Clinical optimization of filters in direct a-se FFDM (full field digital mammography) system

被引:0
|
作者
Uchiyama, Nachiko
Moriyama, Noriyuki
Kitagawa, Mayumi
Gomi, Shiho
Nagai, Yuichi
机构
[1] Natl Canc Ctr, Dept Canc Screening, Res Ctr Canc Screening, Chuo Ku, Tokyo 1040045, Japan
[2] Natl Canc Ctr, Dept Radiol, Res Ctr Canc Screening, Chuo Ku, Tokyo 1040045, Japan
来源
关键词
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We evaluated three combinations of filters (Mo/Mo, Mo/ Rh, and W /Rh) in direct a-Se FFDM system to optimize radiation dose clinically. We measured CNR (Contrast to Noise Ratio) as physical characteristics changing radiation dose and phantom thickness in clinical range. In 20, 30, 40, and 50mm PMMA phantoms, Mo / Mo showed the best performances. On the other hand, in 60 and 70 mm, W/Rh 30kV showed best performance. In addition, in 40 and 50mm PMMA phantoms, W/Rh 30kV showed the second best performance. In direct a-Se FFDM system, W/Rh was valuable in minimizing radiation dose.
引用
收藏
页码:315 / 323
页数:9
相关论文
共 50 条
  • [41] Full-field digital mammography designed as a complete system
    Muller, S
    EUROPEAN JOURNAL OF RADIOLOGY, 1999, 31 (01) : 25 - 34
  • [42] Physical characteristics of a full-field digital mammography system
    Suryanarayanan, S
    Karellas, A
    Vedantham, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 533 (03): : 560 - 570
  • [43] Breast positioning system for full field digital mammography and digital breast tomosynthesis system
    Varjonen, Mari
    Pamilo, Martti
    Hokka, Pirjo
    Hokkanen, Riina
    Strommer, Pekka
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [44] Local diagnostic reference levels (LDRLs) for full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) procedures in Morocco
    Talbi, Mohammed
    El Mansouri, M'hamed
    Nhila, Oussama
    Tahiri, Zakaria
    Eddaoui, Khalida
    Khalis, Mohammed
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2022, 53 (02) : 242 - 247
  • [45] A comparison of full-field digital mammography (FFDM) with digital breast tomosynthesis (DBT) in the detection of microcalcifications by correlating mammographic and pathologic findings
    Farghadani, Maryam
    Ghadiri-faraz, Bahar
    Riahinezhad, Maryam
    Sarami, Sahar
    IMMUNOPATHOLOGIA PERSA, 2024, 10 (02):
  • [46] Full Field Digital Mammography (FFDM) versus CMOS Technology, Specimen Radiography System (SRS) and Tomosynthesis (DBT) - Which System Can Optimise Surgical Therapy?
    Schulz-Wendtland, R.
    Dilbat, G.
    Bani, M.
    Fasching, P. A.
    Heusinger, K.
    Lux, M. P.
    Loehberg, C. R.
    Brehm, B.
    Hammon, M.
    Saake, M.
    Dankerl, P.
    Jud, S. M.
    Rauh, C.
    Bayer, C. M.
    Beckmann, M. W.
    Uder, M.
    Meier-Meitinger, M.
    GEBURTSHILFE UND FRAUENHEILKUNDE, 2013, 73 (05) : 422 - 427
  • [47] Full-field digital mammography
    Bick, U
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (12): : 957 - 964
  • [48] Full field digital mammography scanner
    Tesic, MM
    Piccaro, MF
    Munier, B
    EUROPEAN JOURNAL OF RADIOLOGY, 1999, 31 (01) : 2 - 17
  • [49] Full-field digital mammography
    U Bick
    Breast Cancer Research, 2 (Suppl 2)
  • [50] Benefits of full field digital mammography
    Mundt, Y
    PROCEEDINGS OF THE 23RD INTERNATIONAL CONGRESS OF RADIOLOGY OF THE INTERNATIONAL SOCIETY OF RADIOLOGY, 2004, : 295 - 297