Sorting a large set of heavily used LiF:Mg,Ti thermoluminescent detectors into repeatable subsets of similar response

被引:8
|
作者
Kearfott, Kimberlee J. [1 ]
Newton, Jill P. [1 ]
Rafique, Muhammad [1 ,2 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
[2] Univ Azad Jammu & Kashmir Muzaffarbad, Dept Phys, Azad Kashmir 13100, Pakistan
关键词
Thermoluminescent dosimeters; Cs-137; source; Sensitivity; Statistics; Calibration; PEAK FADING PROPERTIES; DOSIMETERS; PERFORMANCE; CALIBRATION; READINGS; DESIGN; SYSTEM; DRIFT; CT;
D O I
10.1016/j.apradiso.2014.10.019
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A set of 920 heavily used LiF:Mg,Ti thermoluminescent dosimeters (TLDs) was placed into a polymethyl methacrylate (PMMA) plate attached to a 40 x 40 x 15 cm(3) PMMA phantom and irradiated to 4.52 mGy using a Cs-137 source. This was repeated three times to determine the mean and standard deviation of each TLD's sensitivity. Reader drift was tracked over time with 10 control dosimeters. Two test sets of 100 TLDs were divided into subsets with sensitivities within +/- 1% of their subset means. All dosimeters were re-irradiated four times to test the TLDs' response repeatability and determine the sensitivity uniformity within the subsets. Coefficients of variation revealed that, within a given subset, the dosimeters responded within +/- 2.5% of their subset mean in all calibrations. The coefficient of variation in any of the 200 TLDs' calibrations was below 6% across the four calibrations. The work validates the approach of performing three calibrations to separate heavily used and aged TLDs with overall sensitivity variations of +/- 25% into subsets that reproducibly respond within +/- 2.5%. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:180 / 187
页数:8
相关论文
共 37 条
  • [31] Application of phototransferred thermoluminescence (PTTL) for dose re-assessment in routine dosimetry using MTS-N (LiF:Mg,Ti) thermoluminescent detectors
    Sas-Bieniarz, A.
    Budzanowski, M.
    Bubak, A.
    Kopec, R.
    RADIATION MEASUREMENTS, 2014, 71 : 447 - 450
  • [32] ENERGY RESPONSE OF DIFFERENT TYPES OF RADOS PERSONAL DOSEMETERS WITH MTS-N (LiF:Mg,Ti) AND MCP-N (LiF:Mg,Cu,P) TL DETECTORS
    Obryk, B.
    Hranitzky, C.
    Stadtmann, H.
    Budzanowski, M.
    Olko, P.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 211 - 214
  • [33] Design and feasibility of a multi-detector neutron spectrometer for radiation protection applications based on thermoluminescent 6LiF:Ti,Mg (TLD-600) detectors
    Lis, M.
    Gomez-Ros, J. M.
    Bedogni, R.
    Delgado, A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 584 (01): : 196 - 203
  • [34] Study of CaSO4: Dy and LiF:Mg,Ti detectors TL response to electron radiation using a SW Solid Water phantom
    Nunes, M. G.
    Campos, L. L.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 459 - 462
  • [35] Energy response of LiF:Mg,Ti thermoluminescent dosimeters to moderately filtered X-ray spectra in the range of 20 to 250 KV relative to 60Co
    Nunn, A.
    Davis, S.
    DeWerd, L.
    Micka, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2193 - 2194
  • [36] Experimental determination of the photon-energy dependent dose-to-water response of TLD600 and TLD700 (LiF:Mg,Ti) thermoluminescence detectors
    Schwahofer, Andrea
    Feist, Harald
    Georg, Holger
    Haering, Peter
    Schlegel, Wolfgang
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (01): : 13 - 20
  • [37] The relative efficiency of 7LiF:Mg,Ti (MTS-7) and 7LiF:Mg,Cu,P (MCP-7) thermoluminescent detectors for muons, pions and kaons over a broad energy range (2 keV-1 GeV): theoretical calculations using the Microdosimetric d(z) Model
    Parisi, Alessio
    Struelens, Lara
    Vanhavere, Filip
    RADIATION PHYSICS AND CHEMISTRY, 2020, 177