Sorting a large set of heavily used LiF:Mg,Ti thermoluminescent detectors into repeatable subsets of similar response

被引:8
|
作者
Kearfott, Kimberlee J. [1 ]
Newton, Jill P. [1 ]
Rafique, Muhammad [1 ,2 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
[2] Univ Azad Jammu & Kashmir Muzaffarbad, Dept Phys, Azad Kashmir 13100, Pakistan
关键词
Thermoluminescent dosimeters; Cs-137; source; Sensitivity; Statistics; Calibration; PEAK FADING PROPERTIES; DOSIMETERS; PERFORMANCE; CALIBRATION; READINGS; DESIGN; SYSTEM; DRIFT; CT;
D O I
10.1016/j.apradiso.2014.10.019
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A set of 920 heavily used LiF:Mg,Ti thermoluminescent dosimeters (TLDs) was placed into a polymethyl methacrylate (PMMA) plate attached to a 40 x 40 x 15 cm(3) PMMA phantom and irradiated to 4.52 mGy using a Cs-137 source. This was repeated three times to determine the mean and standard deviation of each TLD's sensitivity. Reader drift was tracked over time with 10 control dosimeters. Two test sets of 100 TLDs were divided into subsets with sensitivities within +/- 1% of their subset means. All dosimeters were re-irradiated four times to test the TLDs' response repeatability and determine the sensitivity uniformity within the subsets. Coefficients of variation revealed that, within a given subset, the dosimeters responded within +/- 2.5% of their subset mean in all calibrations. The coefficient of variation in any of the 200 TLDs' calibrations was below 6% across the four calibrations. The work validates the approach of performing three calibrations to separate heavily used and aged TLDs with overall sensitivity variations of +/- 25% into subsets that reproducibly respond within +/- 2.5%. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:180 / 187
页数:8
相关论文
共 37 条
  • [21] Characterisation of thermoluminescent LiF:Mg,Ti (MTS-N) detectors in the 85-550 K temperature range
    Briand, D
    Vincellér, S
    Lacconi, P
    Berkane-Krachai, A
    Bilski, P
    Benabdesselam, M
    Lapraz, D
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 297 - 300
  • [22] Long-term environment monitoring based on MTS-N (LiF:Mg, Ti) and MCP-N (LiF:Mg, Cu, P) thermoluminescent detectors
    Budzanowski, M
    Olko, P
    Obryk, B
    Ryba, E
    Nowak, A
    RADIATION MEASUREMENTS, 2004, 38 (4-6) : 821 - 824
  • [23] DOSE DEPENDENCE OF RADIATION INDUCED DAMAGE IN THE THERMOLUMINESCENT RESPONSE OF LIF:Mg,Ti (TLD-100)
    Reshes, G.
    Biderman, S.
    Horowitz, Y. S.
    Oster, L.
    Eliyahu, I.
    Ginsburg, D.
    Shapiro, A.
    Assor, Y.
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (02) : 232 - 237
  • [24] Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (<1 MeV)
    Tedgren, Asa Carlsson
    Hedman, Angelica
    Grindborg, Jan-Erik
    Carlsson, Gudrun Alm
    MEDICAL PHYSICS, 2011, 38 (10) : 5539 - 5550
  • [25] Measurements and Monte Carlo simulations of the response of the RADOS personal dosemeters with MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P) thermoluminescent detectors to X- and gamma-rays
    Obryk, B.
    Cywicka-Jakiel, T.
    Budzanowski, M.
    Bilski, P.
    Hranitzky, C.
    Olko, P.
    Stadtmann, H.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 616 - 620
  • [26] THERMOLUMINESCENT RESPONSE OF CAF2-DY AND LIF-MG,TI UNDER HEAVY-ION BOMBARDMENT
    FAIN, J
    MONTRET, M
    SAHRAOUI, L
    NUCLEAR INSTRUMENTS & METHODS, 1980, 175 (01): : 37 - 39
  • [27] THERMOLUMINESCENT RESPONSE OF LIF-MG,TI (TLD-100) AS A FUNCTION OF IRRADIATION TEMPERATURE FOR VARIOUS TYPES OF RADIATION
    KITIS, G
    CHARALAMBOUS, S
    TUYN, JWN
    RADIATION PROTECTION DOSIMETRY, 1992, 40 (02) : 117 - 122
  • [28] LIGHT SENSITIVITY OF LIF-MG,TI,CAF2-MN AND VARIOUS AL2O3 THERMOLUMINESCENT DETECTORS
    OSVAY, M
    RANOGAJECKOMOR, M
    GOLDER, F
    KERNENERGIE, 1991, 34 (03): : 116 - 118
  • [29] Measurements of the optical density and the thermoluminescent response of LiF:Mg,Ti exposed to high doses of 60Co gamma rays
    Montano-Garcia, C.
    Gamboa-deBuen, I.
    RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) : 230 - 232
  • [30] Deconvolution study on the glow curve structure of LiF:Mg,Ti and LiF:Mg,Cu,P thermoluminescent detectors exposed to 1H, 4He and 12C ion beams
    Parisi, Alessio
    Van Hoey, Olivier
    Megret, Patrice
    Vanhavere, Filip
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2017, 407 : 222 - 229