evaluation of a lubricious treated cartridge used for implantation of intraocular lenses

被引:1
|
作者
Hu, Can B. [1 ]
Gwon, Arlene [1 ]
Lowery, Mike [1 ]
Makker, Harish [1 ]
Gruber, Lawrence [1 ]
机构
[1] Adv Med Opt, Santa Ana, CA 92705 USA
关键词
lubricious coating; intraocular lenses insertion system; polymer interpenetrating network (IPN); rabbit in vivo study; surface modification;
D O I
10.1163/156856207779116720
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the last decades, many surface modification technologies have been developed in an attempt to improve the function of medical device surfaces by adding or enhancing surface characteristics. These value-added processes included treatment to affect lubricity, hemocompatibility and drug delivery. A unique hydrophilic, lubricious coating was developed to treat hydrophobic polymer surfaces. The coating platforms described are composed of a polyelectrolyte molecular film containing hydrophilic, lubricant molecules. The molecular film is then further cross-linked wish di-functional aldehyde molecules to form an interpenetrating network (IPN). The IPN entraps lubricant molecules in the matrix and provides for prolonged stability of the lubricity. This coating was applied to cartridges which were used to deliver intraocular lenses (IOLs) that replaced the calaractous crystalline lenses in patients In order to determine the safety and effectiveness of the coating, a rabbit in vivo study was designed to evaluate the ease of implantation and postoperative response to implantation of the foldable acrylic IOLs. The performance evaluation of the lubricious treated cartridges focused on the ease of insertion and post-IOL implantation response. It was found that the UNFOLDER (TM) Emerald Insertion System (Advanced Medical Optics) with lubricious treated cartridges generally required lower insertion forces than the standard UNFOLDER (TM) Emerald cartridges. The postoperative inflammatory response following lens extraction and posterior chamber implantation of low (6D), medium (20D) and high (30D) diopter foldable acrylic IOLs with both treated and standard cartridges was mild. Inflammation generally resolved by 3 weeks. Thus, in this animal study, the coating was shown to be effective in assisting the delivery of IOLs through cartridges, without causing any adverse effects.
引用
收藏
页码:179 / 191
页数:13
相关论文
共 50 条
  • [21] Bag-in-the-lens implantation of intraocular lenses
    Tassignon, MJBR
    De Groot, V
    Vrensen, GFJM
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (07): : 1182 - 1188
  • [22] Dissatisfaction after implantation of multifocal intraocular lenses
    Bonnet, Clemence
    Brezin, Antoine P.
    Monnet, Dominique
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [23] Removal of epikeratophakia lenticules and implantation of intraocular lenses
    Bleckmann, H
    Schnoy, H
    Keuch, R
    OPHTHALMOLOGE, 2004, 101 (03): : 285 - 289
  • [24] IMPLANTATION OF SUTURED POSTERIOR CHAMBER INTRAOCULAR LENSES
    MCGUINNESS, R
    KENNEDY, I
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1995, 21 (04): : 373 - 375
  • [25] Implantation of glued intraocular lenses in eyes with microcornea
    Kumar, Dhivya Ashok
    Agarwal, Amar
    Sivangnanam, Soundari
    Chandrasekar, Radika
    Agarwal, Athiya
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (02): : 327 - 333
  • [26] First experiences with implantation of aspheric intraocular lenses
    Werner, W
    Roth, E
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S297 - S297
  • [27] Results of the AcrySof Toric intraocular lenses implantation
    Swiatek, Bozena
    Michalska-Malecka, Katarzyna
    Dorecka, Mariola
    Romaniuk, Dorota
    Romaniuk, Wanda
    MEDICAL SCIENCE MONITOR, 2012, 18 (01): : PI1 - PI4
  • [28] Dissatisfaction after implantation of multifocal intraocular lenses
    de Vries, Niels E.
    Webers, Carroll A. B.
    Touwslager, Wouter R. H.
    Bauer, Noel J. C.
    de Brabander, John
    Berendschot, Tos T.
    Nuijts, Rudy M. M. A.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (05): : 859 - 865
  • [30] Comparison of visual function following implantation of Acrysof Natural intraocular lenses with conventional intraocular lenses
    Landers, John
    Tan, Teng-Han
    Yuen, Joshua
    Liu, Henry
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2007, 35 (02): : 152 - 159