Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system

被引:58
|
作者
Toroi, Paula
Zanca, Federica
Young, Kenneth C.
van Ongeval, Chantal
Marchal, Guy
Bosmans, Hilde
机构
[1] STUK, Radiat & Nucl Safety Author, FIN-00881 Helsinki, Finland
[2] Leuven Univ, Ctr Med Phys Radiol, Dept Radiol, Louvain, Belgium
关键词
digital mammography; optimisation; tungsten; rhodium anode; filter; signal difference-to-noise ratio;
D O I
10.1007/s00330-006-0574-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The signal difference-to-noise ratio (SDNR) between aluminium sheets and a homogeneous background was measured for various radiation qualities and breast thicknesses to determine the optimal radiation quality when using a Novation DR mammography system. Breast simulating phantoms, with a thickness from 2 cm to 7 cm, and aluminium sheet, with a thickness of 0.2 mm, were used. Three different combinations of anode/filter material and a wide range of tube voltages were employed for each phantom thickness. Each radiation quality was studied using three different dose levels. The tungsten (W) anode and rhodium (Rh) filter combination achieved the specified SDNR at the lowest mean glandular dose for all the phantom thicknesses and X-ray tube voltages. The difference between the doses for different anode/filter combinations increased with the phantom thickness. For a 5-cm phantom, with a peak tube voltage of 27 kV and a SDNR of 5, the mean glandular dose associated with the use of W/Rh was reduced by 49% when compared to the molybdenum/molybdenum (Mo/Mo) anode/filter combination and by 33% when compared to Mo/Rh. Based on these measurements, the use of the W/Rh anode/filter can be recommended. It remains important, however, to select the appropriate dose level.
引用
下载
收藏
页码:2368 / 2375
页数:8
相关论文
共 9 条
  • [1] Experimental investigation on the choice of the tungsten/rhodium anode/filter combination for an amorphous selenium-based digital mammography system
    Paula Toroi
    Federica Zanca
    Kenneth C. Young
    Chantal van Ongeval
    Guy Marchal
    Hilde Bosmans
    European Radiology, 2007, 17 : 2368 - 2375
  • [2] Effect of anode/filter combination on the dose and image quality of a digital mammography system based on an amorphous selenium detector
    Baldelli, Paola
    Phelan, Niall
    Egan, Gillian
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 716 - 723
  • [3] Detector evaluation of a prototype amorphous selenium-based full field digital mammography system
    Jesneck, JL
    Saunders, RS
    Samei, E
    Xia, JQ
    Lo, JY
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 478 - 485
  • [4] Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detector
    Baldelli, P.
    Phelan, N.
    Egan, G.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988): : 290 - 295
  • [5] Modulation transfer function of a selenium-based digital mammography system
    Hoheisel, M
    Bätz, L
    Mertelmeier, T
    Giersch, J
    Korn, A
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3589 - 3593
  • [6] Modulation transfer function of a selenium-based digital mammography system
    Hoheisel, M.
    Baetz, L.
    Mertelmeier, T.
    Giersch, J.
    Korn, A.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) : 1118 - 1122
  • [7] Relationships between detector design parameters and Imaging performance of amorphous selenium-based flat-panel detectors for digital mammography
    Zhao, W
    Ji, WG
    Rowlands, JA
    Debrie, A
    IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 120 - 126
  • [8] Evaluation based on breast tomosynthesis of amorphous selenium full field digital mammography system
    Varjonen, M
    Strommer, P
    MEDICAL PHYSICS, 2005, 32 (06) : 1897 - 1897
  • [9] First results with real-time selenium-based full-field digital mammography three-dimensional imaging system
    Lehtimäki, M
    Pamilo, M
    Raulisto, L
    Kalke, M
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 922 - 929