Speech perception, localization, and lateralization with bilateral cochlear implants

被引:355
|
作者
van Hoesel, RJM [1 ]
Tyler, RS
机构
[1] CRC Cochlear Implant & Hearing Aid Innovat, Melbourne, Vic, Australia
[2] Univ Iowa, Dept Otolaryngol Head & Neck Surg, Iowa City, IA 52242 USA
[3] Univ Iowa, Dept Speech Pathol & Audiol, Iowa City, IA 52242 USA
来源
关键词
D O I
10.1121/1.1539520
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Five bilateral cochlear implant users were tested for their localization abilities and speech understanding in noise, for both monaural and binaural listening conditions. They also participated in lateralization tasks to assess the impact of variations in interaural time delays (ITDs) and interaural level differences (ILDs) for, electrical pulse trains under direct computer control. The localization task used pink noise bursts presented from an, eight-loudspeaker array spanning an arc of approximately 108degrees in front of the listeners at ear level (0-degree elevation). Subjects showed large benefits from bilateral device use compared to either side alone. Typical root-mean-square (rms) averaged errors across all eight loudspeakers in the array were about 10degrees for bilateral device use and ranged from 20degrees to 60degrees using either ear alone. Speech reception thresholds (SRTs) were measured for sentences presented from directly in front of the listeners (0degrees) in spectrally matching speech-weighted noise at either 0degrees, +90degrees or -90degrees for four subjects out of five tested who could perform the task. For noise to either side, bilateral device use showed a substantial benefit over unilateral device use when noise was ipsilateral to the unilateral device. This was primarily because of monaural head-shadow effects, which resulted in robust SRT improvements (P < 0.001) of about 4 to 5 dB when ipsilateral and contralateral noise positions were compared. The additional benefit of using both ears compared to the shadowed ear (i.e., binaural unmasking) was only 1 or 2 dB and less robust (P = 0.04). Results from the lateralization studies showed consistently good sensitivity to ILDs; better than the smallest level adjustment available in the implants (0.17 dB) for some subjects. Sensitivity to ITDs was moderate on the other hand, typically of the order of 100 mu s. ITD sensitivity deteriorated rapidly when stimulation rates for unmodulated pulse-trains increased above a few hundred Hz but at 800 pps showed sensitivity comparable to 50-pps pulse-trains when a 50-Hz modulation was applied. In our opinion, these results clearly demonstrate important benefits are available from bilateral implantation, both for localizing sounds (in quiet) and for listening in noise when signal and noise sources are spatially separated. The data do indicate, however, that effects of interaural timing cues are weaker than those from interaural level cues and according to our psychophysical findings rely on the availability of low-rate information below a few hundred Hz. (C) 2003 Acoustical Society of America.
引用
收藏
页码:1617 / 1630
页数:14
相关论文
共 50 条
  • [31] Bilateral Loudness Balancing and Distorted Spatial Perception in Recipients of Bilateral Cochlear Implants
    Fitzgerald, Matthew B.
    Kan, Alan
    Goupell, Matthew J.
    EAR AND HEARING, 2015, 36 (05): : E225 - E236
  • [32] Perception and coding of interaural time differences with bilateral cochlear implants
    Laback, Bernhard
    Egger, Katharina
    Majdak, Piotr
    HEARING RESEARCH, 2015, 322 : 138 - 150
  • [33] Localization ability with bimodal hearing aids and bilateral cochlear implants
    Seeber, BU
    Baumann, U
    Fastl, H
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2004, 116 (03): : 1698 - 1709
  • [34] Development of Sound Localization Strategies in Children with Bilateral Cochlear Implants
    Zheng, Yi
    Godar, Shelly P.
    Litovsky, Ruth Y.
    PLOS ONE, 2015, 10 (08):
  • [35] Sound localization ability of young children with bilateral cochlear implants
    Beijen, Jan-Willem
    Snik, Ad F. M.
    Mylanus, Emmanuel A. M.
    OTOLOGY & NEUROTOLOGY, 2007, 28 (04) : 479 - 485
  • [36] Localization ability with bimodal hearing aids and bilateral cochlear implants
    Seeber, Bernhard U.
    Baumann, Uwe
    Fastl, Hugo
    Journal of the Acoustical Society of America, 2004, 116 (03): : 1698 - 1709
  • [37] Effect of exceeding compliance voltage on speech perception in cochlear implants
    Saoji, Aniket A.
    Adkins, Weston J.
    Olund, Amy P.
    Nelson-Bakkum, Erin R.
    Koka, Kanthaiah
    HEARING RESEARCH, 2021, 400
  • [38] The effect of increased channel interaction on speech perception with cochlear implants
    Tobias Goehring
    Alan W. Archer-Boyd
    Julie G. Arenberg
    Robert P. Carlyon
    Scientific Reports, 11
  • [39] Electrical stimulation rate effects on speech perception in cochlear implants
    Arora, Komal
    Dawson, Pam
    Dowell, Richard
    Vandali, Andrew
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2009, 48 (08) : 561 - 567
  • [40] Prefrontal cortex supports speech perception in listeners with cochlear implants
    Sherafati, Arefeh
    Dwyer, Noel
    Bajracharya, Aahana
    Hassanpour, Mahlega Samira
    Eggebrecht, Adam T.
    Firszt, Jill B.
    Culver, Joseph P.
    Peelle, Jonathan E.
    Griffiths, Timothy D.
    ELIFE, 2022, 11