Radiofrequency ablation of paroxysmal atrial fibrillation with the new irrigated multipolar nMARQ ablation catheter: Verification of intracardiac signals with a second circular mapping catheter

被引:18
|
作者
Rosso, Raphael [1 ]
Halkin, Amir
Michowitz, Yoav
Belhassen, Bernard
Glick, Aharon
Viskin, Sami
机构
[1] Tel Aviv Sourasky Med Ctr, Dept Cardiol, IL-64239 Tel Aviv, Israel
关键词
Atrial fibrillation; Radiofrequency ablation; Pulmonary veins; PULMONARY VEIN POTENTIALS; PATIENT;
D O I
10.1016/j.hrthm.2013.12.029
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND During radiofrequency (RE) ablation of paroxysmal atrial fibrillation, a circular multielectrode recording "lasso" catheter is generally positioned within each pulmonary vein (PV) to determine when pulmonary vein potentials (PVPs) are present and when they have been ablated. The new irrigated multipolar nMARQ circular ablation catheter is positioned within the left atrium to create contiguous circular ablation lines around each PV ostium. OBJECTIVE To determine whether the recordings obtained from the nMARQ catheter position around the PV ostium accurately reproduce the recordings obtained from a Lasso catheter positioned within that vein. METHODS In 10 patients undergoing RF ablation of paroxysmaL atrial fibrillation, we placed an nMARQ and a lasso catheter around and within each PV, respectively. Recordings obtained from both catheters at baseline and after RE ablation were compared. RESULTS At baseline, recordings of PVPs in both catheters were concordant in 92% of all PVs. However, after RF delivery, the concordance between the nMARQ and lasso recordings was poor. The discordant result most commonly observed was disappearance of "PVPs" from the nMARQ catheter with persistence of PVPs in the Lasso catheter (12 of 39 [30%]). Conversely, the delivery of RE frequently resulted in fragmented electrograms (pseudo-PVPs) on the nMARQ catheter despite evidence of PV isolation by lasso catheter recordings. CONCLUSIONS The use of an nMARQ catheter alone, as currently recommended, may Lead to underestimation and overestimation of the number of RE applications required to achieve PV isolation.
引用
收藏
页码:559 / 565
页数:7
相关论文
共 50 条
  • [1] Paroxysmal atrial fibrillation ablation with the multipolar mapping and ablation catheter (Mesh-Bard)
    Pratola, Claudio
    Notarstefano, Pasquale
    Artale, Paolo
    Baldo, Elisa
    Toselli, Tiziano
    Ferrari, Roberto
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (06): : 753 - 756
  • [2] Mapping of arrhythmogenic pulmonary veins in patients with paroxysmal atrial fibrillation with a multipolar basket catheter: implications for radiofrequency catheter ablation
    Arentz, T
    Deisenhofer, I
    von Rosenthal, J
    Blum, T
    Shah, D
    Jais, P
    Hocini, M
    Haissaguerre, M
    Kalusche, D
    EUROPEAN HEART JOURNAL, 2001, 22 : 361 - 361
  • [3] SIMULTANEOUS MAPPING AND RADIOFREQUENCY ABLATION OF PAROXYSMAL ATRIAL FIBRILLATION USING HIGH-DENSITY MULTIPOLAR CATHETER
    Vieira, M. S.
    Vieira, P.
    Roque, C.
    Sa, I.
    Palma, P.
    Antunes, N.
    Meireles, A.
    Gomes, C.
    Anjo, D.
    Santos, M.
    Dias, V.
    Carvalho, H.
    Torres, S.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S111 - S111
  • [4] EXPERIENCE WITH A NOVEL IRRIGATED MULTIPOLAR IRRIGATED ABLATION CATHETER (NMARQ) FOR ATRIAL FIBRILLATION ABLATION: ACUTE SAFETY, EFFICACY AND FOLLOW-UP DATA
    Deneke, Thomas
    Mueller, Patrick
    Schade, Anja
    Krug, Joachim
    Schmitt, Rainer
    Roos, Markus
    Szoelloezi, Atilla
    Nentwich, Karin
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 65 (10) : A426 - A426
  • [5] Radiofrequency ablation of paroxysmal atrial fibrillation by mesh catheter
    Claudio Pratola
    Pasquale Notarstefano
    Paolo Artale
    Tiziano Toselli
    Elisa Baldo
    Lina Marcantoni
    Chiara Carrescia
    Paolo Squasi
    Roberto Ferrari
    Journal of Interventional Cardiac Electrophysiology, 2009, 25 : 135 - 140
  • [6] Radiofrequency ablation of paroxysmal atrial fibrillation by mesh catheter
    Pratola, Claudio
    Notarstefano, Pasquale
    Artale, Paolo
    Toselli, Tiziano
    Baldo, Elisa
    Marcantoni, Lina
    Carrescia, Chiara
    Squasi, Paolo
    Ferrari, Roberto
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2009, 25 (02) : 135 - 140
  • [7] Effect of Different Ablation Settings on Acute Complications Using the Novel Irrigated Multipolar Radiofrequency Ablation Catheter (nMARQ)
    Deneke, Thomas
    Mueller, Patrick
    Halbfass, Philipp
    Szoelloesi, Atilla
    Roos, Markus
    Krug, Joachim
    Fochler, Franziska
    Schade, Anja
    Schmitt, Rainer
    Christopoulos, Georgios
    Muegge, Andreas
    Nentwich, Karin
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (10) : 1063 - 1068
  • [8] First Experience of the New nMARQ Multi-Electrode Catheter for Mapping and Ablation of Paroxysmal and Persistent Atrial Fibrillation
    Chin, Shui Hao
    Varanasi, Srinivas
    Chu, Gavin
    Siddiqui, Shoaib
    Sandilands, Alastair
    Stafford, Peter
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2014, 25 (05) : 559 - 559
  • [9] Safety and efficacy of multipolar pulmonary vein ablation catheter vs. irrigated radiofrequency ablation for paroxysmal atrial fibrillation: a randomized multicentre trial
    McCready, J.
    Chow, A. W.
    Lowe, M. D.
    Segal, O. R.
    Ahsan, S.
    de Bono, J.
    Dhaliwal, M.
    Mfuko, C.
    Ng, A.
    Rowland, E. R.
    Bradley, R. J. W.
    Paisey, J.
    Roberts, P.
    Morgan, J. M.
    Sandilands, A.
    Yue, A.
    Lambiase, P. D.
    EUROPACE, 2014, 16 (08): : 1145 - 1153
  • [10] A circular mapping catheter is not mandatory for isolating pulmonary veins during paroxysmal atrial fibrillation ablation with radiofrequency
    Xhaet, Olivier
    Deceuninck, Olivier
    Robaye, Benoit
    Dormal, Fabien
    Collet, Benoit
    Godeaux, Veronique
    Huys, Florence
    Ballant, Elisabeth
    Gourdin, Maximilien
    Blommaert, Dominique
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2021, 62 (02) : 285 - 292