Managing Uncertainty in Complex Interventional Fluoroscopic Procedures

被引:0
|
作者
Leuenberger, Ronald [1 ]
Meade, Jason A. [1 ]
机构
[1] Louis Stokes Cleveland VA Med Ctr, Cleveland, OH 44106 USA
来源
HEALTH PHYSICS | 2014年 / 107卷 / 05期
关键词
operational topics; exposure; radiation; medical radiation; safety standards; RADIATION; SKIN;
D O I
10.1097/HP.0000000000000174
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
There was a concern for potential patient skin burns indicated by air kerma as recorded by fluoroscopy equipment during two interventional fluoroscopic procedures. Two sentinel events as defined by The Joint Commission were indicated. Dose reconstruction based on measurements of machine output showed the air kerma readings were high by approximately a factor of four and no patient injury or sentinel event occurred. The radiation safety program in effect at the time of the incidents allowed discovery of equipment problems before a serious patient injury occurred, but additional controls have been implemented as a result, including quality control checks performed by cardiology, additional training, a database for tracking equipment maintenance, and mandatory entry of patient dose into electronic health records with follow-up orders automatically generated by the electronic health records. Most unexpected injuries are a result of multiple failures, but there are also multiple opportunities to prevent the injury. This paper provides a comprehensive look at patient safety concerns during two interventional cardiology fluoroscopic procedures and offers ways to further reduce risks to patients. It focuses on lessons learned and a systems-based approach to improving and promoting radiation safety during complex interventional fluoroscopic procedures.
引用
收藏
页码:S181 / S187
页数:7
相关论文
共 50 条
  • [31] PLANNING AND MANAGING NEW AND COMPLEX MEDICAL AND SURGICAL PROCEDURES
    LONG, JM
    FLANIGAN, WJ
    HARA, M
    LEVY, GC
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1966, 196 (11): : 979 - &
  • [32] Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures
    Bednarek, Daniel R.
    Barbarits, Jeffery
    Rana, Vijay K.
    Nagaraja, Srikanta P.
    Josan, Madhur S.
    Rudin, Stephen
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [33] Development of a real-time scattered radiation display for staff dose reduction during fluoroscopic interventional procedures
    Troville, J.
    Kilian-Meneghin, J.
    Guo, C.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [34] A Study to Compare Operator Dosimeter Exposure Versus Fluoroscopic Time and Air Kerma in Fluoroscopically Guided Interventional Procedures
    McKetty, M.
    MEDICAL PHYSICS, 2012, 39 (06) : 3633 - 3633
  • [35] Complex pancreatic surgery: The role of interventional radiologists in managing patients and complications
    Sohn, TA
    Yeo, CJ
    Cameron, JL
    Lillemoe, KD
    Geschwind, JF
    Mitchell, SE
    Venbrux, AC
    GASTROENTEROLOGY, 2002, 123 (01) : 40 - 40
  • [36] Radiation dose to multidisciplinary staff members during complex interventional procedures
    Mussmann, B.
    Larsen, T. R.
    Godballe, M.
    Abdi, A. J.
    Kantso, A.
    Jakobsen, A. R.
    Nielsen, M. V.
    Jensen, J.
    RADIOGRAPHY, 2024, 30 (02) : 512 - 516
  • [37] DETERMINATION OF RADIATION DOSE DURING FLUOROSCOPIC PROCEDURES WITH THE AID OF THE FLUOROSCOPIC SIMULATOR
    MILLER, WB
    ROHRER, RH
    WEENS, HS
    RADIOLOGY, 1962, 78 (02) : 237 - 239
  • [38] Assessment of Maximum Patient Skin Dose for Interventional Radiology Procedures: Method and Uncertainty Analysis
    Farah, J.
    Trianni, A.
    Carinou, E.
    Dabin, J.
    Deangelis, C.
    Domienik, J.
    Knezevic, Z.
    Kopec, R.
    Malchair, F.
    Negri, A.
    Novak, L.
    Siiskonen, T.
    Vanhavere, F.
    Clairand, I.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [39] Why Is the X-Ray Tube Usually Located Underneath the Patient Instead of Above the Patient for Interventional Fluoroscopic Procedures?
    Dave, Jaydev K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (03) : W24 - W25
  • [40] VIDEO CHARACTER GENERATOR FOR FLUOROSCOPIC PROCEDURES