Improved-Resolution, Real-Time Skin-Dose Mapping for Interventional Fluoroscopic Procedures

被引:1
|
作者
Rana, Vijay K. [1 ]
Rudin, Stephen [1 ]
Bednarek, Daniel R. [1 ]
机构
[1] SUNY Buffalo, Dept Physiol, Buffalo, NY 14260 USA
关键词
Interventional fluoroscopic procedures; dose tracking system; skin dose; fluoroscopy exposure; dose mapping; GUIDED PROCEDURES; INJURIES; PATIENT;
D O I
10.1117/12.2042519
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have developed a dose-tracking system (DTS) that provides a real-time display of the skin-dose distribution on a 3D patient graphic during fluoroscopic procedures. Radiation dose to individual points on the skin is calculated using exposure and geometry parameters from the digital bus on a Toshiba C-arm unit. To accurately define the distribution of dose, it is necessary to use a high-resolution patient graphic consisting of a large number of elements. In the original DTS version, the patient graphics were obtained from a library of population body scans which consisted of larger-sized triangular elements resulting in poor congruence between the graphic points and the x-ray beam boundary. To improve the resolution without impacting real-time performance, the number of calculations must be reduced and so we created software-designed human models and modified the DTS to read the graphic as a list of vertices of the triangular elements such that common vertices of adjacent triangles are listed once. Dose is calculated for each vertex point once instead of the number of times that a given vertex appears in multiple triangles. By reformatting the graphic file, we were able to subdivide the triangular elements by a factor of 64 times with an increase in the file size of only 1.3 times. This allows a much greater number of smaller triangular elements and improves resolution of the patient graphic without compromising the real-time performance of the DTS and also gives a smoother graphic display for better visualization of the dose distribution.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures
    Bednarek, Daniel R.
    Barbarits, Jeffery
    Rana, Vijay K.
    Nagaraja, Srikanta P.
    Josan, Madhur S.
    Rudin, Stephen
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [2] Real-Time Patient Skin Dose Mapping for Fluoroscopically Guided Interventional Procedures
    Bednarek, Daniel R.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 33 (03) : 233 - 237
  • [3] Use of a graphics processing unit (GPU) to facilitate real-time 3D graphic presentation of the patient skin-dose distribution during fluoroscopic interventional procedures
    Rana, Vijay
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [4] Development of Graphic Patient Models for a Real-Time Skin Dose Tracking System (DTS) for Fluoroscopic Interventional Procedures
    Rana, V.
    Bednarek, D.
    Wu, J.
    Rudin, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3631 - 3631
  • [5] A Real-Time Skin-Dose Mapping System for Region-Of-Interest (ROI) Fluoroscopy
    Vijayan, S.
    Rana, V.
    Xiong, Z.
    Nagesh, S. Setlur
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2015, 42 (06) : 3717 - 3717
  • [6] Development of a real-time scattered radiation display for staff dose reduction during fluoroscopic interventional procedures
    Troville, J.
    Kilian-Meneghin, J.
    Guo, C.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [7] Extension of a Real-Time Skin Dose Tracking System to Include Pediatric Fluoroscopic Cardiac Procedures
    Rana, V.
    Bednarek, D.
    Wu, J.
    Rudin, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [8] A Real-Time Skin Dose Tracking System for Biplane Neuro-Interventional Procedures
    Rana, Vijay K.
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [9] Automatic Monitoring of Localized Skin Dose with Fluoroscopic and Interventional Procedures
    Yasaman Khodadadegan
    Muhong Zhang
    William Pavlicek
    Robert G. Paden
    Brian Chong
    Beth A. Schueler
    Kenneth A. Fetterly
    Steve G. Langer
    Teresa Wu
    Journal of Digital Imaging, 2011, 24 : 626 - 639
  • [10] Automatic Monitoring of Localized Skin Dose with Fluoroscopic and Interventional Procedures
    Khodadadegan, Yasaman
    Zhang, Muhong
    Pavlicek, William
    Paden, Robert G.
    Chong, Brian
    Schueler, Beth A.
    Fetterly, Kenneth A.
    Langer, Steve G.
    Wu, Teresa
    JOURNAL OF DIGITAL IMAGING, 2011, 24 (04) : 626 - 639