In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair

被引:18
|
作者
Torsello, Giovanni Federico [1 ,2 ]
Herten, Monika [2 ,3 ]
Mueller, Markus [4 ]
Andre-Frank [4 ]
Torsello, Giovanni B. [2 ]
Austermann, Martin [2 ]
机构
[1] Charite Univ Med Berlin, Charite Campus Virchow Klinikum, Dept Diagnost & Intervent Radiol & Nucl Med, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] St Franziskus Hosp Munster, Dept Vasc Surg, Munster, Germany
[3] Essen Univ Hosp, Dept Orthoped & Trauma Surg, Essen, Germany
[4] Univ Hosp Munster, Dept Hand Trauma & Reconstruct Surg, Biomech Lab, Munster, Germany
关键词
bridging stent-graft; displacement force; fenestrated stent-graft; in vitro experiment; in vitro model; material fatigue; permeability; stent fracture; ENDOGRAFTS; COMPONENTS; OUTCOMES; SECURE;
D O I
10.1177/1526602819842569
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To evaluate the safety and integrity of a new stent-graft as a potential bridging device in fenestrated stent-grafts using an in vitro fenestrated model. Materials and Methods: Polyester test sheets with ten 6-mm- or 8-mm-diameter fenestrations were used to simulate a fenestrated main body endoprosthesis. In total, 50 Viabahn balloon-expandable (VBX) stent-grafts of varying lengths (29 and 39 mm) and diameters (6, 7, and 8 mm) were implanted in the fitting fenestrations. After release, the 6- and 7-mm-diameter stent-grafts were flared with a 10x20-mm angioplasty balloon; a 12x20-mm balloon was used in the 8-mm-diameter devices. Safety of the devices was defined as absence of fractures detected on radiography or computed tomography (CT), as well as material failure detected by microscopy and water permeability testing. The forces (in Newtons, N) needed for perpendicular dislocation (pullout force) and axial dislocation (shear stress force) were also evaluated. Results: Forty VBX stent-grafts were subjected to digital radiographic imaging and multiplanar CT. None showed any stent fracture. Subsequent microscopy indicated no damage to the fabric or separation of the graft after flaring. Ten VBX stent-grafts underwent water permeability testing after flaring; no water passed through the graft wall during a 10-minute period under an intraluminal pressure at 120 mm Hg. Testing of 25 VBX stent-grafts revealed initial pullout forces between 11.3 and 31 N. Shear stress tests showed that the average force needed to dislocate the stent-grafts by 50% of their diameter ranged between 5.75 and 6.91 N (mean 6.1 +/- 0.5 N) for the 6-mm stents and between 3.31 and 5.4 N (mean 4.4 +/- 0.8) for the 8-mm stents. Conclusion: This preliminary study demonstrated the applicability of the VBX as a bridging stent-graft in a simulated fenestration model. A comparison with other stent-grafts and clinical assessment are required.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 50 条
  • [21] Balloon-Expandable Aortic Stent-Graft for Treatment of Descending Aortic Rupture Presumably Secondary to Spiroserca lupi Larvae Migration
    Avner, A.
    Willenz, U.
    Sabbag, I
    Moreshet, A.
    Bruckheimer, E.
    ISRAEL JOURNAL OF VETERINARY MEDICINE, 2016, 71 (03): : 43 - 47
  • [22] Preliminary Results of Gore Viabahn Balloon-Expandable Endoprosthesis in Visceral and Renal Arteries of Patients Treated With Branched or Fenestrated Endografts for Complex Aortic Aneurysmal Disease
    Masciello, Fabrizio
    Fargion, Aaron
    Giacomelli, Elena
    Dorigo, Walter
    Pratesi, Carlo
    JOURNAL OF VASCULAR SURGERY, 2020, 72 (01) : E54 - E55
  • [23] Balloon expandable aortic stent-graft for the treatment of aortic aneurysm and rupture in a dog
    Klainbart, Sigal
    Segev, Gilad
    Chen, Hilla
    Peery, Dana
    Shipov, Anna
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2021, 31 (06) : 806 - 812
  • [24] Endovascular stent-graft repair of failed endovascular abdominal aortic aneurysm repair
    Baril, Donald T.
    Silverberg, Daniel
    Ellozy, Sharif H.
    Carroccio, Alflo
    Jacobs, Tikva S.
    Sachdev, Ulka
    Teodorescu, Victoria J.
    Lookstein, Robert A.
    Marin, Michael L.
    ANNALS OF VASCULAR SURGERY, 2008, 22 (01) : 30 - 36
  • [25] Percutaneous Balloon-Expandable Stent Graft Repair of a Ruptured Iliac Aneurysm
    Cookson, Daniel
    Raza, Zahid
    VASCULAR AND ENDOVASCULAR SURGERY, 2011, 45 (05) : 462 - 466
  • [26] Endovascular stent-grafting for traumatic aortic aneurysms with the use of a fenestrated stent-graft
    Sueda, T
    Orihashi, K
    Watari, M
    Okada, K
    Ishii, O
    Naito, A
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 122 (01): : 144 - 146
  • [27] Use of a balloon-expandable sheath for transfemoral complex endovascular aortic repair
    Furlough, Courtney L.
    Hoel, Andrew W.
    JOURNAL OF VASCULAR SURGERY, 2019, 70 (06) : 2089 - +
  • [28] Endovascular Repair of a Thoracoabdominal Aortic Aneurysm With a Patient-Specific Fenestrated-ranched Stent-Graft
    Mendes, Bernardo C.
    Greiten, Lawrence E.
    Oderich, Gustavo S.
    JOURNAL OF ENDOVASCULAR THERAPY, 2017, 24 (05) : 665 - 669
  • [29] Preliminary Outcomes of the LifeStream Balloon-Expandable Covered Stent in Fenestrated and Branched Thoracoabdominal Endovascular Repairs
    Bertoglio, Luca
    Loschi, Diletta
    Cambiaghi, Tommaso
    Mascia, Daniele
    Kahlberg, Andrea
    Melissano, Germano
    Chiesa, Roberto
    JOURNAL OF ENDOVASCULAR THERAPY, 2018, 25 (02) : 230 - 236
  • [30] Treatment of focal distal abdominal aortic stenosis with the GORE VIABAHN VBX balloon expandable covered stent
    Whitbeck, Matthew G.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2020, 95 (03) : 457 - 461