Intra-individual Comparison of Average Glandular Dose of Two Digital Mammography Units using Different Anode/Filter Combinations

被引:3
|
作者
Engelken, Florian J. F. [1 ]
Meyer, Henning [1 ]
Juran, Ralf [1 ]
Bick, Ulrich [1 ]
Fallenberg, Eva [1 ]
Diekmann, Felix [1 ]
机构
[1] Charite Univ Med Berlin, Inst Radiol, D-10117 Berlin, Germany
关键词
Breast; mammography; radiation dosage; radiation quality; tungsten; rhodium; FILM MAMMOGRAPHY; QUALITY; PERFORMANCE; REDUCTION; MORTALITY; CONTRAST; RATIO;
D O I
10.1016/j.acra.2009.05.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The aim of this study was to compare average glandular dose (AGD) in two full-field digital mammography units using different anode/filter combinations. Materials and Methods. Mammographies of 50 consecutive patients on a mammography system using a tungsten/rhodium (W/Rh) anode/filter combination were retrospectively compared to prior examinations on a different mammography unit using combinations of Molybdenum (Mo) and Rhodium (Rh). To exclude effects of increasing patient age, two prior examinations within 5 years were used. Both views of one breast were chosen for analysis. AGD was recorded as stated by each mammography unit. Accuracy of stated AGD and contrast-detail resolution were assessed using different breast phantoms. Results. The mean AGDs from the examinations using W/Rh were 0.95 mGy and 1.01 mGy for craniocaudal (CC) and mediolateral oblique (MLO) views compared to 1.51 mGy and 1.54 mGy, respectively, using combinations of Mo and Rh (P < .001). Relative reduction of AGD was independent of breast thickness but decreased with increasing breast density (partial correlation coefficient of 0.46, P < .005 and 0.57, P < .001, for CC and MLO views, respectively). Low-level contrast resolution was equal in both units using standard acquisition parameters. Conclusion. In clinical mammographies, higher energy beam spectra obtained using W/Rh anode/filter combinations may significantly contribute to lowering AGD compared to Mo/Mo, Mo/Rh, and Rh/Rh in breasts that are not extremely dense.
引用
收藏
页码:1272 / 1280
页数:9
相关论文
共 40 条
  • [1] Comparison of Anode/Filter Combinations in Digital Mammography with Respect to the Average Glandular Dose
    Uhlenbrock, D. F.
    Mertelmeier, T.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (03): : 249 - 254
  • [2] Effect of anode/filter combination on average glandular dose in mammography
    Biegala, Michal
    Jakubowska, Teresa
    Markowska, Karolina
    JOURNAL OF MEDICAL PHYSICS, 2015, 40 (01) : 45 - 51
  • [3] Assessment of Mean Glandular Dose in Mammography System with Different Anode-Filter Combinations Using MCNP Code
    Gholamkar, Lida
    Mowlavi, Ali Asghar
    Sadeghi, Mahdi
    Athari, Mitra
    IRANIAN JOURNAL OF RADIOLOGY, 2016, 13 (04)
  • [4] Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography
    Varjonen, Mari
    Stroemmer, Pekka
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [5] Effect of filter on average glandular dose and image quality in digital mammography
    Songsaeng, C.
    Krisanachinda, A.
    Theerakul, K.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [6] Determination of average glandular dose with modern mammography units for two large groups of patients
    Klein, R
    Aichinger, H
    Dierker, J
    Jansen, JTM
    JoiteBarfuss, S
    Sabel, M
    SchulzWendtland, R
    Zoetelief, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (04): : 651 - 671
  • [7] Average glandular dose in digital mammography and digital breast tomosynthesis: comparison of phantom and patient data
    Bouwman, R. W.
    van Engen, R. E.
    Young, K. C.
    den Heeten, G. J.
    Broeders, M. J. M.
    Schopphoven, S.
    Jeukens, C. R. L. P. N.
    Veldkamp, W. J. H.
    Dance, D. R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (20): : 7893 - 7907
  • [8] Analysis of influence of digital mammography on the individual average glandular dose against screen-film system
    Azuma, Y.
    Goto, S.
    Sumimoto, T.
    Takeda, Y.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1541 - 1550
  • [9] Optimizing the target-filter combination in digital mammography in the sense of image quality and average glandular dose
    Varjonen, Mari
    Strommer, Pekka
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 570 - 576
  • [10] Image Quality and Radiation Dose for Fibrofatty Breast using Target/filter Combinations in Two Digital Mammography Systems
    Alkhalifah, Khaled
    Asbeutah, Akram
    Brindhaban, Ajit
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2020, 10