Performance assessment of a cavity ring-down laser spectrometer: achieving better precision and accuracy in the measurement of δ18O and δ2H in liquid water samples

被引:2
|
作者
Prado-Perez, A. J. [1 ]
Rodriguez-Arevalo, J. [1 ]
Diaz-Teijeiro, M. F. [1 ]
机构
[1] Ctr Estudios & Expt Obras Publ CEDEX, CETA, Madrid 28014, Spain
关键词
cavity ring-down spectroscopy; performance assessment; isotope ratio mass spectrometry; delta O-18; delta H-2; stable isotopes; ENHANCED ABSORPTION; SPECTROSCOPY; ISOTOPES; CALIBRATION; DEUTERIUM; OXYGEN; H-2;
D O I
10.1088/0957-0233/25/1/015801
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The development of new isotopic laser-based analyzers currently represents a clear alternative to conventional isotope ratio mass spectrometers. However, this analytical technique also suffers some disadvantages such as the memory effect, problems related to the overall stability of the equipment and other issues associated with the injection system, essentially regarding the syringe's longevity. This paper aims to minimize these disadvantages in order to increase the overall performance, in terms of precision and accuracy, of these kinds of analyzers. The main results of the experiments carried out in this paper have shown that: (i) the minimum number of discarded injections needed to eliminate the memory effect can be determined just considering the expected isotopic signature difference between two consecutive samples; (ii) both accuracy and precision of the isotopic measurements increase with increasing injection volume up to 2.1-2.2 mu L; (iii) it is possible to extend the syringe lifetime by almost a factor of 6 by using n-methyl 2-pyrrolidone as a lubricant. Besides, it has been concluded that, by using the appropriate procedure, the main disadvantages associated with CRDS laser spectroscopy analyzers can be minimized, achieving measurement accuracy and precision of the order of +/- 0.05 parts per thousand for delta O-18 and +/- 0.3 parts per thousand for delta H-2.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] High-Precision Calculation of the Proportions of Water with δ2H and δ18O, the Cumulative Effect of Evaporation in the Vertical Direction and Depleted δ2H and δ18O of the Shallow Soil Water Caused by Evaporation
    Zeng, Zhenyu
    Zhang, Xiang
    Pan, Guoyan
    Xiao, Yang
    WATER, 2022, 14 (17)
  • [22] Isotopic (δ18O and δ2H) Integrity of Water Samples Collected and Stored by Automatic Samplers
    Williams, Mark R.
    Lartey, Jessica L.
    Sanders, Laura L.
    AGRICULTURAL & ENVIRONMENTAL LETTERS, 2018, 3 (01)
  • [23] High Resolution Pore Water δ2H and δ18O Measurements by H2O(liquid)-H2O(vapor) Equilibration Laser Spectroscopy
    Wassenaar, L. I.
    Hendry, M. J.
    Chostner, V. L.
    Lis, G. P.
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2008, 42 (24) : 9262 - 9267
  • [24] Portable cavity ring-down spectrometer for an HO2 radical measurement: instrument's performance and potential improvement using a narrow linewidth laser
    Wang, Chunhui
    Zhao, Weixiong
    Fang, Bo
    Yang, Nana
    Cheng, Feihu
    Hu, Xiao
    Chen, Yang
    Zhang, Weijun
    Fittschen, Christa
    Chen, Weidong
    OPTICS EXPRESS, 2022, 30 (21) : 37446 - 37456
  • [25] A water isotope (2H, 17O, and 18O) spectrometer based on optical feedback cavity-enhanced absorption for in situ airborne applications
    Kerstel, E. R. T.
    Iannone, R. Q.
    Chenevier, M.
    Kassi, S.
    Jost, H. -J.
    Romanini, D.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2006, 85 (2-3): : 397 - 406
  • [26] A water isotope (2H, 17O, and 18O) spectrometer based on optical feedback cavity-enhanced absorption for in situ airborne applications
    E.R.T. Kerstel
    R.Q. Iannone
    M. Chenevier
    S. Kassi
    H.-J. Jost
    D. Romanini
    Applied Physics B, 2006, 85 : 397 - 406
  • [28] Rapid Scanning Cavity Ring-down Spectrometer for the Precision Measurement of 13C/12C for CO2 in Air
    Yi, Hongming
    Srivastava, Abneesh
    Fleisher, Adam J.
    Hodges, Joseph T.
    2018 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2018,
  • [29] Correcting for Methane Interferences on δ2H and δ18O Measurements in Pore Water Using H2O(liquid)-H2O(vapor) Equilibration Laser Spectroscopy
    Hendry, M. Jim
    Richman, B.
    Wassenaar, L. I.
    ANALYTICAL CHEMISTRY, 2011, 83 (14) : 5789 - 5796
  • [30] Water use by woody plants on contrasting soils in a savanna parkland:: assessment with δ2H and δ18O
    Midwood, AJ
    Boutton, TW
    Archer, SR
    Watts, SE
    PLANT AND SOIL, 1998, 205 (01) : 13 - 24