Technical Note: Homogeneity of Gafchromic® EBT2 film

被引:105
|
作者
Hartmann, Bernadette [1 ,2 ]
Martisikova, Maria [1 ]
Jaekel, Oliver [1 ,3 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, D-69120 Heidelberg, Germany
[2] Univ Halle Wittenberg, Inst Phys, D-06120 Halle, Germany
[3] Univ Klinikum Heidelberg, Heidelberger Lonenstrahl Therapiezentrum HIT, D-69120 Heidelberg, Germany
关键词
dosimeters; dosimetry; radiation therapy; FLAT-BED SCANNER; RADIOCHROMIC FILM; DOSIMETRY;
D O I
10.1118/1.3368601
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A scan protocol established for EBT film and published previously was used. The uniformity of the film coloration was investigated for unirradiated and irradiated EBT2 film sheets. The dose response of EBT2 film was measured and the influence of film inhomogeneities on dose determination was evaluated. Results: Inhomogeneities in pixel values of up to +/- 3.7% within one film were detected. The relative inhomogeneities were found to be approximately independent of the dose. Nonuniformities of the film response lead to uncertainties in dose determination of +/- 8.7% at 1 Gy. When using net optical densities for dose calibration, uncertainties in dose determination amount to more than +/- 6%. Conclusions: EBT2 films from the lot investigated in this study show response inhomogeneities, which lead to uncertainties in dose determination exceeding the commonly accepted tolerance levels. It is important to test further EBT2 lots regarding homogeneity before using the film in clinical routine.
引用
收藏
页码:1753 / 1756
页数:4
相关论文
共 50 条
  • [1] Comparing Homogeneity Between Gafchromic Film EBT2 and EBT3
    Mizuno, H.
    Sumida, I.
    Tanaka, A.
    Ogawa, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 239 - 239
  • [2] Homogeneity of GAFCHROMIC EBT2 film among different lot numbers
    Mizuno, Hirokazu
    Takahashi, Yutaka
    Tanaka, Atsushi
    Hirayama, Takamitsu
    Yamaguchi, Tsuyoshi
    Katou, Hiroaki
    Takahara, Keiko
    Okamoto, Yoshiaki
    Teshima, Teruki
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04): : 198 - 205
  • [3] Neutron response of GafChromic® EBT2 film
    Hsiao, Ming-Chen
    Liu, Yuan-Hao
    Chen, Wei-Lin
    Jiang, Shiang-Huei
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05): : 1391 - 1413
  • [4] Technical Note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry
    Desroches, Joannie
    Bouchard, Hugo
    Lacroix, Frederic
    MEDICAL PHYSICS, 2010, 37 (04) : 1565 - 1570
  • [5] A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT
    Andres, C.
    del Castillo, A.
    Tortosa, R.
    Alonso, D.
    Barquero, R.
    MEDICAL PHYSICS, 2010, 37 (12) : 6271 - 6278
  • [6] A Study of Dose Response Characteristic for Gafchromic EBT2 Film
    Chen, W.
    Di, X.
    Wang, S.
    Wang, B.
    Bai, X.
    Yang, J.
    Guo, B.
    Huang, K.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Absorption spectrum of gafchromic® EBT2 film with angular rotation
    Soah Park
    Kwang-Ho Cheong
    Taejin Hwang
    Jai-Woong Yoon
    Taejin Han
    Haeyoung Kim
    Me-Yeon Lee
    KyoungJu Kim
    Hoonsik Bae
    Sei-Kwon Kang
    Seung Jae Oh
    Yong-Min Huh
    Jin-Suck Suh
    Journal of the Korean Physical Society, 2015, 67 : 52 - 56
  • [8] Evaluation of the Gafchromic® EBT2 film for the dosimetry of radiosurgical beams
    Larraga-Gutierrez, Jose M.
    Garcia-Hernandez, Diana
    Garcia-Garduno, Olivia A.
    Galvan de la Cruz, Olga O.
    Ballesteros-Zebadua, Paola
    Esparza-Moreno, Karina P.
    MEDICAL PHYSICS, 2012, 39 (10) : 6111 - 6117
  • [9] Verification of CyberKnife Dose Distributions with Gafchromic EBT2 Film
    Koksal, Canan
    Akbas, Ugur
    Kesen, Nazmiye Donmez
    Bilge, Hatice
    Kemikler, Gonul
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2018, 33 (04): : 147 - 153
  • [10] Absorption Spectrum of Gafchromic® EBT2 Film with Angular Rotation
    Park, Soah
    Cheong, Kwang-Ho
    Hwang, Taejin
    Yoon, Jai-Woong
    Han, Taejin
    Kim, Haeyoung
    Lee, Me-Yeon
    Kim, KyoungJu
    Bae, Hoonsik
    Kang, Sei-Kwon
    Oh, Seung Jae
    Huh, Yong-Min
    Suh, Jin-Suck
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 52 - 56