Energy calibration of CsI(T1) scintillator in pulse-shape identification technique

被引:16
|
作者
Avdeichkov, V
Ghetti, R
Golubev, P
Jakobsson, B
Colonna, N
机构
[1] Lund Univ, Dept Phys, S-22100 Lund, Sweden
[2] Joint Inst Nucl Res, Dubna 141980, Russia
[3] Ist Nazl Fis Nucl, I-70126 Bari, Italy
关键词
inorganic scintillators; light ions response function; pulse-shape identification; calibration;
D O I
10.1016/S0168-9002(03)00625-9
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A batch of 16 CsI(TI) scintillator crystals, supplied by the Bicron Company, has been studied with respect to precise energy calibration in pulse-shape identification technique. The light corresponding to pulse integration within the time interval 1.6-4.5 mus (long gate) and 0.0-4.5 mus (extra-long gate) exhibits a power law relation, L(E, Z, A) a1 (Z, A)Ea2(Z-4), for (1.2.3) H isotopes in the measured energy range 5-150 MeV. For the time interval 0.0-0.60 mus (short gate), a significant deviation from the power law relation is observed, for energy greater than similar to30 MeV. The character of the a2(p)-a2(d) and a2(p)-a2(t) correlations for protons, deuterons and tritons, reveals 3 types of crystals in the batch. These subbatches differ in the value of the extracted parameter a2 for protons, and in the value of the spread of a2 for deuterons and tritons. This may be explained by the difference in the energy dependence of the fast decay time component and/or by the difference in the light output ratio of the fast/slow components. An accuracy well inside 1.0% is achieved in the energy calibration of the CsI(TI) crystals by using the DeltaE(Si) E(CsI(Tl))/PMT method. The batch of 16 CsI(TI) crystals is utilized to measure the correlations of light charged particles produced in E/A = 61 MeV, (36)Arinduced reactions. The preliminary correlation functions for two protons with small relative momenta are presented. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
下载
收藏
页码:505 / 513
页数:9
相关论文
共 50 条
  • [21] Pulse shape discrimination and dark matter search with NaI(T1) scintillator at the Laboratoire Souterrain de Modane
    Gerbier, G
    Mallet, J
    Mosca, L
    Tao, C
    SOURCES AND DETECTION OF DARK MATTER AND DARK ENERGY IN THE UNIVERSE, 2001, : 333 - 339
  • [22] RECOIL-PROTON SPECTROMETER USING PULSE-SHAPE DISCRIMINATION IN CSI(TI)
    HARIHAR, P
    NUCLEAR INSTRUMENTS & METHODS, 1975, 127 (03): : 387 - 390
  • [23] Pulse-shape discrimination in neutron depth profiling technique
    Vacik, J
    Cervena, J
    Hnatowicz, V
    Havranek, V
    Hoffmann, J
    Posta, S
    Fink, D
    Klett, R
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1998, 142 (03): : 397 - 401
  • [24] A pulse shape discrimination gamma-ray detector based on a silicon drift chamber coupled to a CsI(T1) scintillator: Prospects for a 1 keV-1 MeV monolithic detector
    Marisaldi, M
    Labanti, C
    Soltau, H
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (04) : 1916 - 1922
  • [26] Non-proportionality and energy resolution of CsI(T1)
    Syntfeld-Kazuch, Agnieszka
    Swiderski, Lukasz
    Czarnacki, Wieslaw
    Gierlik, Michal
    Klamra, Wlodzimierz
    Moszynski, Marek
    Schotanus, Paul
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) : 1836 - 1841
  • [27] PULSE-HEIGHT AND PULSE-SHAPE RESPONSE OF CSI(TL) TO P D T HE3 AND HE4
    JOBLIN, AP
    GELLER, KN
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1968, 13 (11): : 1445 - &
  • [28] Pulse-shape discrimination potential of new scintillator material: La-GPS:Ce
    Mizukoshi, K.
    Iida, T.
    Ogawa, I
    Shimizu, K.
    Kurosawa, S.
    Karnada, K.
    Yoshino, M.
    Yoshikawa, A.
    JOURNAL OF INSTRUMENTATION, 2019, 14 (06)
  • [29] PULSE-SHAPE DISCRIMINATION APPLIED TO SCINTILLATING-TARGET TECHNIQUE
    VONMOELL.U
    SLOBODRI.RJ
    NUCLEAR INSTRUMENTS & METHODS, 1974, 120 (02): : 257 - 259
  • [30] PHOTODIODE READOUT AND PULSE SHAPE-ANALYSIS OF CSI(TI) SCINTILLATOR SIGNALS
    KREUTZ, P
    KUHMICHEL, A
    PINKENBURG, C
    POCHODZALLA, J
    GUO, ZY
    LYNEN, U
    SANN, H
    TRAUTMANN, W
    TROCKEL, R
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 260 (01): : 120 - 123