Performance assessment of a cavity ring-down laser spectrometer: achieving better precision and accuracy in the measurement of δ18O and δ2H in liquid water samples

被引:2
|
作者
Prado-Perez, A. J. [1 ]
Rodriguez-Arevalo, J. [1 ]
Diaz-Teijeiro, M. F. [1 ]
机构
[1] Ctr Estudios & Expt Obras Publ CEDEX, CETA, Madrid 28014, Spain
关键词
cavity ring-down spectroscopy; performance assessment; isotope ratio mass spectrometry; delta O-18; delta H-2; stable isotopes; ENHANCED ABSORPTION; SPECTROSCOPY; ISOTOPES; CALIBRATION; DEUTERIUM; OXYGEN; H-2;
D O I
10.1088/0957-0233/25/1/015801
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The development of new isotopic laser-based analyzers currently represents a clear alternative to conventional isotope ratio mass spectrometers. However, this analytical technique also suffers some disadvantages such as the memory effect, problems related to the overall stability of the equipment and other issues associated with the injection system, essentially regarding the syringe's longevity. This paper aims to minimize these disadvantages in order to increase the overall performance, in terms of precision and accuracy, of these kinds of analyzers. The main results of the experiments carried out in this paper have shown that: (i) the minimum number of discarded injections needed to eliminate the memory effect can be determined just considering the expected isotopic signature difference between two consecutive samples; (ii) both accuracy and precision of the isotopic measurements increase with increasing injection volume up to 2.1-2.2 mu L; (iii) it is possible to extend the syringe lifetime by almost a factor of 6 by using n-methyl 2-pyrrolidone as a lubricant. Besides, it has been concluded that, by using the appropriate procedure, the main disadvantages associated with CRDS laser spectroscopy analyzers can be minimized, achieving measurement accuracy and precision of the order of +/- 0.05 parts per thousand for delta O-18 and +/- 0.3 parts per thousand for delta H-2.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Isotope 18O/16O ratio measurements of water vapor by use of the 950-nm wavelength region with cavity ring-down and photoacoustic spectroscopic techniques
    Samura, K
    Hashimoto, S
    Kawasaki, M
    Hayashida, A
    Kagi, E
    Ishiwata, T
    Matsumi, Y
    APPLIED OPTICS, 2002, 41 (12) : 2349 - 2354
  • [42] Simultaneous δ2H and δ18O analyses of water inclusions in halite with off-axis integrated cavity output spectroscopy
    Fourel, Francois
    Lecuyer, Christophe
    Jame, Patrick
    Guironnet, Alexandre
    Boutier, Antoine
    Barbier, Manon
    Blamey, Nigel
    Brand, Uwe
    Fralick, Philip
    JOURNAL OF MASS SPECTROMETRY, 2020, 55 (10):
  • [43] A Compact Diode Laser Cavity Ring-Down Spectrometer for Atmospheric Measurements of NO3 and N2O5 with Automated Zeroing and Calibration
    Odame-Ankrah, Charles A.
    Osthoff, Hans D.
    APPLIED SPECTROSCOPY, 2011, 65 (11) : 1260 - 1268
  • [44] The use of water stable isotopes for a better understanding of hydrogeological processes in Haiti: overview of existing δ18O and δ2H data
    Gourcy, L.
    Adamson, J. K.
    Miner, W. J.
    Vitvar, T.
    Belizaire, D.
    HYDROGEOLOGY JOURNAL, 2022, 30 (05) : 1387 - 1397
  • [45] High-precision δ2H and δ18O measurement for water and volatile organic compounds by continuous-flow pyrolysis isotope ratio mass spectroetry
    Begley, Ian S.
    Scrimgeour, Charles M.
    Analytical Chemistry, 1997, 69 (08):
  • [46] Simultaneous measurement of δ13C, δ18O and δ17O of atmospheric CO2 - performance assessment of a dual-laser absorption spectrometer
    Steur, Pharahilda M.
    Scheeren, Hubertus A.
    Nelson, Dave D.
    McManus, J. Barry
    Meijer, Harro A. J.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2021, 14 (06) : 4279 - 4304
  • [47] Development and verification of a novel isotopic N2O measurement technique for discrete static chamber samples using cavity ring-down spectroscopy
    Bracken, Conor J.
    Lanigan, Gary J.
    Richards, Karl G.
    Muller, Christoph
    Tracy, Saoirse R.
    Well, Reinhard
    Carolan, Rachael
    Murphy, Paul N. C.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2021, 35 (08)
  • [48] Continuous flow 2H/1H and and18O/16O analysis of water samples with dual inlet precision
    Gehre, M
    Geilmann, H
    Richter, J
    Werner, RA
    Brand, WA
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (22) : 2650 - 2660
  • [49] Characterizing spatial and temporal variation in 18O and 2H content of New Zealand river water for better understanding of hydrologic processes
    Yang, Jing
    Dudley, Bruce D.
    Montgomery, Kelsey
    Hodgetts, Will
    HYDROLOGICAL PROCESSES, 2020, 34 (26) : 5474 - 5488
  • [50] A simple rain collector preventing water re-evaporation dedicated for δ18O and δ2H analysis of cumulative precipitation samples
    Groening, M.
    Lutz, H. O.
    Roller-Lutz, Z.
    Kralik, M.
    Gourcy, L.
    Poeltenstein, L.
    JOURNAL OF HYDROLOGY, 2012, 448 : 195 - 200