Optimization of PET/CT image quality using the GE "Sharp IR' point-spread function reconstruction algorithm

被引:17
|
作者
Vennart, Nicholas J. [1 ,2 ]
Bird, Nicholas [1 ]
Buscombe, John [1 ]
Cheow, Heok K. [1 ]
Nowosinska, Ewa [1 ]
Heard, Sarah [1 ]
机构
[1] Cambridge Univ Hosp Fdn Trust, Dept Nucl Med, Cambridge, England
[2] Queen Elizabeth Hosp, Dept Med Phys, Gateshead NE9 6SX, England
关键词
contrast; Discovery; 690; image quality; PET; CT; point-spread function; signal-to-noise; time-of-flight; TIME-OF-FLIGHT; PERFORMANCE; SCANNER; IMPACT; IMPROVEMENT; NOISE; PSF;
D O I
10.1097/MNM.0000000000000669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveThe objective of this study was to quantify any improvement with the GE Sharp IR' point-spread function (PSF) reconstruction algorithm in addition to ordered subsets expectation maximum (OSEM) and time-of-flight (TOF) reconstruction algorithms and establish the optimum parameters to be used in clinical studies.Materials and methodsWe conducted a range of experiments using the National Electrical Manufacturers Association image quality phantom filled with a 4:1 signal-to-background ratio. We scanned the phantom using the GE Discovery 690 PET/CT scanner. We varied iteration number and Gaussian filtration. Results were compared for OSEM, OSEM+TOF and OSEM+TOF+PSF reconstructions. A sample of 15 whole-body fluorine-18-fluorodeoxyglucose were reconstructed with OSEM+TOF and OSEM+TOF+PSF using a selection of optimum reconstruction parameters determined in phantom studies. Clinicians qualitatively ranked their preferred images to choose optimum parameters.ResultsThe addition of PSF improved signal-to-noise ratios (SNRs), contrast, hot contrast recovery coefficients and noise over OSEM and OSEM+TOF reconstruction algorithms. SNRs were the highest at two iterations and with 0 or 2mm filters with OSEM+TOF+PSF reconstruction in all phantom studies. Clinicians generally favoured OSEM+TOF+PSF reconstruction with three iterations and a 2mm filter.ConclusionPSF reconstruction significantly improved image quality for both clinical and phantom studies. We recommended the optimum reconstruction parameters using three iterations, 24 subsets and a 2mm filter, which improved SNRs by up to 28.8% for small lesions (P<0.05).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 50 条
  • [21] High resolution image reconstruction (point spread function based) in FDG brain PET
    Reimold, Matthias
    Pfannenberg, Christina
    Kupferschlaeger, Jurgen
    Bares, Roland
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [22] Applications of both time of flight and point spread function in brain PET image reconstruction
    Shao, Xiaonan
    Shao, Xiaoliang
    Wang, Xiaosong
    Wang, Yuetao
    NUCLEAR MEDICINE COMMUNICATIONS, 2016, 37 (04) : 422 - 427
  • [23] A simple table lookup method for PET/CT partial volume correction using a point-spread function in diagnosing lymph node metastasis
    Yuichi Sakaguchi
    Noriko Mizoguchi
    Tatsuya Mitsumoto
    Katsuhiko Mitsumoto
    Kazuhiko Himuro
    Nobuyoshi Ohya
    Koichiro Kaneko
    Shingo Baba
    Koichiro Abe
    Yoshihiko Onizuka
    Masayuki Sasaki
    Annals of Nuclear Medicine, 2010, 24 : 585 - 591
  • [24] Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PET
    Rapisarda, E.
    Bettinardi, V.
    Thielemans, K.
    Gilardi, M. C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (14): : 4131 - 4151
  • [25] The edge artifact in the point-spread function-based PET reconstruction at different sphere-to-background ratios of radioactivity
    Daisuke Kidera
    Ken Kihara
    Go Akamatsu
    Shohei Mikasa
    Takafumi Taniguchi
    Yuji Tsutsui
    Toshiki Takeshita
    Akira Maebatake
    Kenta Miwa
    Masayuki Sasaki
    Annals of Nuclear Medicine, 2016, 30 : 97 - 103
  • [26] The edge artifact in the point-spread function-based PET reconstruction at different sphere-to-background ratios of radioactivity
    Kidera, Daisuke
    Kihara, Ken
    Akamatsu, Go
    Mikasa, Shohei
    Taniguchi, Takafumi
    Tsutsui, Yuji
    Takeshita, Toshiki
    Maebatake, Akira
    Miwa, Kenta
    Sasaki, Masayuki
    ANNALS OF NUCLEAR MEDICINE, 2016, 30 (02) : 97 - 103
  • [27] Influence of time of flight and point spread function on PET / CT image quality and standard uptake value: A phantom study
    Huang, Kemin
    Feng, Yanlin
    Liu, Dejun
    Liang, Weitang
    Li, Lin
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [28] Upscaling image resolution of compact imaging systems using wavefront coding and a property of the point-spread function
    Lee, Sang-Hyuck
    Park, No-Cheol
    Park, Kyoung-Su
    Park, Young-Pil
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2010, 27 (10) : 2304 - 2312
  • [29] Impact of list-mode reconstruction and image-space point spread function correction on PET image contrast and quantitative value using SiPM-based PET/CT system
    Yuya Shirakawa
    Norikazu Matsutomo
    Radiological Physics and Technology, 2023, 16 : 384 - 396
  • [30] Impact of list-mode reconstruction and image-space point spread function correction on PET image contrast and quantitative value using SiPM-based PET/CT system
    Shirakawa, Yuya
    Matsutomo, Norikazu
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (03) : 384 - 396