Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:17
|
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [31] Optimization of Reconstruction Parameters for Brain PET Studies with Time of Flight and Point Spread Function Modelling
    Prieto, E.
    Marti-Climent, J. M.
    Arbizu, J.
    Ribelles, M. J.
    Sancho, L.
    Quincoces, G.
    Richter, J. A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S172 - S172
  • [32] PARTICLE IMAGE VELOCIMETRY - IMAGE LABELING USING ENCODING OF THE POINT-SPREAD FUNCTION BY APPLICATION OF A POLARIZATION-SENSITIVE PUPIL MASK
    REEVES, M
    LAWSON, NJ
    HALLIWELL, NA
    COUPLAND, JM
    APPLIED OPTICS, 1995, 34 (01): : 194 - 200
  • [33] Deconvolution of the two-dimensional point-spread function of area detectors using the maximum-entropy algorithm
    Graafsma, H
    de Vries, RY
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1999, 32 : 683 - 691
  • [34] PET POINT SPREAD FUNCTION MODELING AND IMAGE DEBLURRING USING A PET/MRI JOINT ENTROPY PRIOR
    Dutta, Joyita
    El Fakhri, Georges
    Zhu, Xuping
    Li, Quanzheng
    2015 IEEE 12th International Symposium on Biomedical Imaging (ISBI), 2015, : 1423 - 1426
  • [35] Optimization of the Algorithm for the Implementation of Point Spread Function in the 3D-OSEM Reconstruction Algorithm Based on the List-Mode Micro PET Data
    Zhao, Jie
    Song, Yunfeng
    Liu, Qiong
    Chen, Shijie
    Chen, Jyh-Cheng
    ELECTRONICS, 2023, 12 (06)
  • [36] Evaluation of the spatial dependence of the point spread function in 2D PET image reconstruction using LOR-OSEM
    Wiant, D.
    Gersh, J. A.
    Bennett, M.
    Bourland, J. D.
    MEDICAL PHYSICS, 2010, 37 (03) : 1169 - 1182
  • [37] Reconstruction of 68-Ga small animal PET images using EM algorithm with system point spread function modelling
    Spinelli, A. E.
    D'Ambrosio, D.
    Di Pierro, D.
    Trespidi, S.
    Nanni, C.
    Ambrosini, V.
    Franchi, R.
    Boschi, S.
    Marengo, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S218 - S218
  • [38] Prospective evaluation of PET/CT with point spread function reconstruction for axillary nodal staging in breast cancer patients
    Besson, Florent
    Switsers, Odile
    Fournier, Cecile Blanc
    Heutte, Natacha
    Ciappuccini, Renaud
    Bardet, Stephane
    Aide, Nicolas
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [39] Impact of FDG PET/CT using point spread function (PSF) reconstruction compared with conventional FDG PET/CT and pelvic MRI for lymph node staging in rectal cancer
    Hotta, Masatoshi
    Minamimoto, Ryogo
    Yano, Hideaki
    Gouda, Yoshimasa
    Shuno, Yasutaka
    Yokoyama, Kota
    Tajima, Tsuyoshi
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [40] Bayesian penalized-likelihood reconstruction algorithm suppresses edge artifacts in PET reconstruction based on point-spread-function
    Yamaguchi, Shotaro
    Wagatsuma, Kei
    Miwa, Kenta
    Ishii, Kenji
    Inoue, Kazumasa
    Fukushi, Masahiro
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 47 : 73 - 79