Effect of anode/filter combination on average glandular dose in mammography

被引:13
|
作者
Biegala, Michal [1 ,2 ]
Jakubowska, Teresa [1 ,2 ]
Markowska, Karolina [3 ]
机构
[1] Med Univ Lodz, Fac Biomed Sci & Postgrad Training, Dept Med Imaging Technol, PL-90251 Lodz, Poland
[2] Copernicus Mem Hosp Lodz, Reg Canc Ctr, Dept Med Phys, Lodz, Poland
[3] Lodz Univ Technol Fdn, Laser Diagnost & Therapy Ctr, Lodz, Poland
关键词
Automatic exposure control system; average glandular dose; digital mammography; threshold contrast visibility;
D O I
10.4103/0971-6203.152251
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A comparative analysis of the mean glandular doses was conducted in 100 female patients who underwent screening mammography in 2011 and 2013. Siemens Mammomat Novation with the application of the W/Rh anode/filter combination was used in 2011, whereas in 2013 anode/filter combination was Mo/Mo or Mo/Rh. The functioning of mammography was checked and the effectiveness of the automatic exposure control (AEC) system was verified by measuring compensation of changes in the phantom thickness and measuring tube voltage. On the base of exposure parameters, an average glandular dose for each of 100 female patients was estimated. The images obtained by using AEC system had the acceptable threshold contrast visibility irrespective of the applied anode/filter combination. Mean glandular doses in the females, examined with the application of the W/Rh anode/filter combination, were on average 23.6% lower than that of the Mo/Mo or Mo/Rh anode/filter combinations. It is recommended to use a combination of the W/Rh anode /filter which exhibited lower mean glandular doses.
引用
收藏
页码:45 / 51
页数:7
相关论文
共 50 条
  • [1] Optimizing the anode-filter combination in the sense of image quality and average glandular dose in digital mammography
    Varjonen, Mari
    Stroemmer, Pekka
    [J]. MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [2] Comparison of Anode/Filter Combinations in Digital Mammography with Respect to the Average Glandular Dose
    Uhlenbrock, D. F.
    Mertelmeier, T.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (03): : 249 - 254
  • [3] Effect of filter on average glandular dose and image quality in digital mammography
    Songsaeng, C.
    Krisanachinda, A.
    Theerakul, K.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [4] Optimizing the target-filter combination in digital mammography in the sense of image quality and average glandular dose
    Varjonen, Mari
    Strommer, Pekka
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 570 - 576
  • [5] NORMALIZED AVERAGE GLANDULAR DOSE IN MAGNIFICATION MAMMOGRAPHY
    LIU, B
    GOODSITT, M
    CHAN, HP
    [J]. RADIOLOGY, 1995, 197 (01) : 27 - 32
  • [6] Patient Specific Average Glandular Dose in Mammography
    Patel, V.
    Highnam, R.
    Tromans, C.
    Pizzutiello, R. J.
    Destounis, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Method for the evaluation of a average glandular dose in mammography
    Okunade, AA
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1153 - 1164
  • [8] Intra-individual Comparison of Average Glandular Dose of Two Digital Mammography Units using Different Anode/Filter Combinations
    Engelken, Florian J. F.
    Meyer, Henning
    Juran, Ralf
    Bick, Ulrich
    Fallenberg, Eva
    Diekmann, Felix
    [J]. ACADEMIC RADIOLOGY, 2009, 16 (10) : 1272 - 1280
  • [9] NORMALIZED AVERAGE GLANDULAR DOSE IN MOLYBDENUM TARGET RHODIUM FILTER AND RHODIUM TARGET RHODIUM FILTER MAMMOGRAPHY
    WU, XZ
    GINGOLD, EL
    BARNES, GT
    TUCKER, DM
    [J]. RADIOLOGY, 1994, 193 (01) : 83 - 89
  • [10] INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY
    DESPONDS, L
    DEPEURSINGE, C
    GRECESCU, M
    HESSLER, C
    SAMIRI, A
    VALLEY, JF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09): : 1165 - 1182