Comparison of HASTE and segmented-HASTE sequences with a T2-weighted fast spin-echo sequence in the screening evaluation of the brain

被引:32
|
作者
Sugahara, T
Korogi, Y
Hirai, T
Hamatake, S
Ikushima, I
Shigematu, Y
Takahashi, M
机构
[1] Department of Radiology, Kumamoto Univ. School of Medicine, Kumamoto 860
关键词
D O I
10.2214/ajr.169.5.9353469
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE, The purpose of this study was to evaluate the neuroradiologic application of half-Fourier acquisition single-shot turbo spin-echo (HASTE) and segmented-HASTE (s-HASTE) sequences in comparison with a T2-weighted fast spin-echo sequence. MATERIALS AND METHODS. First, HASTE, s-HASTE, and fast spin-echo sequences were evaluated for blurring artifacts with a stationary phantom and for motion artifacts with a moving phantom, which repeated constant or intermittent to-and-fro motions at variable intervals. Second, 30 consecutive patients with various intracranial diseases were prospectively examined with the three sequences. Lesions were classified into four groups according to size and signal intensity on fast spin-echo MR images as follows: large hyperintense, small hyperintense, small markedly hyperintense, and hypointense lesions. Signal intensities of the lesion, putamen, and gray matter were compared with the signal intensity of white matter, and contrast-to-noise ratios were calculated. Overall image quality, conspicuity of lesions, delineation of the junction between gray matter and white matter, conspicuity of the putamen, and certain types of artifacts were evaluated qualitatively. RESULTS. Ln the phantom study, the HASTE sequence was least affected by motion artifacts and the fast spin-echo sequence was most affected, although the images of the HASTE sequence were most degraded by blurring artifacts. In the clinical study, we found no significant differences among the three sequences for contrast-to-noise ratios or conspicuity of large hyperintense and small markedly hyperintense lesions. However, the contrast-to-noise ratios of hypointense lesions and gray matter, and the conspicuity of hypointense lesions were significantly poorer for the HASTE sequence than for the fast spin-echo sequence. The contrast-to-noise ratios of small hyperintense lesions and the putamen, conspicuity of small hyperintense lesions and putamen, and delineation of the junction between gray matter and white matter were significantly poorer for HASTE and s-HASTE sequences than for the fast spin-echo sequence. Ghost artifacts, which were observed during the s-HASTE sequence, were sometimes superimposed on the image. CONCLUSION. The HASTE and s-HASTE sequences afford substantial time reduction and also decrease motion artifacts and thus have potential advantages for neuroradiologic application, especially in uncooperative or unsedated children. The s-HASTE sequence may be preferable to the HASTE sequence because of fewer blurring artifacts and higher T2 contrast. However, small hyperintense and hypointense lesions may be overlooked when HASTE and s-HASTE sequences are used.
引用
收藏
页码:1401 / 1410
页数:10
相关论文
共 50 条
  • [21] Hepatic T2-weighted MRI: A prospective comparison of sequences, including breath-hold, half-Fourier turbo spin echo (HASTE)
    Coates, GG
    Borrello, JA
    McFarland, EG
    Mirowitz, SA
    Brown, JJ
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (03): : 642 - 649
  • [22] T2-weighted MRI of rectosigmoid carcinoma: Comparison of respiratory-triggered fast spin-echo, breathhold fast-recovery fast spin-echo, and breathhold single-shot fast spin-echo sequences
    Yamashita, Shuhei
    Masui, Takayuki
    Katayama, Motoyuki
    Sato, Kimihiko
    Yoshizawa, Nobuko
    Seo, Hidekazu
    Sakahara, Harumi
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 511 - 516
  • [23] Comparison of HASTE with multiple signal averaging versus conventional turbo spin echo sequence: a new option for T2-weighted MRI of the female pelvis
    Takahiro Tsuboyama
    Oki Takei
    Atsuhiko Okada
    Toru Honda
    Keiko Kuriyama
    European Radiology, 2020, 30 : 3245 - 3253
  • [24] Comparison of HASTE with multiple signal averaging versus conventional turbo spin echo sequence: a new option for T2-weighted MRI of the female pelvis
    Tsuboyama, Takahiro
    Takei, Oki
    Okada, Atsuhiko
    Honda, Toru
    Kuriyama, Keiko
    EUROPEAN RADIOLOGY, 2020, 30 (06) : 3245 - 3253
  • [25] T2-weighted MRI of the female pelvis: Comparison of breath-hold fast-recovery fast spin-echo and nonbreath-hold fast spin-echo sequences
    Masui, T
    Katayama, M
    Kobayashi, S
    Sakahara, H
    Ito, T
    Nozaki, A
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (06) : 930 - 937
  • [26] Three-dimensional isotropic T2-weighted fast spin-echo (VISTA) ankle MRI versus two-dimensional fast spin-echo T2-weighted sequences for the evaluation of anterior talofibular ligament injury
    Park, H. J.
    Lee, S. Y.
    Park, N. H.
    Rho, M. H.
    Chung, E. C.
    Park, J. H.
    Park, S. J.
    CLINICAL RADIOLOGY, 2016, 71 (04) : 349 - 355
  • [27] Half-Fourier acquisition single-shot turbo spin-echo (HASTE) MR: Comparison with fast spin-echo MR in diseases of the brain
    Patel, MR
    Klufas, RA
    Alberico, RA
    Edelman, RR
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1997, 18 (09) : 1635 - 1640
  • [28] Susceptibility-weighted Imaging for the evaluation of patients with familial cerebral cavernous malformations: A comparison with T2-weighted fast spin-echo and gradient-echo sequences
    de Souza, J. M.
    Domingues, R. C.
    Cruz, L. C. H.
    Domingues, F. S.
    Lasbeck, T.
    Gasparetto, E. L.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2008, 29 (01) : 154 - 158
  • [29] The value of ultrafast Half-Fourier Single-Shot Turbo Spin-Echo (HASTE) sequence for T2-weighted imaging of the kidney: Comparison with respiratory-triggered turbo SE sequence
    Nakayama, Y
    Yamashita, Y
    Matsuno, Y
    Sakaguchi, T
    Tang, Y
    Takahashi, M
    RADIOLOGY, 1998, 209P : 337 - 337
  • [30] Three-dimensional isotropic T2-weighted fast spin-echo (VISTA) knee MRI at 3.0T in the evaluation of the anterior cruciate ligament injury with additional views: comparison with two-dimensional fast spin-echo T2-weighted sequences
    Park, Hee Jin
    Lee, So Yeon
    Park, Noh Hyuck
    Ahn, Jin Hwan
    Chung, Eun Chul
    Kim, Sung Jun
    Cha, Jang Gyu
    ACTA RADIOLOGICA, 2016, 57 (11) : 1372 - 1379