An industrial design of a pulsed field magnetometer (PFM) which generates in a bore of 30 nun a maximum field of 5 tesla with two different pulse durations (40 and 57 ms) is described. The field is calibrated using a small pickup coil where the effective winding area is known from a nuclear magnetic resonance calibration. The magnetization is calibrated using Fe3O4, an industrial ferrite, but also Fe and Ni.
引用
收藏
页码:2982 / 2984
页数:3
相关论文
共 50 条
  • [42] Magnetometer Calibration Scheme for Quadrotors with On-Board Magnetic Field of Multiple DC Motors
    Liu, Haiwei
    Liu, Ming
    Ge, Yunjian
    Shuang, Feng
    PROCEEDINGS OF 2013 CHINESE INTELLIGENT AUTOMATION CONFERENCE: INTELLIGENT AUTOMATION & INTELLIGENT TECHNOLOGY AND SYSTEMS, 2013, 255 : 409 - 419
  • [43] Co-estimation of geomagnetic field and in-orbit fluxgate magnetometer calibration parameters
    Patrick Alken
    Nils Olsen
    Christopher C. Finlay
    Earth, Planets and Space, 72
  • [44] Measurement accuracy of a pulsed field magnetometer designed for rare earth based permanent magnets
    Song, MS
    Kim, YB
    Kim, CS
    Kim, TK
    IEEE TRANSACTIONS ON MAGNETICS, 2000, 36 (05) : 3637 - 3639
  • [45] A High Precision Pulsed Field Magnetometer for Magnetic Properties Measurements of Rare Earth Magnets
    Kim, Y. B.
    Kapustin, G. A.
    JOURNAL OF THE KOREAN MAGNETICS SOCIETY, 2005, 15 (04): : 250 - 255
  • [46] Co-estimation of geomagnetic field and in-orbit fluxgate magnetometer calibration parameters
    Alken, Patrick
    Olsen, Nils
    Finlay, Christopher C.
    EARTH PLANETS AND SPACE, 2020, 72 (01):
  • [47] In flight magnetometer calibration in the projectile frame
    Adam, Ronan
    Combettes, Christophe
    Pecheur, Emmanuel
    Changey, Sebastien
    2018 5TH IEEE INTERNATIONAL SYMPOSIUM ON INERTIAL SENSORS & SYSTEMS (INERTIAL 2018), 2018, : 49 - 52
  • [48] High frequency magnetization dynamics metrology using a pulsed field inductive microwave magnetometer
    Vopson, M. M.
    Lees, K.
    Hall, M.
    Cain, M. G.
    Stewart, M.
    Tran, Y.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (01)
  • [49] THE PULSED HIGH-FIELD APPARATUS AND MAGNETOMETER FOR INVESTIGATIONS OF HARD MAGNETIC-MATERIALS
    JUSZCZYK, S
    DUDA, H
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1984, 44 (1-2) : 133 - 138
  • [50] CALIBRATION OF SPINNER MAGNETOMETER WITH WIRE LOOP
    NOLTIMIER, HC
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1964, 41 (01): : 55 - &