In-house quality check of external beam plans using 3D treatment planning systems - a DVH comparison

被引:1
|
作者
Kumar, Ayyalasomayajula Anil [1 ]
Akula, Roopa Rani [4 ]
Ayyangar, Komanduri [4 ]
Reddy, Krishna P. [2 ,3 ]
Vuppu, Srinivas [2 ,3 ]
Narayana, P. V. Lakshmi [1 ]
Rao, A. Durga Prasada [1 ]
机构
[1] Andhra Univ, Dept Nucl Phys, Swami Jnanananda Labs Nucl Res, Visakhapatnam 530003, Andhra Pradesh, India
[2] MNJ Inst Oncol, Hyderabad, Andhra Pradesh, India
[3] Reg Canc Ctr, Hyderabad, Andhra Pradesh, India
[4] MGMMT, Int Canc Ctr, Bhimavaram, India
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2016年 / 17卷 / 03期
关键词
TPS; DVH; cobalt-60; Eclipse; ROPS; RADIOTHERAPY; ASSURANCE; PHOTON;
D O I
10.1120/jacmp.v17i3.6020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper presents a new approach towards the quality assurance of external beam plans using in-house-developed DICOM import and export software in a clinical setup. The new approach is different from what is currently used in most clinics, viz., only MU and point dose are verified. The DICOM-RT software generates ASCII files to import/export structure sets, treatment beam data, and dose-volume histograms (DVH) from one treatment planning system (TPS) to the other. An efficient and reliable 3D planning system, ROPS, was used for verifying the accuracy of treatment plans and treatment plan parameters. With the use of this new approach, treatment plans planned using Varian Eclipse planning system were exported to ROPS planning system. Important treatment and dosimetrical data, such as the beam setup accuracy, target dose coverage, and dose to critical structures, were also quantitatively verified using DVH comparisons. Two external beam plans with diverse photon energies were selected to test the new approach. The satisfactory results show that the new approach is feasible, easy to use, and can be used as an adjunct test for patient treatment quality check.
引用
收藏
页码:138 / 146
页数:9
相关论文
共 50 条
  • [41] Open hole tensile testing of 3D printed parts using in-house fabricated PLA filament
    Dave, Harshit K.
    Prajapati, Ashish R.
    Rajpurohit, Shilpesh R.
    Patadiya, Naushil H.
    Raval, Harit K.
    RAPID PROTOTYPING JOURNAL, 2020, 26 (01) : 21 - 31
  • [42] The effect of location and size of planning target volume (PTV) on the dose-volume histogram (DVH) of the ipsilateral breast in 3D conformal, external beam partial breast radiotherapy (PBRT)
    Kee, AY
    Choo, R
    Debra, B
    Petersen, I
    Schomberg, P
    Schickell, D
    Pisansky, T
    Garces, Y
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S233 - S234
  • [43] Dosimetric Comparison of 2D and 3D Episcleral Plaque Treatment Planning
    Dapaah-Afriyie, B.
    Wilkinson, D.
    Astrahan, M.
    Singh, A.
    Berry, J.
    MEDICAL PHYSICS, 2017, 44 (06) : 3080 - 3080
  • [44] Evaluation of the BacT/ALERT® 3D system for the implementation of in-house quality control sterility testing at Canadian Blood Services
    Mastronardi, Cherie
    Perkins, Heather
    Derksen, Paul
    denAdmirant, Michael
    Ramirez-Arcos, Sandra
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (08) : 1179 - 1187
  • [45] ELECTRON-BEAM TREATMENT PLANNING USING A 3-D PENCIL BEAM ALGORITHM
    JAPP, B
    CUNNINGHAM, JR
    MCPARLAND, BF
    MEDICAL PHYSICS, 1986, 13 (04) : 604 - 604
  • [46] Comparison of accelerated partial breast irradiation (APBI) with whole breast irradiation (WBI) using 3D conformal external beam radiation therapy (3D CRT)
    Yadav, B. S.
    Sharma, S. C.
    Singh, R.
    Dahiya, D.
    CANCER RESEARCH, 2019, 79 (04)
  • [47] A dosimetric comparison of 3 types of breast treatment plans. 3D conformal, RapidArc and IMRT
    Pella, S.
    Dumitru, N.
    Pudasani, M.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S995 - S995
  • [48] 3D gamma analysis between treatment plans for nominally beam-matched medical linear accelerators using PyMedPhys
    Guan, Fada
    Donahue, William
    Biggs, Simon
    Jennings, Matthew
    Draeger, Emily
    Chen, Huixiao
    Wang, Yuenan
    Nguyen, Ngoc
    Carlson, David J.
    Chen, Zhe
    Han, Dae Yup
    PRECISION RADIATION ONCOLOGY, 2024,
  • [49] 3D TREATMENT PLANNING OF CONICAL BREAST BRACHYTHERAPY APPLICATORS USING CONVENTIONAL TREATMENT PLANNING SOFTWARE
    Yang, Y.
    Melhus, C.
    Sioshansi, S.
    Rivard, M.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S402 - S402
  • [50] Automated prostate localization in external-beam radiotherapy using mutual information-based registration of treatment planning CT and daily 3D ultrasound images
    Castro-Pareja, CR
    Zagrodsky, V
    Bouchet, L
    Shekhar, R
    CARS 2005: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2005, 1281 : 435 - 440