Managing Uncertainty in Complex Interventional Fluoroscopic Procedures

被引:0
|
作者
Leuenberger, Ronald [1 ]
Meade, Jason A. [1 ]
机构
[1] Louis Stokes Cleveland VA Med Ctr, Cleveland, OH 44106 USA
来源
HEALTH PHYSICS | 2014年 / 107卷 / 05期
关键词
operational topics; exposure; radiation; medical radiation; safety standards; RADIATION; SKIN;
D O I
10.1097/HP.0000000000000174
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
There was a concern for potential patient skin burns indicated by air kerma as recorded by fluoroscopy equipment during two interventional fluoroscopic procedures. Two sentinel events as defined by The Joint Commission were indicated. Dose reconstruction based on measurements of machine output showed the air kerma readings were high by approximately a factor of four and no patient injury or sentinel event occurred. The radiation safety program in effect at the time of the incidents allowed discovery of equipment problems before a serious patient injury occurred, but additional controls have been implemented as a result, including quality control checks performed by cardiology, additional training, a database for tracking equipment maintenance, and mandatory entry of patient dose into electronic health records with follow-up orders automatically generated by the electronic health records. Most unexpected injuries are a result of multiple failures, but there are also multiple opportunities to prevent the injury. This paper provides a comprehensive look at patient safety concerns during two interventional cardiology fluoroscopic procedures and offers ways to further reduce risks to patients. It focuses on lessons learned and a systems-based approach to improving and promoting radiation safety during complex interventional fluoroscopic procedures.
引用
收藏
页码:S181 / S187
页数:7
相关论文
共 50 条
  • [21] Radiation exposure during standard and complex interventional procedures
    Vehmas, T
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 : 296 - 298
  • [22] Investigation of skin reactions in complex interventional radiology procedures
    Kulkarni, A. R.
    Akhilesh, P.
    Mahalakshmi, S.
    Sharma, S. D.
    RADIOPROTECTION, 2019, 54 (01) : 61 - 65
  • [23] Approaches to interventional fluoroscopic dose curves
    Wunderle, Kevin A.
    Rakowski, Joseph T.
    Dong, Frank F.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 342 - 352
  • [24] Lateral Head Shielding for Eye Lens Dose Reduction for Neuro-CBCT and Interventional Fluoroscopic Procedures
    Xiong, Z.
    Vijayan, S.
    Guo, C.
    Choudhary, P.
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [25] Using a convolutional neural network for human recognition in a staff dose management software for fluoroscopic interventional procedures
    Troville, J.
    Dhonde, R. S.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [26] Improved-Resolution, Real-Time Skin-Dose Mapping for Interventional Fluoroscopic Procedures
    Rana, Vijay K.
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [27] Fluoroscopic Procedures Performed Teleradiologically
    Berlin, Leonard
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (04) : W777 - W777
  • [28] RADIANT WARMER FOR FLUOROSCOPIC PROCEDURES
    SLOVIS, TL
    POLAND, RL
    RADIOLOGY, 1980, 134 (02) : 540 - 541
  • [29] Interventional procedures
    Holm, HH
    Skjoldbye, B
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 : S131 - S134
  • [30] Interventional procedures
    Wassef, W
    O'Keefe, S
    CURRENT OPINION IN GASTROENTEROLOGY, 2000, 16 (06) : 508 - 515