Objective: To compare BacT/ALERT (BTA) and BACTEC 9240 (BAC), two continuously monitoring automated blood culture systems, for the recovery of bloodstream pathogens and standard media available for these systems. Materials and Methods: Blood samples from 100 adults and 50 paediatric patients suspected of having bloodstream infections were inoculated at the bedside into non-vented BTA and BAC standard blood culture bottles and incubated in their respective instruments. The time to growth detection (TD) was recorded for each bottle that became positive. A quantitative assay was also carried out with 5 standard bloodstream pathogens to assess TD of each pathogen as well as the quantity of organisms recovered. Results: A total of 23 isolates representing true infections were recovered by both BTA and BAC bottles, indicating a blood culture positivity rate of 15.3%, 18 (78.3%) by BTA bottles and 13 (56.5%) by BAC. Proteus mirabilis, Pseudomonas aeruginosa and Clostridium perfringens were recovered only by the BTA system. The average TDs were 19.0 and 24.6 h for BTA and BAC, respectively. Analysis of the quantitative growth of known pathogens in both systems was more or less the same for Staphylococcus aureus, Escherichia coli and P. aeruginosa but slightly different for Haemophilus influenzae and Streptococcus pneumoniae. The anaerobic bottle of the BTA did not support the growth of H. influenzae below an inoculum of 10(10) CFU/ml whereas the BAC did so at a lower inoculum of 10(8) CFU/ml. TD for S. pneumoniae in the BTA was about half of that in the BAC. Conclusions: The BTA system appears to be more efficient in detecting common bloodstream pathogens as a higher inoculum is needed for the BAC system to detect the same organism. Copyright (C) 2006 S. Karger AG, Basel.
引用
收藏
页码:223 / 227
页数:5
相关论文
共 50 条
  • [41] Evaluation of the BACT/ALERT 3D system for testing susceptibility of Mycobacterium tuberculosis to first-line antituberculosis agents:: comparison with BacTec MGIT 960 and Lowenstein-Jensen proportion method
    Guney, C.
    Suel, A.
    Karagoz, T.
    Yorulmaz, Y.
    INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2007, 29 : S463 - S463
  • [42] Evaluation of a new generation culture bottle with the BacT/ALERT® 3D culture system with 8 contaminating organisms in platelets
    Brecher, ME
    Heath, D
    Hay, S
    Rothenberg, S
    Stutzman, LC
    TRANSFUSION, 2001, 41 (09) : 33S - 33S
  • [43] Evaluation of BacT/Alert 3D SA bottles for accurate detection of mycobacteremia with special-reference to Mycobacterium abscessus
    Kasuga, E.
    Matsumoto, T.
    Oana, K.
    Shiohara, M.
    Okabe, T.
    Yamauchi, K.
    Honda, T.
    Ota, H.
    Kawakami, Y.
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2007, 12 (01) : 43 - 46
  • [44] Effect of Blood Volume in Standard Anaerobic Blood Culture Bottles of the BacT/ALERT 3D System Used for the Detection of Pathogens and Time to Detection
    Kim, Seong Chun
    Kim, Sunjoo
    Lee, Dong-Hyun
    Choi, Sae-Rom
    Kim, Jeong-Sook
    PLOS ONE, 2015, 10 (02):
  • [45] Comparative evaluation of Lowenstein-Jensen proportion method, BacT/ALERT 3D system, and enzymatic pyrazinamidase assay for pyrazinamide susceptibility testing of Mycobacterium tuberculosis
    Singh, Pushpendra
    Wesley, Clement
    Jadaun, G. P. S.
    Malonia, Sunil Kumar
    Das, R.
    Upadhyay, P.
    Faujdar, J.
    Sharma, P.
    Gupta, P.
    Mishra, Abhay Kumar
    Singh, Kalpana
    Chauhan, D. S.
    Sharma, V. D.
    Gupta, U. D.
    Mishra, Abhay Kumar
    Singh, Kalpana
    Chauhan, D. S.
    Sharma, V. D.
    Gupta, U. D.
    Venkatesan, K.
    Katoch, V. A.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (01) : 76 - 80
  • [46] Validation of the BacT/ALERT 3D System for Sterility Testing 18F-FDG
    Woods, Michael
    Frantzen, Kari
    Benard, Francois
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [47] Evaluation of negative results of BacT/alert 3D automated blood culture system (vol 43, pg 257, 2005)
    Kocoglu, ME
    Bayram, A
    Balci, I
    JOURNAL OF MICROBIOLOGY, 2005, 43 (04) : 381 - 381
  • [48] BacT/ALERT 3D在罐头商业无菌检测中的应用
    郑晶
    黄晓蓉
    陈彬
    汤敏英
    林杰
    张永祥
    食品安全导刊, 2008, (04) : 50 - 52
  • [49] Monitoring of apheresis platelet bacterial contamination with the BacT/ALERT 3D - A university experience.
    Brecher, ME
    Hay, SN
    Rothenberg, S
    TRANSFUSION, 2002, 42 (09) : 11S - 11S
  • [50] Evaluation of BacT/Alert 3D liquid culture system for recovery of mycobacteria from clinical specimens using sodium dodecyl (lauryl) sulfate-NaOH decontamination
    Carricajo, A
    Fonsale, N
    Vautrin, AC
    Aubert, G
    JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (10) : 3799 - +