EXPOSURE DOSE INDEX BASED ON NOISE FACTOR ANALYSIS IN DIGITAL MAMMOGRAPHY: VERIFICATION USING DIRECT-TYPE FLAT-PANEL DETECTOR SYSTEM

被引:1
|
作者
Maruyama, Sho [1 ]
机构
[1] Gunma Paz Univ, Fac Hlth Sci, Sch Radiol Sci, 3-3-4 Tonyamachi, Takasaki, Gunma 3700006, Japan
关键词
BASIC IMAGING PROPERTIES; PHYSICAL-CHARACTERIZATION; BREAST-CANCER; RADIOGRAPHY; PERFORMANCE; QUALITY; PHANTOMS;
D O I
10.1093/rpd/ncab003
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this study, the development of a reasonable index of optimized exposure dose is attempted. Using a direct-type flat-panel system, noise factors contained in the image are analyzed based on the relative standard deviation method, and it is verified that the proposed index conforms with the appropriate standards regarding minimum exposure dose. The findings indicate that Poisson noise is dominant in the general clinical dose range; this noise fraction formed 90% of the total noise in the system considered in this study. Considering the results of previous surveys, it is estimated that the point at which the content rate of the Poisson noise starts to decrease can be considered as the lower limit, below which the dose cannot be reduced further without compromising on image quality. As the 'index' obtained by this method can be easily measured, it is useful as a quantitative indicator for dose determination and optimization.
引用
收藏
页码:473 / 481
页数:9
相关论文
共 50 条
  • [21] Contrast-to-noise and exposure measurements of an aSi:H/CsI(Tl) flat-panel based digital radiography system using a QC chest phantom
    Liu, XM
    Shaw, CC
    Balter, P
    Kappadath, SC
    Lai, CJ
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 826 - 832
  • [22] Comparison of the AEC Thickness Tracking Using the Signal-Difference-To-Noise Ratio in 113 Flat-Panel Digital Mammography Units
    Hwang, Y.
    Chen, C.
    Tsai, H.
    Chen, C.
    Chui, C.
    Wan, Y.
    Liu, H.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [23] Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detector
    Baldelli, P.
    Phelan, N.
    Egan, G.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988): : 290 - 295
  • [24] Image quality of digital direct flat-panel mammography versus an analog screen-film technique using a phantom model
    Krug, Kathrin Barbara
    Stuetzer, Hartmut
    Girnus, Ralf
    Zaehringer, Markus
    Gossmann, Axel
    Winnekendonk, Guido
    Lackner, Klaus
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (02) : 399 - 407
  • [25] Theoretical Analysis of Hybrid Flat-Panel Detector Arrays for Digital X-Ray Fluoroscopy: General System Architecture, Signal, and Noise Processes
    Suryanarayanan, Sankararaman
    Karellas, Andrew
    Vedantham, Srinivasan
    IEEE SENSORS JOURNAL, 2001, 1 (02) : 168 - 174
  • [26] Incorporation of anatomical noise in generalized DQE analysis of advanced flat-panel detector-based imaging systems
    Richard, S
    Siewerdsen, J
    Jaffray, D
    Moseley, D
    Bakhtiar, B
    MEDICAL PHYSICS, 2004, 31 (06) : 1810 - 1810
  • [27] Monte Carlo analysis of relation between patient dose and noise characteristic of flat-panel detector for cone-beam CT
    Nakamori, N
    Yang, YQ
    Yoshida, Y
    Tsunoo, T
    Endo, M
    Sato, K
    MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 724 - 731
  • [28] Automated analysis for the respiratory kinetics with the screening dynamic chest radiography using a flat-panel detector system
    Tanaka, R
    Sanada, S
    Kobayashi, T
    Suzuki, M
    Matsui, T
    Inoue, H
    CARS 2003: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2003, 1256 : 179 - 186
  • [29] A contrast-detail phantom study to compare a slot scanning digital mammography system with an ASi/CsI flat-panel based system
    Lai, C
    Shaw, C
    Geiser, W
    Liu, X
    Kappadath, S
    Altunbas, M
    Tu, S
    Wang, T
    Whitman, G
    Dempsey, P
    MEDICAL PHYSICS, 2004, 31 (06) : 1796 - 1796
  • [30] Comparison of Digital Mammograms Obtained with the Cassette-type Retrofit Mammography Flat Panel Detector Installed on an Analog System and the Conventional Full-field Digital Mammography
    Seo, Wontaek
    Lee, Su Hyun
    Jung, Kahee
    Jo, Mijung
    Lee, Hyeonran
    Shin, Choul Woo
    Chang, Jung Min
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948