Application of phototransferred thermoluminescence (PTTL) for dose re-assessment in routine dosimetry using MTS-N (LiF:Mg,Ti) thermoluminescent detectors

被引:7
|
作者
Sas-Bieniarz, A. [1 ]
Budzanowski, M. [1 ]
Bubak, A. [1 ]
Kopec, R. [1 ]
机构
[1] Polish Acad Sci IFJ PAN, Inst Nucl Phys, PL-31342 Krakow, Poland
关键词
Dose reassessment; Phototransferred thermoluminescence (PTTL); MTS-N; (LiF:Mg; Ti);
D O I
10.1016/j.radmeas.2014.09.001
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Erasure of the thermoluminescence (TL) signal on detector readout is considered to be a disadvantage of TL dosimetry, as post-readout dose reassessment is then impossible in principle. A method of dose reassessment based on phototransferred thermoluminescence (PTTL) has been developed at the Institute of Nuclear Physics, Polish Academy of Sciences (IFJ PAN) and applied to MTS-N (LiF:Mg,Ti) detectors. We demonstrate the possibility of applying PTTL for dose reassessment in MTS-N TL detectors routinely applied in the dosimetric service at IFJ PAN. Readings of TL detectors exposed to relatively high doses by the customers of our dosimetry service can now be reassessed using our automatic readers. A major obstacle in applying the PTTL method at lower exposures is the presence of residual dose accumulated in LiF:Mg,Ti detectors after many field exposure and readout cycles. Since most of the TL detectors in our service have been already used for a long time (e.g. for over 10 years in the case of some detector batches), we find that our PTTL method of dose reassessment is possible only in detectors which had received doses exceeding 5 mSv. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:447 / 450
页数:4
相关论文
共 15 条
  • [1] Dose reassessment by using PTTL method in MTS-N (LiF:Mg, Ti) thermoluminescent detectors
    Budzanowski, M.
    Sas-Bieniarz, A.
    Bilski, P.
    Bubak, A.
    Kopec, R.
    RADIATION MEASUREMENTS, 2013, 56 : 389 - 392
  • [2] Application of individually calibrated solid LiF:Mg,Ti (MTS-N) detectors in clinical dosimetry
    Waligórski, MPR
    Lesiak, J
    Bubula, E
    Byrski, E
    Ryba, E
    Olko, P
    Bilski, P
    RADIATION PROTECTION DOSIMETRY, 1999, 85 (1-4) : 377 - 380
  • [3] Characterisation of thermoluminescent LiF:Mg,Ti (MTS-N) detectors in the 85-550 K temperature range
    Briand, D
    Vincellér, S
    Iacconi, P
    Berkane-Krachai, A
    Bilski, P
    Benabdesselam, M
    Lapraz, D
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 297 - 300
  • [4] Characterisation of thermoluminescent LiF:Mg,Ti (MTS-N) detectors in the 85-550 K temperature range
    Briand, D
    Vincellér, S
    Lacconi, P
    Berkane-Krachai, A
    Bilski, P
    Benabdesselam, M
    Lapraz, D
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 297 - 300
  • [5] Supralinearity of peak 4 and 5 in thermoluminescent lithium fluoride MTS-N (LiF : Mg, Ti) detectors at different Mg and Ti concentration
    Olko, P
    Bilski, P
    Budzanowski, M
    Molokanov, A
    Ochab, E
    Waligorski, MPR
    RADIATION MEASUREMENTS, 2001, 33 (05) : 807 - 812
  • [6] Long-term environment monitoring based on MTS-N (LiF:Mg, Ti) and MCP-N (LiF:Mg, Cu, P) thermoluminescent detectors
    Budzanowski, M
    Olko, P
    Obryk, B
    Ryba, E
    Nowak, A
    RADIATION MEASUREMENTS, 2004, 38 (4-6) : 821 - 824
  • [7] Dose re-evaluation in personal dosimetry using the phototransferred thermoluminescence method of LiF:Mg,Cu,Si TLD
    Kim, Hyoungtaek
    Kim, Min Chae
    Lee, Jungil
    Chang, Insu
    Lee, Seung Kyu
    Kim, Jang-Lyul
    RADIATION MEASUREMENTS, 2018, 118 : 20 - 25
  • [8] Mailed dosimetric audit of therapeutic proton beams using thermoluminescence MTS-N (LiF:Mg,Ti) powder - First results
    Kunst, J.
    Kopec, R.
    Kukolowicz, P.
    Mojeszek, N.
    Sadowski, B.
    Stolarczyk, L.
    Slusarczyk-Kacprzyk, W.
    Tobola, A.
    Olko, P.
    RADIATION MEASUREMENTS, 2017, 106 : 312 - 314
  • [9] ENERGY RESPONSE OF DIFFERENT TYPES OF RADOS PERSONAL DOSEMETERS WITH MTS-N (LiF:Mg,Ti) AND MCP-N (LiF:Mg,Cu,P) TL DETECTORS
    Obryk, B.
    Hranitzky, C.
    Stadtmann, H.
    Budzanowski, M.
    Olko, P.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 211 - 214
  • [10] Measurements and Monte Carlo simulations of the response of the RADOS personal dosemeters with MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P) thermoluminescent detectors to X- and gamma-rays
    Obryk, B.
    Cywicka-Jakiel, T.
    Budzanowski, M.
    Bilski, P.
    Hranitzky, C.
    Olko, P.
    Stadtmann, H.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 616 - 620