Correlation Between Contrast-Detail Analysis and Clinical Image Quality Assessment of Intrapulmonary Lesions in Dual-Energy Subtraction Chest Radiography Using the TwoShot Method: A Phantom Study

被引:0
|
作者
Kuramoto, Taku [1 ]
Takarabe, Shinya [2 ]
Kanzaki, Yui [2 ]
Shibayama, Yusuke [2 ]
Yamasaki, Yuzo [3 ]
Kitamura, Yoshiyuki [3 ]
机构
[1] Kobe Tokiwa Univ, Fac Hlth Sci, Dept Radiol Technol, 2-6-2 Otanicho,Nagata Ku, Kobe 6530838, Japan
[2] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi,Higashi Ku, Fukuoka 8128582, Japan
[3] Kyushu Univ, Grad Sch Med Sci, Dept Clin Radiol, 3-1-1,Maidashi,Higashi Ku, Fukuoka 8128582, Japan
基金
日本学术振兴会;
关键词
Dual-energy subtraction imaging; Contrast-detail phantom; Inverse image quality figure; Visual grading analysis; Image quality; PULMONARY NODULES; LUNG-CANCER; COMPUTED RADIOGRAPHY; OPTIMIZATION;
D O I
10.1016/j.acra.2023.10.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: Dual -energy subtraction (DES) imaging constitutes a technique that has demonstrated efficacy in enhancing the detectability of pulmonary nodules on chest radiographs. However, a simple and quantitative methodology for evaluating the clinical image quality of DES images is currently lacking. The objective of this study was to investigate the applicability of contrast -detail (C -D) phantom analysis to the visual clinical image quality evaluation of chest DES images. Materials and Methods: We employed a custom-made phantom incorporating the C -D phantom and a multipurpose anthropomorphic adult chest phantom. Two phantom sizes were utilized to simulate standard- and large -bodied adult patients for each phantom. The custom-made phantom images were scored automatically using dedicated software, yielding an inverse image quality figure (IQF inv ) value. The multipurpose anthropomorphic adult chest phantom was employed in a visual grading analysis (VGA) study that was conducted by two experienced radiologists and five radiological technologists. Each nodule placed in the chest phantom image was rated on a 4 -point Likert scale. Results: A statistically significant correlation was observed between the VGA scores of the seven observers and the obtained IQF inv values. Conclusion: The findings of this study suggest that DES image analysis of the C -D phantom possesses the potential to be utilized for the evaluation of clinical DES image quality based on chest lesion detectability.
引用
收藏
页码:2118 / 2127
页数:10
相关论文
共 8 条
  • [1] Correlation of Contrast-Detail Analysis and Clinical Image Quality Assessment in Chest Radiography with a Human Cadaver Study
    De Crop, An
    Bacher, Klaus
    Van Hoof, Tom
    Smeets, Peter V.
    Smet, Barbara S.
    Vergauwen, Merel
    Kiendys, Urszula
    Duyck, Philippe
    Verstraete, Koenraad
    D'Herde, Katharina
    Thierens, Hubert
    RADIOLOGY, 2012, 262 (01) : 298 - 304
  • [2] Effects of dual-energy subtraction chest radiography on detection of small pulmonary nodules with varying attenuation: receiver operating characteristic analysis using a phantom study
    Seitaro Oda
    Kazuo Awai
    Yoshinori Funama
    Daisuke Utsunomiya
    Yumi Yanaga
    Koichi Kawanaka
    Yasuyuki Yamashita
    Japanese Journal of Radiology, 2010, 28 : 214 - 219
  • [3] Effects of dual-energy subtraction chest radiography on detection of small pulmonary nodules with varying attenuation: receiver operating characteristic analysis using a phantom study
    Oda, Seitaro
    Awai, Kazuo
    Funama, Yoshinori
    Utsunomiya, Daisuke
    Yanaga, Yumi
    Kawanaka, Koichi
    Yamashita, Yasuyuki
    JAPANESE JOURNAL OF RADIOLOGY, 2010, 28 (03): : 214 - 219
  • [4] Analysis of dual-energy mammography subtraction technique for the dose and image quality evaluation using 3D-printed breast phantom
    Endarko, Fitria M.
    Celina, Fitria M.
    Gani, M. Roslan A.
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2023, 46 (04) : 1693 - 1701
  • [5] Analysis of dual-energy mammography subtraction technique for the dose and image quality evaluation using 3D-printed breast phantom
    Fitria M. Endarko
    M. Roslan A. Celina
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1693 - 1701
  • [6] Evaluation of image quality and radiation dose using gold nanoparticles and other clinical contrast agents in dual-energy Computed Tomography (CT): CT abdomen phantom
    Zukhi, J.
    Yusob, D.
    Tajuddin, A. A.
    Vuanghao, L.
    Zainon, R.
    10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [7] Treatment planning based on water density image generated using dual-energy computed tomography for pancreatic cancer with contrast-enhancing agent: Phantom and clinical study
    Ohira, Shingo
    Yagi, Masashi
    Iramina, Hiraku
    Karino, Tsukasa
    Washio, Hayate
    Ueda, Yoshihiro
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Teshima, Teruki
    MEDICAL PHYSICS, 2018, 45 (11) : 5208 - 5217
  • [8] A phantom study using dual-energy spectral computed tomography imaging: Comparison of image quality between two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms for evaluating ground-glass nodules of the lung
    Li, Jun
    Gao, Jun
    Wang, Guangli
    Cui, Dingye
    Deng, Kai
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2021, 17 (07) : 1742 - 1747