Assessment of the Quality of Breast MR Imaging Using the Modified Dixon Method and Frequency-selective Fat Suppression: A Phantom Study

被引:5
|
作者
Takatsu, Yasuo [1 ,2 ]
Kyotani, Katsusuke [3 ]
Ueyama, Tsuyoshi [4 ]
Miyati, Tosiaki [2 ]
Yamamura, Kenichirou [2 ,5 ]
Andou, Atsushi [6 ]
机构
[1] Osaka Red Cross Hosp, Dept Radiol, Tennouji Ku, 5-30 Fudegasaki, Osaka, Osaka 5438555, Japan
[2] Kanazawa Univ, Grad Sch Med Sci, Div Hlth Sci, Kanazawa, Ishikawa, Japan
[3] Kobe Univ Hosp, Div Radiol, Kobe, Hyogo, Japan
[4] Univ Tokyo Hosp, Dept Radiol, Tokyo, Japan
[5] Osaka Med Coll Hosp, Dept Radiol, Osaka, Japan
[6] Takarazuka City Hosp, Dept Radiol, Takarazuka, Hyogo, Japan
关键词
breast; Dixon; e-Thrive; fat suppression; magnetic resonance imaging; TESLA;
D O I
10.2463/mrms.tn.2017-0144
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To obtain objective and concrete data by physically assessing the quality of breast magnetic resonance images based on the fat-suppression effect by the modified Dixon method (mDixon) and frequency-selective fat suppression (e-Thrive) using an original lipid-content breast phantom that could easily reveal the influence of non-uniform fat suppression in breast magnetic resonance imaging. The fat-suppression uniformity was approximately seven times superior when using mDixon compared with when using e-Thrive. mDixon appears to have a significant advantage.
引用
收藏
页码:350 / 355
页数:6
相关论文
共 34 条
  • [31] Quantitative Assessment of Intrahepatic Lipids Using Fat-Selective Imaging With Spectral-Spatial Excitation and In-/Opposed-Phase Gradient Echo Imaging Techniques Within a Study Population of Extremely Obese Patients Feasibility on a Short, Wide-Bore MR Scanner
    Springer, Fabian
    Machann, Juergen
    Schwenzer, Nina F.
    Ballweg, Verena
    Wuerslin, Christian
    Schneider, Joachim H.
    Fritsche, Andreas
    Claussen, Claus D.
    Schick, Fritz
    INVESTIGATIVE RADIOLOGY, 2010, 45 (08) : 484 - 490
  • [32] Quantitative assessment of acquisition imaging parameters on MRI radiomics features: a prospective anthropomorphic phantom study using a 3D-T2W-TSE sequence for MR-guided-radiotherapy
    Yuan, Jing
    Xue, Cindy
    Lo, Gladys
    Wong, Oi Lei
    Zhou, Yihang
    Yu, Siu Ki
    Cheung, Kin Yin
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2021, 11 (05) : 1870 - +
  • [33] Correlation Between Contrast-Detail Analysis and Clinical Image Quality Assessment of Intrapulmonary Lesions in Dual-Energy Subtraction Chest Radiography Using the TwoShot Method: A Phantom Study
    Kuramoto, Taku
    Takarabe, Shinya
    Kanzaki, Yui
    Shibayama, Yusuke
    Yamasaki, Yuzo
    Kitamura, Yoshiyuki
    ACADEMIC RADIOLOGY, 2024, 31 (05) : 2118 - 2127
  • [34] Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test
    Wong, Oi Lei
    Yuan, Jing
    Yu, Siu Ki
    Cheung, Kin Yin
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2017, 7 (02) : 205 - 214