In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair

被引:18
|
作者
Torsello, Giovanni Federico [1 ,2 ]
Herten, Monika [2 ,3 ]
Mueller, Markus [4 ]
Andre-Frank [4 ]
Torsello, Giovanni B. [2 ]
Austermann, Martin [2 ]
机构
[1] Charite Univ Med Berlin, Charite Campus Virchow Klinikum, Dept Diagnost & Intervent Radiol & Nucl Med, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] St Franziskus Hosp Munster, Dept Vasc Surg, Munster, Germany
[3] Essen Univ Hosp, Dept Orthoped & Trauma Surg, Essen, Germany
[4] Univ Hosp Munster, Dept Hand Trauma & Reconstruct Surg, Biomech Lab, Munster, Germany
关键词
bridging stent-graft; displacement force; fenestrated stent-graft; in vitro experiment; in vitro model; material fatigue; permeability; stent fracture; ENDOGRAFTS; COMPONENTS; OUTCOMES; SECURE;
D O I
10.1177/1526602819842569
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To evaluate the safety and integrity of a new stent-graft as a potential bridging device in fenestrated stent-grafts using an in vitro fenestrated model. Materials and Methods: Polyester test sheets with ten 6-mm- or 8-mm-diameter fenestrations were used to simulate a fenestrated main body endoprosthesis. In total, 50 Viabahn balloon-expandable (VBX) stent-grafts of varying lengths (29 and 39 mm) and diameters (6, 7, and 8 mm) were implanted in the fitting fenestrations. After release, the 6- and 7-mm-diameter stent-grafts were flared with a 10x20-mm angioplasty balloon; a 12x20-mm balloon was used in the 8-mm-diameter devices. Safety of the devices was defined as absence of fractures detected on radiography or computed tomography (CT), as well as material failure detected by microscopy and water permeability testing. The forces (in Newtons, N) needed for perpendicular dislocation (pullout force) and axial dislocation (shear stress force) were also evaluated. Results: Forty VBX stent-grafts were subjected to digital radiographic imaging and multiplanar CT. None showed any stent fracture. Subsequent microscopy indicated no damage to the fabric or separation of the graft after flaring. Ten VBX stent-grafts underwent water permeability testing after flaring; no water passed through the graft wall during a 10-minute period under an intraluminal pressure at 120 mm Hg. Testing of 25 VBX stent-grafts revealed initial pullout forces between 11.3 and 31 N. Shear stress tests showed that the average force needed to dislocate the stent-grafts by 50% of their diameter ranged between 5.75 and 6.91 N (mean 6.1 +/- 0.5 N) for the 6-mm stents and between 3.31 and 5.4 N (mean 4.4 +/- 0.8) for the 8-mm stents. Conclusion: This preliminary study demonstrated the applicability of the VBX as a bridging stent-graft in a simulated fenestration model. A comparison with other stent-grafts and clinical assessment are required.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 50 条
  • [41] Endovascular Stent-Graft Repair of a Tuberculous Mycotic Aortic Aneurysm
    Han, Daniel K.
    Chung, Christine
    Walkup, Maggie H.
    Faries, Peter L.
    Marin, Michael L.
    Ellozy, Sharif H.
    [J]. ANNALS OF VASCULAR SURGERY, 2011, 25 (05) : 699.e13 - 699.e16
  • [42] Use of the Gore Viabahn VBX balloon-expandable endoprosthesis in the congenital heart disease population
    Cole, Jason M.
    Salavitabar, Arash
    Armstrong, Aimee K.
    Berman, Darren P.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2019, 94 (03) : 416 - 421
  • [43] Experimental research on balloon-expandable endovascular stent expansion
    Wang, YueXuan
    Yi, Hong
    Ni, ZhongHua
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 2272 - 2275
  • [44] Balloon-expandable stent complications during endovascular grafting
    Ombrellaro, MP
    Stevens, SL
    Kerstetter, K
    Freeman, MB
    Goldman, MH
    [J]. VASCULAR SURGERY, 1996, 30 (04): : 293 - 299
  • [45] Performance of a new generation balloon expandable stent-graft (Gore VBX) as bridging stent for B-EVAR
    Ferrer, Ciro
    Borghese, Ottavia
    Orellana, Bernardo
    Mancusi, Carlotta
    Pennetta, Federico F.
    Coscarella, Carlo
    Giudice, Marco
    Giudice, Rocco
    [J]. JOURNAL OF CARDIOVASCULAR SURGERY, 2024,
  • [46] Outcomes of balloon-expandable versus self-expandable stent graft for endovascular repair of iliac aneurysms using iliac branch endoprosthesis
    Lima, Guilherme B.
    Tenorio, Emanuel R.
    Marcondes, Giulianna B.
    Khasawneh, Mohammad A.
    Mendes, Bernardo C.
    DeMartino, Randall R.
    Shuja, Fahad
    Colglazier, Jill J.
    Kalra, Manju
    Oderich, Gustavo S.
    [J]. JOURNAL OF VASCULAR SURGERY, 2022, 75 (05) : 1616 - +
  • [47] Outcomes of Zone 1 Thoracic Endovascular Aortic Repair With Fenestrated Surgeon-Modified Stent-Graft for Aortic Arch Pathologies
    Li, Xiaoye
    Zhang, Lei
    Song, Chao
    Zhang, Hao
    Xia, Shibo
    Zhu, Longtu
    Guo, Wenying
    Li, Haiyan
    Jing, Zaiping
    Lu, Qingsheng
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2024, 31 (01) : 62 - 68
  • [48] Outcomes of directional branches using self-expandable or balloon-expandable stent grafts during endovascular repair of thoracoabdominal aortic aneurysms
    Tenorio, Emanuel R.
    Karkkainen, Jussi M.
    Mendes, Bernardo C.
    DeMartino, Randall R.
    Macedo, Thanila A.
    Diderrich, Alisa
    Hofer, Jan
    Oderich, Gustavo S.
    [J]. JOURNAL OF VASCULAR SURGERY, 2020, 71 (05) : 1489 - +
  • [49] Balloon-expandable stent graft for coral reef aorta
    Isoda, Ryutaro
    Morita, Ichiro
    Mano, Masayuki
    Ishida, Atsuhisa
    [J]. JOURNAL OF VASCULAR SURGERY CASES INNOVATIONS AND TECHNIQUES, 2020, 6 (03): : 422 - 424
  • [50] Fenestrated Stent-Graft Facilitates Emergency Endovascular Therapy for Blunt Aortic Injury DISCUSSION
    Rodriguez, Aurelio
    Britt, L. D.
    Kurimoto, Yoshihko
    Karmy-Jones, Riyad
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2009, 66 (04): : 978 - 979