Analysis of dual-energy mammography subtraction technique for the dose and image quality evaluation using 3D-printed breast phantom

被引:0
|
作者
Endarko, Fitria M. [1 ]
Celina, Fitria M. [1 ]
Gani, M. Roslan A. [2 ]
机构
[1] Kampus ITS, Lab Med Phys & Biophys, Inst Teknol Sepuluh Nopember, Surabaya 60111, East Java, Indonesia
[2] Dharmais Natl Canc Ctr Hosp, Dept Radiodiagnost, Jakarta, Indonesia
关键词
Mammography; Dual-energy; Doses; Image quality; ENHANCED SPECTRAL MAMMOGRAPHY; DIGITAL MAMMOGRAPHY; DIAGNOSTIC PERFORMANCE; TOMOSYNTHESIS; CESM;
D O I
10.1007/s13246-023-01330-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study aimed to quantitatively assess the radiation dose using XR-QA2 and the image quality of the dual-energy subtraction mammography technique on an in-house phantom. The analysis was carried out to investigate the effect of targets/filters on dose value and image quality using an in-house phantom made of PLA + as an object representing compressed breasts. All irradiation parameters were performed in the craniocaudal position with manual mode. Mean glandular dose (MGD) was recorded, followed by the calculation of the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and modulation transfer function (MTF) for image quality assessment parameters. The results showed that the image quality was accepted at dose levels within the IAEA and BAPETEN tolerance limit for 60 mm equivalent compressed breast using dual-energy mammography. Furthermore, the target/filter (W/Rh) reduced the dose by 1.03 mGy compared to the Mo/Mo and Mo/Rh with an enhancement in image quality. This indicated that the target/filter (W/Rh) combination was optimal due to the image quality improvement obtained with lower MGD.
引用
收藏
页码:1693 / 1701
页数:9
相关论文
共 50 条
  • [1] Analysis of dual-energy mammography subtraction technique for the dose and image quality evaluation using 3D-printed breast phantom
    Fitria M. Endarko
    M. Roslan A. Celina
    Physical and Engineering Sciences in Medicine, 2023, 46 : 1693 - 1701
  • [2] Design, Manufacturing and Quality Assessment of 3D-Printed Anthropomorphic Breast Phantom for Mammography
    Huseljic, Elma
    Odzak, Senad
    Beganovic, Adnan
    Odzak, Almasa
    Pandzic, Adi
    Jusufbegovic, Merim
    MEDICON 2023 AND CMBEBIH 2023, VOL 1, 2024, 93 : 221 - 231
  • [3] Dual-energy computed tomography of the neck-optimizing tube current settings and radiation dose using a 3D-printed patient phantom
    Diekhoff, Torsten
    Scheel, Michael
    Kress, Wiebke
    Hamm, Bernd
    Jahnke, Paul
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2021, 11 (04) : 1144 - U17
  • [4] Quantification of breast density using dual-energy mammography with liquid phantom calibration
    Lam, Alfonso R.
    Ding, Huanjun
    Molloi, Sabee
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (14): : 3985 - 4000
  • [5] Estimates of the Image Quality and the Radiation Dose for Head and Abdomen Phantom Image Acquisition by Using Dual-energy CT
    Kim, Dae-Hong
    Kim, Hee-Joung
    Lee, Chang-Lae
    Jeon, Pil-Hyun
    Lee, Won-Hyung
    Jeon, Sung-Soo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 60 (07) : 1150 - 1156
  • [6] Estimates of the image quality and the radiation dose for head and abdomen phantom image acquisition by using dual-energy CT
    Dae-Hong Kim
    Hee-Joung Kim
    Chang-Lae Lee
    Pil-Hyun Jeon
    Won-Hyung Lee
    Sung-Soo Jeon
    Journal of the Korean Physical Society, 2012, 60 : 1150 - 1156
  • [7] Dual-Energy CT Angiography in the Evaluation of Intracranial Aneurysms: Image Quality, Radiation Dose, and Comparison With 3D Rotational Digital Subtraction Angiography
    Zhang, Long-Jiang
    Wu, Sheng-Yong
    Niu, Jing-Bo
    Zhang, Zhuo-Li
    Wang, Henry Z.
    Zhao, Yan-E
    Chai, Xue
    Zhou, Chang-Sheng
    Lu, Guang-Ming
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (01) : 23 - 30
  • [8] A dual-energy subtraction technique for microcalcification imaging in digital mammography - A signal-to-noise analysis
    Lemacks, MR
    Kappadath, SC
    Shaw, CC
    Liu, XM
    Whitman, GJ
    MEDICAL PHYSICS, 2002, 29 (08) : 1739 - 1751
  • [9] Evaluation of image quality and radiation dose of abdominal dual-energy CT
    Schmidt, David
    Soderberg, Marcus
    Nilsson, Mats
    Lindvall, Hakan
    Christoffersen, Christina
    Leander, Peter
    ACTA RADIOLOGICA, 2018, 59 (07) : 845 - 852
  • [10] Evaluation of exposure factors of dual-energy contrast-enhanced mammography to optimize radiation dose with improved image quality
    Niroshani, Sachila
    Nakamura, Tokiko
    Michiru, Nikaidou
    Negishi, Toru
    ACTA RADIOLOGICA OPEN, 2022, 11 (08)