Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detector

被引:38
|
作者
Baldelli, P. [1 ]
Phelan, N. [1 ]
Egan, G. [1 ]
机构
[1] Natl Canc Screening Serv, Dublin, Ireland
来源
BRITISH JOURNAL OF RADIOLOGY | 2010年 / 83卷 / 988期
关键词
SPECTRAL SHAPE; OPTIMIZATION; MODEL;
D O I
10.1259/bjr/60404532
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A comparison, in terms of image quality and glandular breast dose, was carried out between two similar digital mammography systems using amorphous selenium flat panel detectors. The two digital mammography systems currently available from Lorad-Hologic were compared. The original system utilises Mo/Mo and Mo/Rh as target/filter combinations, while the new system uses W/Rh and W/Ag. Images of multiple mammography phantoms with simulated compressed breast thicknesses of 4 cm, 5 cm and 6 cm and various glandular tissue equivalency were acquired under different spectral conditions. The contrast of five details, corresponding to five glandular compositions, was calculated and the ratio of the square of the contrast-to-noise ratio to the average glandular dose was used as a figure-of-merit (FOM) to compare results. For each phantom thickness and target/filter combination, there is an optimum voltage that maximises the FOM. Results show that the W/Rh combination is the best choice for all the detection tasks studied, but for thicknesses greater than 6 cm the W/Ag combination would probably be the best choice. In addition, the new system with W filter presents a better optimisation of the automatic exposure control in comparison with the original system with Mo filter.
引用
收藏
页码:290 / 295
页数:6
相关论文
共 50 条
  • [1] Effect of anode/filter combination on the dose and image quality of a digital mammography system based on an amorphous selenium detector
    Baldelli, Paola
    Phelan, Niall
    Egan, Gillian
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 716 - 723
  • [2] Investigation of imaging performance of amorphous selenium flat-panel detectors for digital mammography
    Zhao, W
    Ji, WG
    Rowlands, JA
    Debrie, A
    [J]. MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 536 - 546
  • [3] Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system
    Paula Toroi
    Federica Zanca
    Kenneth C. Young
    Chantal van Ongeval
    Guy Marchal
    Hilde Bosmans
    [J]. European Radiology, 2007, 17 : 2368 - 2375
  • [4] Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system
    Toroi, Paula
    Zanca, Federica
    Young, Kenneth C.
    van Ongeval, Chantal
    Marchal, Guy
    Bosmans, Hilde
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (09) : 2368 - 2375
  • [5] Image quality and radiation dose on digital chest imaging: Comparison of amorphous silicon and amorphous selenium flat-panel systems
    Bacher, Klaus
    Smeets, Peter
    Vereecken, Ludo
    De Hauwere, An
    Duyck, Philippe
    De Man, Robert
    Verstraete, Koenraad
    Thierens, Hubert
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (03) : 630 - 637
  • [6] Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detector
    Zhao, W
    Ji, WG
    Debrie, A
    Rowlands, JA
    [J]. MEDICAL PHYSICS, 2003, 30 (02) : 254 - 263
  • [7] Digital breast tomosynthesis using an amorphous selenium flat panel detector
    Bissonnette, M
    Hansroul, M
    Masson, E
    Savard, S
    Cadieux, S
    Warmoes, P
    Gravel, D
    Agopyan, J
    Polischuk, B
    Haerer, W
    Mertelmeier, T
    Lo, JY
    Chen, Y
    Dobbins, JT
    Jesneck, JL
    Singh, S
    [J]. Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 529 - 540
  • [8] Reduction of patient dose in digital mammography: Simulation of low-dose image using computed radiography system and flat panel detector system
    Saito, Yuki
    Sakai, Maki
    Fujita, Naotoshi
    Kodera, Yoshie
    [J]. MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [9] Effect of filter on average glandular dose and image quality in digital mammography
    Songsaeng, C.
    Krisanachinda, A.
    Theerakul, K.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [10] Relationships between detector design parameters and Imaging performance of amorphous selenium-based flat-panel detectors for digital mammography
    Zhao, W
    Ji, WG
    Rowlands, JA
    Debrie, A
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 120 - 126