Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:17
|
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [1] Improvement in PET/CT Image Quality with a Combination of Point-Spread Function and Time-of-Flight in Relation to Reconstruction Parameters
    Akamatsu, Go
    Ishikawa, Kaori
    Mitsumoto, Katsuhiko
    Taniguchi, Takafumi
    Ohya, Nobuyoshi
    Baba, Shingo
    Abe, Koichiro
    Sasaki, Masayuki
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (11) : 1716 - 1722
  • [2] A combined nonlinear contrast image reconstruction algorithm under inexact point-spread function
    Yu. E. Voskoboinikov
    Optoelectronics, Instrumentation and Data Processing, 2007, 43 (6) : 489 - 499
  • [3] A Combined Nonlinear Contrast Image Reconstruction Algorithm under Inexact Point-Spread Function
    Voskoboinikov, Yu. E.
    OPTOELECTRONICS INSTRUMENTATION AND DATA PROCESSING, 2007, 43 (06) : 489 - 499
  • [4] Impact of Point-Spread Function Modeling on PET Image Quality in Integrated PET/MR Hybrid Imaging
    Aklan, Bassim
    Oehmigen, Mark
    Beiderwellen, Karsten
    Ruhlmann, Marcus
    Paulus, Daniel H.
    Jakoby, Bjoern W.
    Ritt, Philipp
    Quick, Harald H.
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (01) : 78 - 84
  • [5] Impact of Point-Spread Function Reconstruction on Dynamic and Static FDOPA PET/CT Quantitative Parameters in Glioma
    Girard, A.
    Madani, F.
    Chaboub, N.
    Le Reste, P.
    Devillers, A.
    Saint-Jalmes, H.
    Le Jeune, F.
    Palard-Novello, X.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 48 (SUPPL 1) : S238 - S239
  • [6] Benefits of Point-Spread Function and Time of Flight for PET/CT Image Quality in Relation to the Body Mass Index and Injected Dose
    Akamatsu, Go
    Mitsumoto, Katsuhiko
    Ishikawa, Kaori
    Taniguchi, Takafumi
    Ohya, Nobuyoshi
    Baba, Shingo
    Abe, Koichiro
    Sasaki, Masayuki
    CLINICAL NUCLEAR MEDICINE, 2013, 38 (06) : 407 - 412
  • [7] Optimal reconstruction matrix and PET image filtration for point-spread function and time-of-flight reconstruction - A phantom study
    Ptacek, Jaroslav
    Karhan, Pavel
    Fiala, Petr
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 39 : 95 - 99
  • [8] Impact of Point-Spread Function Reconstruction on 68Ga-DOTATATE PET/CT Quantitative Imaging Parameters
    You, Helena
    Sanli, Yasemin
    Subramaniam, Rathan M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 213 (03) : 683 - 688
  • [9] Reconstruction of the Point-Spread Function of the Human Eye Using a New Phase-Retrieval Algorithm
    Yukio Iida
    Masahiro Shibutani
    Katsuhiko Kobayashi
    Kazuhiko Ohnuma
    Yoichi Miyake
    Toru Noda
    Optical Review, 2006, 13 : 87 - 92
  • [10] Reconstruction of the point-spread function of the human eye using a new phase-retrieval algorithm
    Iida, Yukio
    Shibutani, Masahiro
    Kobayashi, Katsuhiko
    Ohnuma, Kazuhiko
    Miyake, Yoichi
    Noda, Toru
    OPTICAL REVIEW, 2006, 13 (02) : 87 - 92